Measuring principle
Ozone
-
Application
Drinking water – ensures adequate disinfection
Process water – for hygienic packaging and filling
Wastewater – to ensure safe discharge water -
Characteristic
Amperometric measurement of dissolved ozone
-
Measurement range
0 to 2 mg/l
-
Measuring method
– closed 2 electrode system
– conversion of ozone into a signal current in nA
– signal current is proportional to the concentration of ozone
– the measurement is not pH dependent -
Design
– exchangeable, electrolyte-filled membrane cap
– working and counter electrode
– adaptable shaft for flexible installation in assemblies -
Material
Sensor shaft: PVC
Membrane: Silicone
Membrane cap: PVC -
Dimension
Diameter: 25 mm (0.98 in)
Length: 161 mm (6.34 in) -
Process temperature
+0 to 45 °C (32 to 110 °F), non freezing
-
Process pressure
1 bar (14.5 psi), 2 bar (29 psi) absolute
-
Temperature sensor
10k NTC integrated (Memosens)
-
Connection
Inductive, digital connection head with Memosens
Field of application
Memosens CCS58D is a low-maintenance sensor for ozone. It provides high-precision measurement with an almost exclusive specificity for ozone, ensuring reliable disinfection monitoring. The robust sensor membrane with its high surfactant resistance makes the sensor suitable for diverse applications, e.g. bottle cleaning. Memosens CCS58D is not the latest Memosens generation. To get information on the new Memosens CCS58E sensor with extended functionality, click here.
Memosens CCS58D is an ozone sensor for skid builders and end customers. It measures reliably in:Drinking water – to ensure sufficient disinfectionProcess water – to provide hygienic packaging and bottlingWastewater – to guarantee safe discharge water
Benefits
Almost exclusive specificity for ozone ensures reliable measured values for safe disinfection processes.Robust membrane technology guarantees high surfactant resistance and optimum suitability for cleaning processes such as bottle cleaning. Integrated Heartbeat Technology provides, for example, an electrolyte counter for predictive, application-specific maintenance intervals. Connection to the Liquiline multiparameter transmitter allows for easy combination with other relevant parameters of liquid analysis such as pH and ORP. Fast commissioning is ensured by factory-calibration and plug & play installation of the sensor.