Endress+Hauser CLM223 and CLS21 Wiring & Commissioning Guide: Step-by-Step Setup and Troubleshooting

Installing a conductivity measurement system may appear straightforward: connect the sensor, switch on the transmitter, configure several parameters, and check the reading.

However, experienced instrumentation engineers know that small mistakes during installation can lead to unstable measurements, incorrect temperature compensation, or mismatched 4–20 mA signals.

The Endress+Hauser Liquisys M CLM223 conductivity transmitter and Condumax CLS21 conductivity sensor are a typical example.

Although both devices are widely used in industrial liquid analysis, their wiring arrangements and configuration menus require careful attention.

This guide explains how to connect, configure, calibrate, and troubleshoot a CLM223 and CLS21 measurement system, following a practical commissioning sequence.

1. Understanding the CLM223 and CLS21 Measurement System

The Endress+Hauser CLM223 is a panel-mounted conductivity transmitter designed for industrial process monitoring.

It accepts conductive or inductive conductivity sensors, depending on the instrument version and configuration.

The Condumax CLS21 is a two-electrode conductive conductivity sensor suitable for applications involving moderate to relatively high conductivity levels.

Typical applications include:

  • Industrial process water monitoring

  • Cooling water systems

  • Wastewater treatment

  • Chemical processing

  • Water treatment equipment

  • General liquid conductivity monitoring

Main Technical Specifications

ParameterCLM223 / CLS21
TransmitterEndress+Hauser Liquisys M CLM223
SensorEndress+Hauser Condumax CLS21
Measurement principleConductive conductivity
Sensor designTwo-electrode
Typical cell constantk = 1.0 cm-1
CLS21 measurement range10 µS/cm to 20 mS/cm
Temperature measurementPt100, depending on sensor version
Analog output0/4–20 mA, depending on configuration
Transmitter installationPanel mounting
CLM223 front protectionIP54

Important: Always check the exact transmitter order code, sensor version, and manufacturer documentation before installation. The available functions and electrical connections can vary between configurations.

2. Safety Checks Before Wiring

Before opening the transmitter or connecting the sensor, several safety precautions must be completed.

Disconnect the Power Supply

Ensure that the transmitter is completely disconnected from its electrical supply.

Verify the absence of voltage before touching the terminals.

Only qualified electrical or instrumentation personnel should perform installation and maintenance.

Verify the Supply Voltage

CLM223 transmitters are available in different supply configurations.

Check the nameplate to confirm whether the installed unit requires AC or low-voltage AC/DC power.

Never assume that all CLM223 transmitters operate from the same supply voltage.

Applying an incorrect voltage can permanently damage the electronics.

Check the Sensor and Cable

Before installation, inspect:

  • Sensor body and electrodes

  • Cable insulation

  • Connector condition

  • Cable shielding

  • Terminal identification

  • Moisture or corrosion inside connection points

Damaged insulation and moisture contamination can cause unstable conductivity readings.

3. CLM223 and CLS21 Wiring Connections

Correct wiring is one of the most important steps in the commissioning process.

The CLS21 operates according to the conductive measurement principle.

It must therefore be connected using the appropriate conductive sensor input arrangement.

Do not use wiring instructions intended for inductive conductivity sensors.

3.1 Conductivity Signal Connections

For the CLM223 conductive input, terminals 83 and 84 are associated with the conductivity measurement circuit.

These connections carry the sensor measurement signal.

However, the exact assignment of conductors must be verified using the applicable E+H wiring diagram for the sensor and cable assembly.

Do not identify the connections solely by assuming that one conductor represents an inner electrode and another represents an outer electrode.

3.2 Temperature Sensor Connections

When using a CLS21 fitted with a Pt100 temperature element, the CLM223 uses the designated temperature input terminals.

For the applicable three-wire configuration, terminals 11, 12, and 13 are used.

For the CYK71 cable arrangement shown in the manufacturer’s documentation, the temperature conductor colors are identified as:

TerminalCable identification
11Green (GN)
12White (WH)
13Yellow (YE)

These conductor colors apply to the specified cable arrangement and should not be treated as universal for every sensor cable.

Check the actual cable documentation before making connections.

3.3 Cable Shielding

Conductivity measurements are sensitive to electrical interference, particularly when low conductivity values are being measured.

The sensor cable shield must be terminated according to the manufacturer’s conductive sensor wiring instructions.

For the applicable CLM223 conductive sensor arrangement, the shield connection is associated with the designated PE terminal.

The sensor cable shield, protective earth, and measurement signal conductors must not be confused with one another.

Avoid routing the conductivity sensor cable alongside:

  • Variable-frequency drive output cables

  • High-current motor power cables

  • Switching power conductors

  • Other strong sources of electromagnetic interference

Where possible, use separate cable routes and maintain sufficient separation from electrical power circuits.

3.4 Analog Output Connections

For the standard analog output arrangement:

TerminalFunction
31 (+)Current output 1 positive
32 (-)Current output 1 negative
33 (+)Current output 2 positive, if fitted
34 (-)Current output 2 negative, if fitted

Current output 1 is typically used to transmit conductivity measurements.

If the transmitter is equipped with a second analog output, it may be configured for temperature or another supported measured variable.

Check the specific hardware version before connecting the PLC or DCS.

Engineering tip: Never assume that an unused terminal is available for another purpose. Refer to the terminal diagram supplied with the actual transmitter.

4. Understanding the CLM223 Operating Keys

The CLM223 has several dedicated operating keys.

Understanding their functions helps prevent mistakes during configuration and calibration.

ENTER (E)

Used to:

  • Access configuration menus

  • Confirm parameter changes

  • Navigate between parameter fields

  • Save selected settings

PLUS (+) and MINUS (-)

Used to:

  • Select menu groups

  • Increase or decrease numerical values

  • Change configuration options

  • Navigate through available settings

Pressing PLUS and MINUS simultaneously provides the escape function.

Depending on the current menu level, this returns the operator to the preceding menu or measurement display.

CAL

The CAL key is used to access calibration functions.

REL and AUTO

These keys are associated with relay operation and automatic/manual operating modes where supported.

They are not substitutes for the normal configuration keys.

Accessing Configuration Mode

  1. Start from the measurement display.

  2. Press ENTER.

  3. Enter access code 22 using PLUS and MINUS.

  4. Press ENTER to confirm.

  5. Select the required configuration group.

The transmitter will now allow the relevant parameters to be modified.

Important: Calibration and configuration use different entry keys. Do not confuse the CAL menu with the normal parameter setup menu.

5. Setup 1: Configuring Conductivity Measurement

The SETUP 1 function group defines the basic conductivity measurement parameters.

These settings should be checked before calibration or process commissioning.

A1: Measurement Mode

For the CLS21 conductive sensor, select the appropriate conductive conductivity measurement mode:

cond

The selected mode must match the sensor measurement principle and installed transmitter version.

Changing the measurement mode may affect other stored parameters, so record the existing settings before making modifications.

A4: Conductivity Display Unit

Select a suitable engineering unit according to the expected conductivity range.

Common options include:

  • µS/cm

  • mS/cm

For example, a process operating around 800 µS/cm is normally easier to monitor in µS/cm.

For a process around 12 mS/cm, using mS/cm is more convenient.

A5: Sensor Cell Constant

The cell constant is one of the most important parameters affecting measurement accuracy.

A CLS21 sensor commonly has a nominal cell constant of:

k = 1.000 cm-1

However, the actual cell constant may differ slightly from the nominal value.

For example, if the sensor calibration certificate specifies:

k = 0.998 cm-1

Enter 0.998 rather than automatically using 1.000.

An incorrect cell constant introduces a systematic error into the conductivity measurement.

Practical recommendation: Always obtain the actual sensor cell constant from the sensor certificate or a valid calibration result.

A6: Cable Resistance Compensation

For conductive conductivity measurement, cable resistance can influence accuracy.

This becomes especially relevant when long measurement cables are installed.

The CLM223 provides a cable resistance compensation setting for conductive sensors.

Cable resistance should be determined using the relevant cable specification or measurement procedure.

Do not automatically apply a resistance-per-metre value without confirming whether it represents an individual conductor, conductor pair, or the relevant measurement circuit.

A7: Measurement Damping

Measurement damping helps reduce fluctuations in the displayed conductivity value.

A setting of 1 corresponds to no averaging-related damping.

Higher settings provide a smoother displayed value but reduce the speed at which rapid process changes are reflected.

Use minimal damping for stable measurement conditions.

Increase damping only when necessary, particularly where the measured signal is naturally unstable.

Excessive damping can hide rapid process changes.

6. Setup 2: Temperature Compensation

Conductivity changes with temperature because the mobility of ions in a liquid is temperature-dependent.

Without appropriate temperature compensation, conductivity readings may vary significantly even when the liquid composition remains unchanged.

The CLM223 provides several temperature compensation methods.

B1: Temperature Sensor Type

Select the temperature sensor fitted to the actual CLS21 assembly.

For a CLS21 equipped with a Pt100 element:

B1 = Pt100

Other selections may be available in the transmitter, but the configured value must match the installed sensor.

An incorrect selection can cause temperature measurement errors and therefore incorrect temperature-compensated conductivity readings.

B2: Temperature Compensation Method

Common compensation modes include:

ModeApplication
NoneUncompensated conductivity
LinLinear temperature compensation
NaClSodium chloride-related compensation
PureUltrapure water compensation
PureHUltrapure water compensation for the relevant acidic contamination model
TabUser-defined compensation table, where supported

For many ordinary industrial water applications, linear compensation is a practical starting point.

However, the correct compensation method depends on the actual chemical composition of the liquid.

B3: Temperature Coefficient

When linear compensation is selected, the transmitter uses a temperature coefficient.

An example setting is:

alpha = 2.10 %/K

This indicates the assumed relative conductivity change per degree of temperature change.

The coefficient of 2.10 %/K is not universally applicable.

Different process liquids can have significantly different temperature coefficients.

For accurate monitoring, verify the coefficient using process data, laboratory measurements, or a validated application reference.

B5 and B6: Temperature Adjustment

The temperature input can be checked against a reliable reference thermometer.

If the displayed temperature differs from the reference value, the relevant adjustment procedure can be used.

Before applying a temperature offset, check:

  1. Sensor element selection

  2. Temperature wiring

  3. Cable condition

  4. Sensor immersion

  5. Reference thermometer accuracy

Do not use temperature offset adjustment to conceal an unresolved wiring problem.

B7: Reference Temperature

A commonly used reference temperature for conductivity reporting is:

25 °C

Ensure that the conductivity values expected by the laboratory, PLC, DCS, or quality control system use the same reference conditions.

Otherwise, two correct measurements may appear inconsistent because they are reported at different reference temperatures.

7. Configuring the 4–20 mA Output

The analog output allows the CLM223 to transmit conductivity measurements to a PLC, DCS, recorder, or other control system.

The relevant configuration is performed through the CURRENT OUTPUT function group.

O1: Select the Current Output

Select the required output channel.

For example:

Out1

If a second output is installed, configure it separately as required.

O3: Output Characteristic

For normal linear conductivity transmission, select:

Lin

O311: Current Range

Select:

4–20 mA

This is commonly used in industrial automation systems.

O312: Lower Range Value

Define the conductivity corresponding to 4 mA.

Example:

4 mA = 0 µS/cm

O313: Upper Range Value

Define the conductivity corresponding to 20 mA.

Example:

20 mA = 2000 µS/cm

Example: Checking the Analog Output

With the above configuration, the theoretical current output is:

I = 4 + 16 * (C - Cmin) / (Cmax - Cmin)

Where:

  • I = output current in mA

  • C = measured conductivity

  • Cmin = conductivity at 4 mA

  • Cmax = conductivity at 20 mA

For this example:

I = 4 + 16 * C / 2000

The expected current values are:

ConductivityCurrent Output
0 µS/cm4 mA
500 µS/cm8 mA
1000 µS/cm12 mA
1500 µS/cm16 mA
2000 µS/cm20 mA

These values assume normal measurement operation and a linear output without an active fault-current condition.

Common PLC/DCS Scaling Problem

Suppose the transmitter is configured for:

0–2000 µS/cm

But the PLC is configured for:

0–5000 µS/cm

The transmitter may display the correct conductivity while the PLC displays a completely different value.

This is not necessarily a sensor problem.

It is a scaling mismatch.

Always verify that the lower and upper range values match between the transmitter and the receiving control system.

Also confirm the analog input type, wiring arrangement, and actual measured loop current.

8. Sensor Calibration and Commissioning

The CLS21 sensor is supplied with a defined cell constant.

For a new sensor, the first step is to enter the correct value into the transmitter and verify the measurement.

Additional calibration may be required depending on process accuracy requirements, maintenance history, and quality procedures.

When Should Calibration Be Performed?

Calibration or verification is recommended when:

  • Installing a sensor with uncertain calibration history

  • Replacing an existing sensor

  • Investigating persistent measurement deviation

  • Performing scheduled calibration

  • Returning a cleaned sensor to a critical measurement service

Calibration Procedure

  1. Prepare a suitable certified conductivity standard.

  2. Check the standard’s reference temperature and stated conductivity.

  3. Clean and rinse the sensor according to the manufacturer’s instructions.

  4. Immerse the electrodes correctly in the calibration solution.

  5. Remove trapped air bubbles.

  6. Allow the sensor and solution temperature to stabilize.

  7. Press the CAL key.

  8. Enter access code 22 and confirm using CAL.

  9. Select the appropriate conductive sensor calibration procedure.

  10. Enter the required calibration solution parameters.

  11. Follow the displayed instructions.

  12. Review the resulting cell constant and calibration status.

  13. Accept and save the result only when it is technically reasonable.

The detailed calibration fields may vary with the transmitter software version.

Important Calibration Precautions

Do not use a calibration solution that has been contaminated by previous measurements.

Do not return used solution to its original container.

Avoid direct sunlight or unnecessary heating of the calibration container.

Use a conductivity standard appropriate to the expected sensor range.

After calibration, verify the measured value and record the results.

Note: Air-set calibration procedures intended for inductive conductivity measurement must not be applied to the conductive CLS21 sensor.

9. Troubleshooting Common CLM223 and CLS21 Problems

Problem 1: Conductivity Reading Fluctuates Continuously

Possible causes include:

  • Poor electrical connections

  • Damaged sensor cable

  • Incorrect shielding

  • Moisture ingress

  • Electrical interference

  • Air bubbles around the electrodes

  • Unstable process conditions

Recommended troubleshooting:

  1. Inspect the sensor and cable.

  2. Check terminal tightness.

  3. Verify shield termination against the wiring diagram.

  4. Inspect the installation for electromagnetic interference.

  5. Confirm proper sensor immersion.

  6. Compare the measurement with a stable reference solution.

Do not assume that every fluctuating signal is caused by poor grounding.

Problem 2: Conductivity Reading Is Consistently Incorrect

Possible causes include:

  • Incorrect cell constant

  • Contaminated electrodes

  • Incorrect temperature compensation

  • Unsuitable calibration

  • Air bubbles

  • Actual changes in process composition

Start by checking the A5 cell constant and the temperature compensation settings.

Compare the transmitter reading with a suitable conductivity reference.

If contamination is suspected, clean the sensor using a method appropriate to the deposit and wetted materials.

Problem 3: Temperature Reading Is Incorrect

Check:

  • Pt100 selection in B1

  • Temperature wiring

  • Cable continuity

  • Loose or corroded connections

  • Sensor installation

  • Temperature offset settings

Compare the displayed temperature with a calibrated reference thermometer.

Correct electrical or installation problems before making software adjustments.

Problem 4: PLC Reading Does Not Match the Transmitter

Possible causes include:

  • Different analog scaling ranges

  • Incorrect PLC input configuration

  • Incorrect output channel selection

  • Wiring problems

  • Active fault-current output

  • Incorrect analog input interpretation

Use a suitable current measurement instrument to verify the actual output.

For example, if the transmitter is configured for 0–2000 µS/cm and displays 1000 µS/cm, the expected normal output is approximately 12 mA.

If the measured current is correct but the PLC value is wrong, investigate the PLC scaling.

If the measured current is incorrect, check the transmitter configuration and electrical connections.

Problem 5: Temperature Coefficient Cannot Be Changed

Check the compensation mode in B2.

The linear temperature coefficient setting applies when linear compensation is selected.

With certain other compensation modes, the transmitter uses an internal compensation characteristic rather than the manually entered linear coefficient.

This behavior does not necessarily indicate an instrument fault.

10. Cleaning and Maintaining the CLS21 Sensor

Electrode contamination is a common cause of conductivity measurement problems.

Different deposits require different cleaning methods.

Examples include:

ContaminationGeneral Maintenance Approach
Mineral scaleUse a manufacturer-approved cleaning method suitable for the deposit
Oil or greaseUse a compatible cleaning agent
Biological depositsUse an appropriate approved cleaning procedure
Loose particlesRinse carefully with clean water
Persistent depositsInspect and clean according to the sensor documentation

Always consider the chemical compatibility of the sensor materials before using cleaning chemicals.

Never assume that acid cleaning is appropriate for every type of contamination.

After cleaning:

  1. Rinse the sensor thoroughly.

  2. Inspect the electrodes.

  3. Check for mechanical damage.

  4. Reinstall the sensor correctly.

  5. Verify its response.

  6. Perform calibration if required.

Regular inspection can prevent unexpected measurement errors and unnecessary sensor replacement.

11. Factory Reset and Parameter Backup

The CLM223 provides reset functions through the SERVICE menu.

The S9 function includes different reset options.

Typical selections include:

  • No: No reset

  • Sens: Reset relevant sensor calibration data

  • Facty: Restore the applicable factory settings

A factory reset can remove important user-defined parameters, although certain settings are excluded according to the manufacturer’s documentation.

Therefore, a reset should never be the first troubleshooting action.

Before performing any reset, record the existing configuration.

Recommended Parameter Backup Checklist

ParameterInformation to Record
A1Measurement mode
A5Cell constant
A6Cable resistance
A7Measurement damping
B1Temperature sensor type
B2Compensation mode
B3Temperature coefficient
B7Reference temperature
O1/O2Output selection and measured variable
O311Current range
O312Lower output value
O313Upper output value

Also document relevant alarm, relay, HOLD, and communication settings if the installed transmitter uses these functions.

Taking photographs of the configuration screens is a simple and effective way to preserve settings during maintenance.

12. Final Commissioning Checklist

Before handing over the conductivity measurement system, verify the following items.

Inspection ItemRequirement
Transmitter supplyMatches the nameplate
Sensor compatibilityConfirmed for conductive measurement
Conductivity wiringVerified against the applicable wiring diagram
Temperature wiringCorrect for the installed sensor
Cable shieldingInstalled according to manufacturer instructions
Cell constantMatches certificate or valid calibration
Temperature compensationAppropriate for the liquid
Reference temperatureMatches process requirements
Analog outputCorrect range and measured variable
PLC/DCS scalingMatches transmitter configuration
Sensor installationProper immersion, no trapped air
Measurement verificationAcceptable against reference
Parameter backupCompleted
Alarm and control functionsChecked where applicable

A properly commissioned measurement system should provide stable conductivity readings, reasonable temperature measurements, and consistent analog output signals.

Conclusion

Successful commissioning of an Endress+Hauser CLM223 and CLS21 conductivity measurement system depends on more than simply connecting the sensor and entering a few parameters.

The most important steps are:

  1. Verify the sensor and transmitter compatibility.

  2. Follow the correct conductive sensor wiring diagram.

  3. Enter the actual sensor cell constant.

  4. Configure the appropriate temperature compensation.

  5. Match the analog output range with the PLC or DCS.

  6. Verify the measurement using a suitable reference.

  7. Record the final configuration before handing over the system.

When troubleshooting, always start with basic installation checks before replacing the transmitter or sensor.

A loose terminal, incorrect cell constant, unsuitable compensation coefficient, or PLC scaling mismatch can produce symptoms that look like serious instrument failure.

A systematic commissioning procedure saves maintenance time, reduces unnecessary replacement costs, and improves long-term measurement reliability.

For industrial instrumentation engineers, a few additional minutes spent verifying wiring, calibration parameters, and output signals can prevent hours of troubleshooting later.


Technical References

  • Endress+Hauser — Liquisys M CLM223/253 Operating Instructions, BA00193C

  • Endress+Hauser — Liquisys M CLM223/253 Technical Information, TI00193C

  • Endress+Hauser — Condumax CLS21 Sensor Technical Documentation

  • Endress+Hauser — Conductivity Measurement Cable and Installation Documentation

Technical note: This article provides general commissioning guidance. Exact connections, available menus, and calibration procedures depend on instrument hardware, software version, and sensor configuration. Always follow the official documentation supplied with the equipment.