Measuring principle
Amperometric oxygen measurement
-
Application
Typical applications are:
– Wastewater treatment plants: oxygen control in aeration basin, process water treatment and monitoring
.
– Water plants: status monitoring of drinking water, water quality monitoring in rivers, lakes or seas
.
– Utilities of all industries: Oxygen control in biological treatment, process water treatment and monitoring. -
Characteristic
Digital amperometric 3-electrode longterm-stable sensor to measure dissolved oxygen
-
Measurement range
0.01 to 100 mg/l
0.00 to 1000 %SAT
0 to 2000 hPa -
Measuring principle
Measuring principle
-
Design
Design
-
Material
Sensor shaft: POM
Membrane cap: POM
Working electrode: Gold
Counter and reference electrode: silver/silver halide
Membrane: ETFE (COS51-TN), FEP (COS51-TF) -
Dimension
Diameter: 40mm (1.57 inch)
Shaft length: 146 mm (5.74 inch) -
Process temperature
-5 to 60 °C
(20 to 140 °F) -
Process pressure
Max. 5 bar abs
(Max. 72.5 psi abs) -
Temperature sensor
NTC 30k
-
Ex certification
With ATEX, IECEx, CSA C/US, NEPSI, JapanEx and INMETRO approval for use in Zone 0, Zone 1 and Zone 2 hazardous areas. With CSA
C/US approval also in Class I Division 1 hazardous areas in the
Gas Ex area. Additionally suitable for Class I Division 2. -
Connection
Inductive, contactless connection head with Memosens 2.0 technology
-
Ingress protection
IP 68
Field of application
Memosens COS51E is perfectly suited for all wastewater and utility applications including hazardous areas. The sensor is highly accurate and moisture-resistant ensuring the safety of your processes. COS51E features Memosens 2.0 digital technology. It is able to store more calibration and process data and thus provides the perfect basis for predictive maintenance and IIoT services. Designed to be low-maintenance with a long operating life, the sensor offers outstanding value for money.
Memosens COS51E measures the dissolved oxygen concentration in:
Wastewater treatment plants:- O2 control in the aeration basin- Process water treatment and monitoring
Utilities of all industries:- O2 control in the biological treatment- Process water treatment and monitoring Water works:- Status monitoring of drinking water (oxygen enrichment, corrosion protection, etc.)- Water quality monitoring in rivers, lakes or seas
The sensor is suitable for hazardous area use (ATEX, FM, CSA).
Benefits
Wide measuring range: The amperometric measuring principle and cathode design make the sensor suitable for all wastewater applications from industrial to municipal plants. High accuracy: Thanks to its amperometric, three-electrode design, the sensor provides long-term stable measurement with low drift. Increased plant availability: Memosens technology and Liquiline transmitters enable plug & play for fast commissioning and sensor exchange. Non-contact, inductive signal transmission ensures maximum process integrity.Memosens 2.0 digital technology makes the sensor ready for predictive maintenance and IIoT services since it offers extended storage of calibration and process data.