Measuring principle
Raman spectroscopy
-
Temperature
-30 to 150 °C / -22 to 302 °F
-
Sampling probe compatibility
Raman Rxn-10
-
Pressure
Bio Sleeve:
Max pressure: 13.8 barg / 200 psig -
Wetted materials
Bio sleeve:
Body: 316L stainless steel
Window: Proprietary material, optimized for bioprocesses
Surface finish: Ra 15 with Electropolish to ASME BPE SF4 finish
Adhesive: USP Class VI and ISO 10993 compatible -
Length
120 or 220 mm
-
Diameter
Bio sleeve:
12 mm -
Sterilization method
Bio sleeve:
Autoclave, gamma irradiation -
Process connection
Bio Sleeve:
PG13.5
Field of application
The bio multi optic and bio sleeve duo delivers reliable, real-time bioprocess measurements with maximum sterility and efficiency. The non-contact multi optic features a patented self-aligning lens for top spectra quality and reproducibility. It pairs with the Rxn-10 probe and connects to an autoclavable (or gamma sterilizable), disposable sleeve via an easy snap-in-place design. The system reduces maintenance and cross-contamination risk for reusable or single-use benchtop bioreactor setups.
Benchtop bioreactors to measure glucose, lactate, amino acids, cell density, titer and more Benchtop fermenters to measure glycerol, methanol, ethanol, sorbitol, biomass, and more Use with flow assembly CYA680 for select downstream bioprocessing applications
Benefits
Patented self-focusing lens with automatic alignment for optimal spectral quality and reproducibility Improved ease of use with less probe service and maintenance Simplified workflow for greater process efficiency – swap out probes mid-batch, use multiple sleeves in the same reactor for simplified redundancy, and calibrate / verify probes at any time Autoclavable, gamma sterilizable, and compatible with standard cleaning protocols (bio sleeve), with reduced risk of cross-contamination (via non-contact multi optic) 120 mm version is compatible with Endress+Hauser flow assembly CYA680