Measuring principle
Differential pressure
-
Characteristic
Digital transmitter with metallic measuring diaphragms diaphragm seals and capillaries
Modular transmitter
Long-term stability
Secondary process barrier
enhanced safety via self diagnostic functions. -
Supply voltage
4…20 mA HART
10,5…45V DC (Non Ex):
Ex ia: 10,5…30V DC
PROFIBUS PA:
9…32 V DC (Non Ex)
FOUNDATION Fieldbus:
9…32 V DC (Non Ex) -
Reference Accuracy
0,075% + influence of diaphragm seal
-
Long term stability
0.08 % of URL/ year
0.14 % of URL/ 5 years
0.27 % of URL/ 10 years -
Process temperature
-70°C…400°C
(-94°F…752°F) -
Ambient temperature
-50°C…85°C
(-58°F…185°F) -
Measuring cell
100 mbar…40 bar
(40 inH2O…600 psi) -
Smallest calibratable span
5 mbar (0.075 psi)
-
Vacuum resistance
50 mbar (0.73 psi)
-
Max. Turn down
100:1
-
Max. overpressure limit
160 bar
(2400 psi) -
Process connection
Diaphragm seal (Pancake, threads)
Flanges (DIN, ASME)
Hygienic connections
Tri-Clamp ISO2852 -
Process connection hygienic
Tri-Clamp
DIN11851
NEUMO
Varivent
SMS
DRD
Universal adapter -
Material process membrane
316L, AlloyC,
Tantal,
Gold-Rhodium
PTFE -
Material gasket
None, diaphragm seal welded
-
Fill fluid
Silicone oil,
Inert oil,
Vegetable oil
Low- temperature oil
High- temperature oil -
Material housing
316L, Die-cast aluminum
-
Communication
4…20 mA HART
PROFIBUS PA
FOUNDATION Fieldbus -
Certificates / Approvals
ATEX, FM, CSA, CSA C/US, IEC Ex, INMETRO, NEPSI, EAC, UK Ex
-
Safety approvals
SIL
-
Design approvals
NACE MR0175
EN10204-3.1 -
Hygienic approvals
3A, FDA
-
Marine approvals
GL/ ABS
-
Specialities
Diagnostic functions
TempC Membrane -
Successor
PMD78B
-
Measuring principle
Differential pressure
-
Characteristic / Application
Digital transmitter with metallic measuring diaphragms diaphragm seals and capillaries
Modular transmitter
Long term stability
Enhanced safety via self diagnostic functions
Secondary process barrier