Chromate analyzer Liquiline System CA80CR

  • Measuring principle

    Colorimetric

  • Characteristic

    Analyzer for Chromate (Chrome VI) in aqueous solutions

  • Size

    Housing (open version):
    793 x 530 x 417 mm
    31.22 x 20.87 x 16.42 in
    Housing (closed version):
    793 x 530 x 463 mm
    31.22 x 20.87 x 18.23 in
    Housing with base:
    1723 x 530 x 463 mm
    67.83 x 20.87 x 18.23 in

  • Design

    Open design, cabinet and stand housing
    High-Performance plastic ASA-PC, additional stand coated steel

  • Process temperature

    4 to 40 °C (39 to 104 °F)

  • Ambient temperature

    5 to 40 °C (41 to 104 °F)
    Outdoor version: – 20 to 40 °C (-4 to 104 °F)

  • Process pressure

    At atmospheric pressure, < 0.2 bar absolute

  • Sample flow rate

    Min. 5 ml/min (0.17 fl.oz/min)

  • Consistency of the sample

    Suspended solids content
    Turbitity < 50 NTU, aqueous, homogenized

  • Specials

    Easy upgrade to measuring station with up to four digital Memosens sensors
    Automatic calibration and cleaning
    User-configurable measuring, cleaning and calibration intervals
    Optional cooling module for longer reagent shelf life
    2 channel version optional
    User-definable measuring ranges
    Modular design for easily extensible functionality
    Digital communication for remote access

  • Application

    Limit value monitoring of Chromate in potable water and waste water
    Common applications are industrial waste water Monitoring for tanneries and the galvanic industry

  • Power supply

    100 to 120 VAC / 200 to 240 VAC ± 10%
    24 VDC ± 10%
    50 ± 1 or 60 ± 1,2 Hz

  • Output / communication

    2x 0/4 to 20 mA
    Modbus RS485/TCP (optional)
    Ethernet (optional)
    Alarmrelay

  • Input

    1 or 2 measuring channels
    1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)

  • Measurement range

    0,03 to 2.5 mg/l Cr(VI)
    0.2 to 5.0 mg/l Cr(VI)
    0.2 to 5.0 mg/l with dilution function to maximum 10 to 250 mg/l Cr(VI)

  • Consumables

    Reagents and standard solutions CY80CR as well as cleaner CY800 are necessary for the operation
    Regular maintenance is done with the parts of the maintenance kit CAV800

Field of application
Liquiline System CA80CR offers highly accurate online measurement for optimized chromate removal in industrial wastewater. It supports you in achieving environmentally compliant electroplating and tanning processes. Like all Liquiline System analyzers, CA80CR enables plug & play of up to four Memosens sensors – minimizing the installation effort. Automatic calibration and cleaning plus advanced diagnostics with remote access ensure consistent process safety.

Liquiline System CA80CR supports you in improving the environmental performance of your processes by:
Monitoring and optimizing the cleaning capacity of industrial wastewater treatment plants.Effluent monitoring in tanneries and electroplating plants.
Monitoring of membrane filtration.Limit monitoring in drinking water applications.

Benefits
Standardized colorimetric measuring principle – diphenyle carbazide method – guarantees compliance with discharge regulations. Reduced operating costs through automatic calibration and cleaning and low reagent consumption. Simple maintenance: no tools required. Advanced diagnostics with remote access for increased process safety. Fast commissioning and plug & play thanks to Memosens technology and user-friendly operation. Easy upgrade of functionality – even to a complete measuring station – simply by adding modules and connecting Memosens sensors. Reduces installation effort.