Measuring principle
Potentiometric
-
Application
pH / ORP sensor for heavily contaminated media
and suspended solids
• Chemical processes
• Pulp and paper industry
• Flue gas cleaning
• Contaminated media:
• Solids
• Emulsions
• Precipitation reactions -
Characteristic
Gel pH/ORP electrode with open aperture and double gel reference with ion trap for chemical applications
-
Measurement range
ORP: –1 500 to 1 500 mV
pH: 0 to 14 -
Measuring principle
Pt-disc as additional ORP element
rH measurement and control of the reference impedance -
Design
All shaft lengths with temperature sensor
Advanced gel technology
Hydrolysis stable gel -
Material
Sensor shaft: Glass to suit process
pH membrane glass: Type B
Metal lead: Ag/AgCl
ORP measuring element: Platinum
O-ring: FKM
Process coupling: PPS fiber-glass reinforced
Nameplate: Ceramic metal oxide -
Dimension
Diameter: 12 mm (0.46 inch)
Shaft lengths: 120, 225, 360 and 425 mm
(4.68, 8.77, 14.04 and 16.57 inch) -
Process temperature
0 to 110 °C (32 to 230 °F)
-
Process pressure
0.8 to 14 bar (11.6 to 203 psi) absolute
-
Temperature sensor
NTC 30k
-
Ex certification
With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in
hazardous areas Zone 0, Zone 1 and Zone 2 -
Connection
Inductive, digital connection head with Memosens 2.0 technology
-
Ingress protection
IP68
-
Measuring principle
Sensor ORP / Redox
-
Application
pH / ORP sensor for heavily contaminated media
and suspended solids
• Chemical processes
• Pulp and paper industry
• Flue gas cleaning
• Contaminated media:
• Solids
• Emulsions
• Precipitation reactions -
Characteristic
Gel pH/ORP electrode with open aperture and double gel reference with ion trap for chemical applications
-
Measurement range
ORP: –1 500 to 1 500 mV
pH: 0 to 14 -
Measuring principle
Pt-disc as additional ORP element
rH measurement and control of the reference impedance -
Design
All shaft lengths with temperature sensor
Advanced gel technology
Hydrolysis stable gel -
Material
Sensor shaft: Glass to suit process
pH membrane glass: Type B
Metal lead: Ag/AgCl
ORP measuring element: Platinum
O-ring: FKM
Process coupling: PPS fiber-glass reinforced
Nameplate: Ceramic metal oxide -
Dimension
Diameter: 12 mm (0.46 inch)
Shaft lengths: 120, 225, 360 and 425 mm
(4.68, 8.77, 14.04 and 16.57 inch) -
Process temperature
0 to 110 °C (32 to 230 °F)
-
Process pressure
0.8 to 14 bar (11.6 to 203 psi) absolute
-
Temperature sensor
NTC 30k
-
Ex certification
With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in
hazardous areas Zone 0, Zone 1 and Zone 2 -
Connection
Inductive, digital connection head with Memosens 2.0 technology
-
Ingress protection
IP68
Field of application
Memosens CPS96E is the specialist for media with a high content of suspended solids such as dispersions, precipitations or emulsions. The open aperture prevents fouling and ensures reliable simultaneous pH and ORP measurement. Thanks to Memosens 2.0 digital technology, CPS96E combines maximum process integrity with simple operation. It resists moisture, enables lab calibration and offers extended storage of calibration and process data providing the perfect basis for predictive maintenance.
Chemical processesPulp and paper industryContaminated media and suspended solids- Dispersions- Precipitations- EmulsionsWith ATEX, IECEx, CSA C/US, NEPSI, Japan, INMETRO approvals for use in hazardous areas, zones 0, 1 and 2.
Benefits
Memosens 2.0 offers extended storage of calibration and process data, enabling better trend identification and offering a future-proof basis for predictive maintenance and enhanced IIoT services. Simultaneous measurement of pH, ORP and rH values (in rH mode) provides a better process overview for tighter process control. Additional platinum electrode enables constant monitoring of reference impedance and thus sensor quality. Rob