Aluminum analyzer Liquiline System CA80AL

  • Measuring principle

    Colorimetric

  • Characteristic

    Process analyzer for aluminum in aqueous solutions
    Comply with standard colorimetric measuring principle – pyrocatechol violet methot

  • Size

    Housing (open version):
    793 x 530 x 417 mm
    31.22 x 20.87 x 16.42 in
    Housing (closed version):
    793 x 530 x 463 mm
    31.22 x 20.87 x 18.23 in
    Housing with base:
    1723 x 530 x 463 mm
    67.83 x 20.87 x 18.23 in

  • Design

    Open design, cabinet and stand housing
    High-Performance plastic ASA-PC, additional stand coated steel

  • Process temperature

    4 … 40 °C (39.2 … 104 °F)

  • Ambient temperature

    5 … 40C ( 41 … 104F)
    Outdoor version: – 20 to 40 °C (-4 to 104 °F)

  • Process pressure

    at atmospheric pressure, < 0.2 bar

  • Sample flow rate

    min. 5ml/min (0,17 fl.oz/min)

  • Consistency of the sample

    low solid content (< 50NTU), aqueous homogeneous

  • Application

    Monitoring of the aluminum content of potable water, mineral water and industrial waste water

  • Power supply

    100 … 120 VAC / 200 … 240 VAC ± 10%
    24 VDC ± 10%
    50 ± 1 oder 60 ± 1,2 Hz

  • Output / communication

    2x 0/4 … 20 mA
    Webserver, Modbus, Ethernet/IP, Profibus DP

  • Input

    1 or 2 measuring channel
    1 … 4 digital sensor inputs for sensors with Memosens protocol (optional)

  • Measurement range

    15 to 1000 µg/l Al
    15 to 1000 µg/l with dilution function to maximum 300 to 20 000 µg/l Al

  • Consumables

    Reagents and standard solutions CY80AL as well as cleaner CY800 are necessary for the operation. Regular maintenance is done with the parts of the maintenance kit CAV800.

Field of application
Liquiline System CA80AL offers precise, regulation-compliant online measurement of aluminum residues in drinking water and wastewater. The analyzer helps you to optimize aluminum dosing in flocculation processes and phosphate removal. It enables plug & play of up to 4 Memosens sensors – minimizing the installation effort. Automatic calibration and cleaning and the low reagent consumption reduce operating costs while advanced diagnostics with remote access ensure process safety.

Liquiline System CA80AL aluminum analyzer supports:Limit monitoring in drinking waterOptimization of aluminum-based coagulation processesPhosphate elimination in wastewater treatment plants

Benefits
The aluminum analyzer uses the pyrocatechol violet method according to DIN ISO 10566, ensuring direct comparability to lab results
Supports predictive maintenance by advanced diagnostics Reduces operating costs thanks to automatic calibration and cleaning as well as low reagent consumption
Easy upgrade of functionality to a complete measuring station – by adding modules and connecting Memosens sensors. Comfortable remote access via integrated web serverSeamless integration into process control systems thanks to digital field buses such as Modbus, PROFIBUS, EtherNet/IP

Spread the love

Contact Us

    Please prove you are human by selecting the cup.