Measuring principle
Total chlorine
-
Application
Wastewater treatment plants, disinfection monitoring and control in wastewater outlet, reuse of outlet water, utilities of all industries, all disinfection measuring points
-
Characteristic
Amperometric measurement of total chlorine
-
Measurement range
0.1 to 10 mg/l total chlorine
-
Measuring method
Total chlorine consists of free chlorine (HOCl, OCl-) and bound chlorine (chloramines)
All components are reduced at the working electrode
Almost pH independent -
Design
Closed, membrane covered measuring cell
-
Material
Sensor shaft: PVC
Membrane: PET
Membrane cap: PPE -
Dimension
Diameter: 25 mm (0.98 inch)
Length: 148 mm (5.83 inch) -
Process temperature
5 to 45 °C
(41 to 113 °F) -
Process pressure
1 to 4 bar (14.5 to 58 psi)
-
Temperature sensor
10k NTC integrated (Memosens)
-
Connection
Inductive, digital connection head with Memosens
Field of application
Total chlorine is the mixture of free and bound chlorine components and thus a good indicator of residual disinfectants in discharge water. The sensor Memosens CCS120D provides reliable monitoring of total chlorine limits in the outlet. It reduces your maintenance effort due to easy replacement of electrolyte and membrane cap. Memosens CCS120D is not the latest Memosens generation. More information on the new sensor Memosens CCS53E.
Memosens CCS120D measures total chlorine in:Wastewater treatment plants- Disinfection monitoring and control in wastewater outlet- Reuse of outlet waterUtilities of all industries- All disinfection measuring points
Benefits
Easy handling thanks to Memosens digital technology Enables predictive maintenance due to storage of sensor and process-specific data Low maintenance thanks to fast exchange of membrane cap and electrolyte Flexible installation in CYA112 immersion assembly or CCA250 flow assembly Easy combination with other relevant parameters of liquid analysis thanks to connection to the Liquiline multichannel transmitter