Measuring principle
Thermocouple
-
Characteristic / Application
metric style
modular temperature assembly
universal range of application
suitable for hazardous areas
suitable for high process pressures
flanged process connection
with neck
incl. thermowell / protection tube (metal) -
Thermowell / protection tube
bar stock (drilled)
-
Insert / probe
mineral insulated (MI), flexible
-
Outer diameter protection tube / Insert
18,0 mm (0,71”)
24,0 mm (0,94”) -
Max. immersion length on request
up to 30.000,00 mm (1.181,10”)
-
Material protection tube/ thermowell
1.0460 (A105)
1.4571 (316Ti)
1.7335 (13CrMo4-5; F-11)
Alloy C276 (2.4819)
Duplex SAF 2205 (1.4462)
Titan Gr2 (3.7035 ) -
Optional coating
Not defined
-
Process connection
weld in version
flange:
DN25 PN40 B1 (EN1092)
DN40 PN40 B1(EN1092)
DN50 PN40 B1 (EN1092)
ASME 1″ 150 RF (B16.5)
ASME 1″ 300 RF (B16.5) -
Tip shape
tapered
-
Surface roughness Ra
0,8 μm (31,5 μin.)
-
Operating temperature range
Type K:
max. 1.100 °C
(max. 2.012 °F)
Type J:
max. 750 °C
(max. 1.382 °F) -
Max. process pressure (static)
at 20 °C: 400 bar (5.802 psi)
-
Accuracy
class 1 acc. to IEC 60584
-
Response time
depending on configuration
-
Integration head transmitter
yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)
-
Ex – approvals
ATEX II
NEPSI
IECEx
EAC Ex
UKCA -
Certification
Gost Metrology
SIL (transmitter only)
Field of application
The high modular and robust thermometer finds its use in several applications in almost all industries. An optional head transmitter, with all common communication protocols, makes the device ready to use with enhanced measurement accuracy and reliability compared to directly wired sensors. A variety of process connections, dimensions and materials offer flexible application possibilities.
Universal range of application
Particularly suitable for steam and gas applications with high process pressures and temperatures
Measuring range: –40 to 1100 °C (–40 to 2012 °F)
Static pressure range up to 400 bar (5800 psi)
Degree of protection: up to IP68
Benefits
High degree of flexibility thanks to modular design with standard terminal heads as per DIN EN 50446 and customer-specific immersion lengthsHigh degree of insert compatibility and design as per DIN 43772Extension neck to protect the head transmitter from overheatingFast response time with reduced/tapered tip formTypes of protection for use in hazardous locations: Intrinsic safety (Ex i.a.), non-sparking (Ex nA)Head transmitter with easy selection: Analog output 4 to 20 mA, HART®, PROFIBUS® PA or FOUNDATION Fieldbus™