Measuring principle
Absorption
-
Application
Monitoring of WWTP in- and outlet
Drinking water and surface water -
Installation
Insitu (immersion in process), extractive with sample and flow cell
-
Characteristic
Insitu (immersion in process), extractive with sample and flow cell
UV/VIS spectrometer for measurement of TOCeq, CODeq, BODeq, SAC254, TU/TSS, Nitrate, APHA/Hazen color -
Measurement range
“Depending on the path length and the application the measuring range vary:
TOCeq: 0 to 400 mg/l
CODeq: 0 to 20 000 mg/l
BODeq: 0 to 5000 mg/l
SAC254: 0 to 1000 /m
TU: 0 to 800 FAU
TSS: 0 to 10 000 mg/l
NO3-N: 0 to 500 mg/l
APHA/Hazen: 0 to 500 Hazen -
Measuring principle
UV/VIS absorbance at 200 to 800 nm
-
Design
40 mm
-
Material
Housing: stainless steel or titanium
Windows: sapphire or quartz
O-rings: EPDM -
Dimension
571.4 x 40 mm
21.5 x 1.6 in -
Process temperature
0 to 50°C
32 to 122 °F -
Process pressure
0.5 to 10 bar abs.
7.3 to 145 psi -
Temperature sensor
-20 to 60°C
-4 to 140 °F -
Connection
G1 and NPT 3/4″
-
Ingress protection
IP68
-
Measuring principle
UV absorbance
-
Application
Monitoring of WWTP in- and outlet, drinking water and surface water
-
Characteristic
Insitu (immersion in process)
Extractive with sample and flow cell UV/VIS spectrometer for measurement of TOCeq, CODeq, BODeq, SAC254, TU/TSS, Nitrate, APHA/Hazen color -
Measurement range
Depending on the path length and the application the measuring range vary:
TOCeq: 0 to 400 mg/l
CODeq: 0 to 20 000 mg/l
BODeq: 0 to 5000 mg/l
SAC254: 0 to 1000 /m
TU: 0 to 800 FAU
TSS: 0 to 10 000 mg/l
NO3-N: 0 to 500 mg/l
APHA/Hazen: 0 to 500 Hazen -
Design
40 mm
-
Material
Housing: stainless steel or titanium
Windows: sapphire or quartz
O-rings: EPDM -
Process temperature
0 to 50°C
32 to 122 °F -
Process pressure
0.5 to 10 bar abs.
7.3 to 145 psi -
Temperature sensor
-20 to 60°C
-4 to 140 °F -
Connection
G1 and NPT 3/4″
-
Ingress protection
IP68
-
Measuring principle
UV absorbance
-
Characteristic
Insitu (immersion in process), extractive with sample and flow cell UV/VIS spectrometer for measurement of TOCeq, CODeq, BODeq, SAC254, TU/TSS, Nitrate, APHA/Hazen color
-
Measuring method
UV/VIS Absorbance at 200 to 800 nm
-
Size
571.4 x 40 mm
21.5 x 1.6 in -
Design
40 mm
-