Measuring principle
Differential conductivity
-
Characteristic
Total carbon (TOC) analyzer for trace levels
-
Measuring method
TOC determination by UV digestion and measurement of the differential conductivity
-
Size
Housing:
500 x 290 x 200 mm
19.68 x 11.41 x 7.87 in -
Design
Stainless steel housing;
IP 42 (standard), IP54 (optional) -
Process temperature
< 50 °C (122 °F)
-
Ambient temperature
10 to 45 °C (50 to 113 °F)
-
Process pressure
max. 0.5 bar (7.25 psi)
-
Sample flow rate
Min. 5 ml/min (0.17 fl.oz/min)
-
Consistency of the sample
max. conductivity 2 μS/cm, particle free
-
Specials
The operating concept and data storage meet the requirements of 21 CFR, Part 11
UV reactor with continuous function monitoring -
Application
Determination of total carbon in ultrapure water applications, e.g. in the lifescience industry, that meet the following conditions:
Conductivity < 2 μS/cm
pH range: neutral -
Power supply
100/240 V AC, 47 – 63 Hz
-
Output / communication
0/4 to 20 mA, galvanically isolated
-
Input
1 to optional 3 measuring channels
Optional control input 24 V (for 1 channel instruments) -
Measurement range
0.5 to 1 000 μg/l (ppb)
Field of application
The total organic carbon (TOC) content is an important quality parameter in the production and use of ultrapure water. If the TOC content is too high, the performance of water purification systems can be impaired or pharmaceutical batches can even be contaminated. The CA79 online TOC analyzer provides continuous, accurate TOC monitoring of your water for injection (WFI). Your product batches can be produced safely and in compliance with regulations.
The CA79 TOC analyzer is perfectly suited for TOC monitoring of ultrapure water in life sciences applications with the following process conditions:Conductivity < 2 µS/cmneutral pH range
Benefits
Real-time overview of water quality: The CA79 TOC analyzer measures continuously providing a fast response time (t90) of 50 seconds. This enables you to react immediately and to protect your product effectively.The online analyzer uses proven UV oxidation and differential conductivity measurement which has become the most-established method for TOC trace analysis in ultrapure water.Compliance:CA79 meets the requirements of the European and US Pharmacopeias and allows working according to FDA 21 CFR Part 11. CA79 provides clear documentation of relevant events, regular quality reports and system suitability tests (SSTs).The analyzer’s maintenance-friendly design in combination with our worldwide service network offers you complete support for the measuring point, including the installation qualification (IQ) and regular operational qualifications (OQ).