Measuring principle
Sensor ORP / Redox
-
Application
Flotation, leaching, neutralization, outlet monitoring
-
Characteristic
Resistent against electrolyte poisoning and dirt repellent gel-electrode including process connection NPT 3/4″
-
Measurement range
-1500mV to +1500mV
-
Measuring principle
Gel-compact electrode with double chamber reference system, potassium nitrate bridge electrolyte and PTFE-diaphragm, platinum ring
-
Design
Compact electrode in PPS housing with NPT3/4″ process connection
-
Material
Housing: PPS
ORP electrode: platinum ring
Double chamber reference system: KNO3 and KCl/AgCl -
Dimension
Diameter: 22 mm (0.87 inch)
Length: 150 mm (5.91 inch) -
Process temperature
0 to 80 °C (32 to 170 °F)
-
Process pressure
1 to 10 bar abs at 80 °C
(15 to 145 psi abs at 176 °F) -
Temperature sensor
NTC
-
Ex certification
(optional) FM IS NI Cl. I Div.1&2, Groups A-D
-
Connection
Process connection: NPT 3/4″
Memosens plug -
Ingress protection
IP 68
Field of application
Memosens CPF82E is the digital, robust solution for ORP measurement in harsh environments. Its large, dirt-repellent PTFE junction guarantees low maintenance while its integrated assembly offers easy and compact installation. Thanks to Memosens 2.0 digital technology, CPF82E combines maximum process integrity with simple operation. It resists moisture, enables lab calibration and offers extended storage of calibration and process data providing the perfect basis for predictive maintenance.
Memosens CPF82E is the robust ORP sensor for treatment of industrial water and wastewater and for processes in mining industries. It monitors:Inlet and outletNeutralizationFlotationLeachingWith CSA C/US approval for use in hazardous areas, zones 0, 1 and 2.
Benefits
Memosens 2.0 offers extended storage of calibration and process data, enabling better trend identification and providing a future-proof basis for predictive maintenance and enhanced IIoT services. Long time stable: Double junction provides better protection against electrode poisoning ions such as S2- or CN-. Robust polymer housing protects against mechanical damage. The sensor is suitable for flow and immersion installation Non-contact, inductive signal transmission maximizes process safety.Minimum process downtime and extended sensor lifetime reduce operating costs.