Digital pH sensor Memosens CPS71E

  • Measuring principle

    Potentiometric

  • Application

    Process technology and monitoring of processes with:
    • Rapidly changing pH values
    • High proportion of electrode poisons such as H2S

  • Characteristic

    Digital pH electrode for chemical process with an ion trap for poison-resistant reference

  • Measurement range

    Application B:
    • pH: 0 to 14
    Application H:
    • pH: 0 to 12

  • Measuring principle

    Gel compact electrode with ceramic junction and ion trap

  • Design

    All shaft lengths with temperature sensor
    Advanced gel technology

  • Material

    Sensor shaft: Glass to suit process
    pH membrane glass: Type B, Type N
    Metal lead: Ag/AgCl
    Open aperture: Ceramic junction, zirconium dioxide
    O-ring: FKM
    Process coupling: PPS fiber-glass reinforced
    Nameplate: Ceramic metal oxide

  • Dimension

    Diameter: 12 mm (0.47 inch)
    Shaft lengths: 120, 225, 360 and 425 mm
    (4.72, 8.86, 14.2 and 16.7 inch)

  • Process temperature

    1 to 140 °C (32 to 284 °F)

  • Process pressure

    0,8 to 14 bar (11,6 to 203 psi) (absolute)

  • Temperature sensor

    NTC 30k

  • Ex certification

    With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in
    hazardous areas Zone 0, Zone 1 and Zone 2.

  • Connection

    Inductive, contactless connection head with Memosens 2.0 technology

  • Ingress protection

    IP68

  • Additional certifications

    Additional certifications

Field of application
Memosens CPS71E is specially designed for demanding processes. Its unique, contamination-resistant reference and its resistance to moisture ensure reliable measurement even in heavily polluted, aggressive media such as strong acids and bases. Thanks to Memosens 2.0 digital technology, CPS71E offers extended storage of calibration and process data providing the perfect basis for predictive maintenance. Pre-calibration in the lab and quick sensor exchange on site maximize process uptime.

Processes in chemical industry and monitoring of processes with:High proportion of electrode poisons such as H2SQuickly changing pH values With ATEX, IECEx, CSA C/US, NEPSI, Japan, INMETRO approvals for use in hazardous areas, zones 0, 1 and 2.

Benefits
The electrode’s unique ion trap prevents poisoning of the junction and reference, making it resistant to strong acids and bases and ensuring a long sensor lifetime. The optional pressurized reference allows for reliable measurement in blocking media such as dispersions.Flexible installation thanks to optional upside-down mounting Inductive cable connection and non-contact signal transmission eliminate any problems due to moisture or corrosion and increase process integrity.Fast sensor exchange on site reduces process downtime and operating costs. Memosens 2.0 digital technology makes trend identification and predictive maintenance possible thanks to its extended storage of calibration and process data. The technology paves the way for advanced IIoT services.