Measuring principle
Potentiometric
-
Application
Long-term monitoring and limit control in processes with stable process conditions
• Chemical processes
• Pulp and paper industry
• Waste incinerators and gas scrubbers
• Mining
• Wastewater -
Characteristic
Combination of pH and ORP electrode for standard applications in process technology and environmental engineering
Poison-resistant reference with ion trap -
Measurement range
ORP: –1 500 to 1 500 mV
Application A
• pH: 1 to 12
Application B
• pH: 0 to 14 -
Measuring principle
Gel compact electrode with PTFE ring-junction and with ion trap
Pt-disc as additional ORP element
rH measurement and control of the reference impedance -
Design
All shaft lengths with temperature sensor
Advanced gel technology -
Material
Sensor shaft: Glass to suit process
pH membrane glass: Type A, B
Metal lead: Ag/AgCl
Open aperture: PFTE junction, sterilizable
ORP measuring element: Platinum
O-ring: FKM
Process coupling: PPS fiber-glass reinforced
Nameplate: Ceramic metal oxide -
Dimension
Diameter: 12 mm (0.47 inch)
Shaft lengths: 120, 225 and 360 mm
(4.72, 8.86, 14.17 and 16.73 inch) -
Process temperature
Application A: –15 to 80 °C (5 to 176 °F)
Application B: 0 to 135 °C (32 to 275 °F) -
Process pressure
0.8 to 17 bar (11.6 to 246,5 psi) absolute
-
Temperature sensor
NTC 30k
-
Ex certification
With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in
hazardous areas Zone 0, Zone 1 and Zone 2 -
Connection
Inductive, digital connection head with Memosens 2.0 technology
-
Ingress protection
IP68
-
Measuring principle
Sensor ORP / Redox
-
Application
Long-term monitoring and limit control in processes with stable process conditions
Chemical processes
• Pulp and paper industry
• Waste incinerators and gas scrubbers
• Mining
• Wastewater -
Characteristic
Combination of pH and ORP electrode for standard applications in process technology and environmental engineering
Poison-resistant reference with ion trap -
Measurement range
ORP: –1 500 … 1 500 mV
Application: A
• pH: 1 to 12
Application B
• pH: 0 to 14 -
Measuring principle
Gel compact electrode with PTFE ring-junction with ion trap
Pt-disc as additional ORP element
rH measurement and control of the reference impedance -
Design
All shaft lengths with temperature sensor
Advanced gel technology -
Material
Sensor shaft: Glass to suit process
pH membrane glass: Type A, B
Metal lead: Ag/AgCl
Open aperture: PFTE junction, sterilizable
ORP measuring element: Platinum
O-ring: FKM
Process coupling: PPS fiber-glass reinforced
Nameplate: Ceramic metal oxide -
Dimension
Diameter: 12 mm (0.47 inch)
Shaft lengths: 120, 225 and 360 mm
(4.72, 8.86, 14.17 and 16.73 inch) -
Process temperature
Application A: –15 to 80 °C (5 to 176 °F)
Application B: 0 to 135 °C (32 to 275 °F) -
Process pressure
0.8 to 17 bar (11.6 to 246,5 psi) absolute
-
Temperature sensor
NTK 30k
-
Ex certification
With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in
hazardous areas Zone 0, Zone 1 and Zone 2 -
Connection
Inductive, digital connection head with Memosens 2.0 technology
-
Ingress protection
IP68
Field of application
Memosens CPS16E is suitable for simultaneous pH and ORP measurement in applications with long-term stable conditions. Thanks to its pH and ORP elements, the sensor offers a better process overview and enables tighter process control. Memosens CPS16E features Memosens 2.0 digital technology, offering extended data storage for predictive maintenance. The pH/ORP sensor resists moisture and enables lab calibration which makes the operator’s daily tasks easier and maximizes process integrity.
Long-term monitoring or limit control in processes with stable process conditions:- Chemical processes- Pulp and paper industry- Power stations (e.g. flue gas cleaning)- Incinerator plants- Mining- Wastewater
Water treatment:- Drinking water- Cooling water- Well water With ATEX, IECEx, CSA C/US, NEPSI, Japan, INMETRO approvals for use in hazardous areas, zones 0, 1 and 2.