Liquid Analysis Physical Medium Properties Liquitrend QMW43

  • Measuring principle

    Medium consistency

  • Measuring method

    Capacitive- conductive

  • Application

    Continuous measurement of build-up and liquid conductivity

  • Supply voltage

    10…30 VDC
    IO-Link: 18…30 VDC

  • Measurement range

    Conductivity:
    0 µS/ cm…100 mS/ cm
    Build-up thickness:
    0 mm…10 mm

  • Accuracy

    Build-up:
    <= 1% of measuring range
    Conductivity:
    0…2 mS/ cm: <= 5% of reading
    2…20 mS/ cm: <= 7% of reading
    20…50 mS/ cm: <= 10% of reading
    50…100 mS/ cm: <= 15% of reading

  • Ambient temperature

    -40°C…+70°C (-40°F…+158°F)

  • Process temperature

    -20°C…+100°C
    (-4°F…+212°F)
    +150°C (+302°F) for 1h

  • Process pressure

    Vacuum…25 bar
    (Vacuum…362.5 psi)

  • Output / communication

    4…20 mA
    Frequency
    IO-Link

  • Connection

    Threads G1″, G3/4″, M24,
    Tri-Clamp ISO2852, DIN11851

  • Ex certification

    No Ex certificates
    CSA C/US General Purpose

  • Additional certifications

    EN10204- 3.1
    EHEDG
    3-A
    EG1935/2004
    CRN approval

  • Options

    Customer specific configuration
    Cleaned from oil and grease
    Surface roughness test
    Welding- and process adapters

Field of application
The Liquitrend QMW43 is developed for the food and beverage industry to help customers to ensure product safety, improve quality and increase the availability of installations. Base your decisions on the unique and reliable measurement of buildup thickness and conductivity via the combination of conductive and capacitive measuring modes.

Cleaning optimization, blending homogeneity monitoring and media differentiation of liquids, suspensions, pasty media etc.For the use in storage, mixing and process vessels with or without an agitator, pipes and equipment peripheryProcess connections: Threads and hygienic connectionsSensor design: Flush-mounted, surface roughness Ra<0.76µm Temperature: -20 to +100°C (-4 to +212°F), CIP/SIP to 150°C (302°F) for 1 hourCertificates and declarations: 3A, FDA, EHEDG and EG1935/2005, EN10204- 3.1

Benefits
Reliable, repeatable and accurate measurement of buildup thickness and conductivity regardless of media type Conductive as well as capacitive measuring cycles, the device automatically detects the most accurate measuring mode and uses it automatically Verification of cleaning status without the need to open the tank Compact and flush-mounted stainless-steel sensor offers the perfect insight into the conditions in pipelines or tanks at all critical locations Plug and Play – simple commissioning without presettings Hygienic safety by Design and material traceability are a matter of course Integrated digital communication option via IO-Link