Measuring principle
Hydrostatic pressure
-
Characteristic
Compact pressure transmitter for hydrostatic level measurement with flush mounted metallic Contite measuring cell:
Hermetically sealed, condensate-resistant, climatic-proofed and with lowest temperature influences -
Supply voltage
4…20 mA HART
10,5…45 VDC (Non Ex):
Ex ia: 10,5…30 VDC
PROFIBUS PA:
9…32 VDC (Non Ex)
FOUNDATION Fieldbus:
9…32 VDC (Non Ex) -
Reference Accuracy
Standard 0.2 %
Optional 0.1 % -
Long term stability
0.05 % of URL/ year
-
Process temperature
-10°C…100°C
(14°F…212°F) -
Ambient temperature
-40°C…85°C
(-40°F…185°F) -
Measuring cell
100 mbar…10 bar
(1.5 psi…150 psi) -
Smallest calibratable span
10 mbar (1.45 psi)
-
Vacuum resistance
0 mbar abs.
-
Max. Turn down
100:1
-
Max. overpressure limit
40 bar (600 psi)
-
Process connection
Thread:
G1 1/2, MNPT1 1/2
Flange:
DN40…DN100,
ASME 1 1/2″…4″,
JIS 10K -
Process connection hygienic
Tri-Clamp
DIN11851
DIN11864-1
NEUMO
Varivent N
SMS
DRD -
Material process membrane
316L, AlloyC
Gold-Rhodium -
Material gasket
None, measuring cell welded
-
Fill fluid
Inert oil,
Synthetic oil -
Material housing
316L, Die-cast aluminum
-
Communication
4…20 mA HART
PROFIBUS PA
FOUNDATION Fieldbus
IO-Link -
Certificates / Approvals
ATEX, FM, CSA, CSA C/US, IEC Ex, INMETRO, NEPSI, UK Ex
-
Safety approvals
Overfill protection WHG
SIL -
Design approvals
EN10204-3.1
NACE MR0103 -
Hygienic approvals
3A, EHEDG
CoC ASME-BPE -
Marine approvals
GL/ ABS/ LR/ BV/ DNV
-
Drinking water approvals
KTW/ NSF/ ACS
-
Specialities
Modularity to differential pressure and process pressure devices (replacable display, universal electronics)
Diagnostic functions
-
Measuring principle
Hydrostatic
-
Characteristic / Application
Compact pressure transmitter for hydrostatic level measurement with flush mounted metallic Contite measuring cell:
Hermetically sealed, condensate-resistant and climatic-proofed -
Specialities
Modularity to differential pressure and process pressure devices (replacable display, universal electronics)
Diagnostic functions
Hermetically sealed Contite measuring cell (condensate-resistant and climatic-proofed) -
Supply / Communication
4…20mA HART:
10,5…45V DC
Ex ia: 10,5…30V DC
PROFIBUS PA /
FOUNDATION Fieldbus:
9…32V DC (Non Ex) -
Accuracy
Standard 0.2%
Optional 0.1% -
Long term stability
0.05% of URL/year
-
Ambient temperature
-40°C…85°C
(-40°F…185°F) -
Process temperature
-10°C…100°C