Nitrite analyzer Liquiline System CA80NO

  • Measuring principle

    Colorimetric

  • Characteristic

    Process analyzer for Nitrite in potable water

  • Size

    Housing (open version):
    793 x 530 x 417 mm
    31.22 x 20.87 x 16.42 in
    Housing (closed version):
    793 x 530 x 463 mm
    31.22 x 20.87 x 18.23 in
    Housing with base:
    1723 x 530 x 463 mm
    67.83 x 20.87 x 18.23 in

  • Design

    Open design, cabinet and stand housing
    High-Performance plastic ASA-PC, additional stand coated steel

  • Process temperature

    4 to 40 °C (39 to 104 °F)

  • Ambient temperature

    5 to 40 °C (41 to 104 °F)
    Outdoor version: – 20 to 40 °C (-4 to 104 °F)

  • Process pressure

    At atmospheric pressure, < 0.2 bar absolute

  • Sample flow rate

    Min. 5 ml/min (0.17 fl.oz/min)

  • Consistency of the sample

    Suspended solids content
    Turbitity < 50 NTU, aqueous, homogenized

  • Specials

    Easy upgrade to measuring station with up to four digital Memosens sensors
    Automatic calibration and cleaning
    User-configurable measuring, cleaning and calibration intervals
    Optional cooling module for standard solution
    2 channel version optional
    User-definable measuring ranges
    Modular design for easily extensible functionality

  • Application

    Limit value monitoring of nitrite for potable water and mineral water applications

  • Power supply

    100 to 120 VAC / 200 to 240 VAC ± 10%
    24 VDC ± 10%
    50 ± 1 or 60 ± 1,2 Hz

  • Output / communication

    2x 0/4 to 20 mA
    Modbus RS485/TCP (optional)
    Ethernet (optional)
    Alarmrelay

  • Input

    1 or 2 measuring channel
    1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)

  • Measurement range

    10 to 500 µg/l NO2-N
    0.2 to 3.0 mg/l NO2-N
    0.1 to 1.0 mg/l NO2-N
    0.1 to 1.0 mg/l with dilution function to maximum 5 to 50 mg/l NO2-N

  • Consumables

    Reagents and standard solutions CY80NO as well as cleaner CY800 are necessary for the operation
    Regular maintenance is done with the parts of the maintenance kit CAV800

Field of application
Liquiline System CA80NO offers precise, regulation-compliant online measurement of nitrite. It supports you in meeting the specified limits in drinking water, mineral water or food production. Like all Liquiline System analyzers, it enables plug & play of up to four Memosens sensors – minimizing the installation effort. Automatic calibration and cleaning and the low reagent consumption reduce operating and maintenance costs. Advanced diagnostics with remote access ensure process safety.

Liquiline System CA80NO supports you in keeping the strict nitrite limits for: Drinking waterMineral waterRaw water for food production

Benefits
Compliance with standard colorimetric measuring principle – naphthylamine method – following ISO 6777 and DIN EN 26777 ensures comparability to lab measurements. Reduced operating costs through automatic calibration and cleaning. Simple maintenance: no tools required. Advanced diagnostics with remote access for increased process safety. Fast commissioning and plug & play thanks to Memosens technology and user-friendly operation.Easy upgrade of functionality – even to a complete measuring station – simply by adding modules and connecting Memosens sensors. Reduces installation effort.

Spread the love

Contact Us

    Please prove you are human by selecting the heart.