Ammonium analyzer Liquiline System CA80AM
Colorimetric system for online monitoring of water and wastewater treatment
-
Measuring principle
Colorimetric
-
Characteristic
Analyzer for ammonium in aqueous solutions
-
Measuring method
Comply with standard colorimetric measuring principle – indophenol blue method following ISO 7150-1, DIN 38406-5 and GB 7481-87
-
Size
Housing (open version):
793 x 530 x 417 mm
31.22 x 20.87 x 16.42 in
Housing (closed version):
793 x 530 x 463 mm
31.22 x 20.87 x 18.23 in
Housing with base:
1723 x 530 x 463 mm
67.83 x 20.87 x 18.23 in -
Design
Open design / cabinet:
Plastic ASA-PC
Stand housing:
Plastic ASA-PC, steel coated -
Process temperature
4 to 40 °C (39 to 104 °F)
-
Ambient temperature
5 to 40 °C (41 to 104 °F)
Outdoor version: – 20 to 40 °C (-4 to 104 °F) -
Process pressure
Unpressurized
-
Sample flow rate
Min. 5 ml/min (0.17 fl.oz/min)
-
Consistency of the sample
Low solids content
Turbitity < 50 NTU, aqueous, homogenized -
Specials
Easy upgrade to measuring station with up to four digital Memosens sensors
Automatic calibration and cleaning
User-configurable measuring, cleaning and calibration intervals
Optional cooling module for longer reagent life time
Automatic calibration and cleaning
Optional 2 channel version
User-definable measuring ranges
Flexibale functionality and modular expandable
Digital communication for remote access -
Application
Monitoring and optimization of the cleaning capacity of municipal and industrial wastewater treatment plants
Monitoring of the wastewater treatment plant outlet for documentation purposes
Monitoring and optimization of activated sludge basins -
Power supply
100 to 120 VAC / 200 to 240 VAC ± 10%
24 VDC ± 10%
50 ± 1 or 60 ± 1.2 Hz -
Output / communication
2x 0/4 to 20 mA
Optional: Webserver, Modbus, Ethernet/IP, Profibus DP -
Input
1 or 2 measuring channels
Optinonal 1 to 4 digital sensor inputs for sensors with Memosens protocol -
Measurement range
0.05 to 20 mg/l NH4-N
0.5 to 50 mg/l NH4-N
1 to 100 mg/l NH4-N
0.5 to 50 mg/l with dilution function to maximum 10 to 1000 mg/l NH4-N -
Consumables
Reagents and standard solutions CY80AM
Cleaner CY800
Maintenance kit CAV800
Field of application
Liquiline System CA80AM offers high-precision ammonium measurement in all critical control points. As member of the Liquiline platform, it enables plug & play and easy upgrade to a measuring station – minimizing the installation effort. Automatic calibration and cleaning as well as the low consumption of reagents save you operating and maintenance costs. Advanced diagnostics with remote access ensure process safety and support you in providing process documentation to the authorities.
Liquiline System CA80AM monitors and controls:
Inlets of wastewater treatment plants
Outlets of wastewater treatment plants for documentation purposes
Aeration basins
Water treatment
Cleaning capacity of municipal and industrial wastewater plants
Benefits
Fast commissioning and plug & play thanks to Memosens technology and user-friendly operation Easy upgrade of functionality – even to a complete measuring station – simply by adding modules and connecting Memosens sensorsAdvanced diagnostics with remote access for increased process safetyReduced operating costs through automatic, configurable calibration and cleaning and long reagent shelf life with optional cooling moduleCompliance with standard colorimetric measuring principle – indophenol blue method – following ISO 7150-1, DIN 38406-5 and GB 7481-87
Proline Prosonic Flow G 300 ultrasonic flowmeter
Highly robust gas specialist for fluctuating conditions with compact, easily accessible transmitter
-
Measuring principle
Ultrasonic flow
-
Product headline
Highly robust gas specialist for fluctuating process conditions with compact, easily accessible transmitter.
Flexible device with user-definable gas mixtures for demanding measuring tasks.
Accurate measurement of natural and process gas in the chemical as well as oil and gas industries. -
Sensor features
Maximum reliability even with humid or wet gas – sensor design insensitive to condensate. High-performance process control – real-time pressure- and temperature-compensated values. Efficient solution – multivariable, no pressure loss.
Direct measurement: flow, pressure & temperature. Wetted parts: titanium / 316L. Maximum measuring accuracy: 0.5 %. -
Transmitter features
Full access to process and diagnostic information – numerous, freely combinable I/Os. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology.
Compact dual-compartment housing with up to 3 I/Os. Backlit display with touch control and WLAN access. Remote display available. -
Nominal diameter range
DN 25 to 300 (1 to 12″)
-
Wetted materials
Measuring tube: 1.4408/1.4409 (CF3M)
Transducer: 1.4404 (316, 316L), Titan Grade 2 -
Measured variables
Volume flow, corrected volume flow, mass flow, flow velocity, speed of sound, pressure, temperature, density, dynamic viscosity, energy flow, Wobbe index, methane fraction, calorific value, molar mass
-
Max. measurement error
Volume flow (standard):
‐ ±1.0 % o.r. for 3 to 40 m/s (9.84 to 131.23 ft/s)
‐ ±2 % o.r. for 0.3 to 3 m/s (0.98 to 9.84 ft/s)
Volume flow (optional calibration):
‐ ±0.5 % o.r. for 3 to 40 m/s (9.84 to 131.23 ft/s)
‐ ±1.0 % o.r. for 0.3 to 3 m/s (0.98 to 9.84 ft/s)
Corrected volume flow (standard):
‐ ±1.5 % o.r. for 3 to 40 m/s (9.84 to 131.23 ft/s)
‐ ±2.5 % o.r. for 0.3 to 3 m/s (0.98 to 9.84 ft/s)
Corrected volume flow (optional calibration):
‐ ±1.0 % o.r. for 3 to 40 m/s (9.84 to 131.23 ft/s)
‐ ±1.5 % o.r. for 0.3 to 3 m/s (0.98 to 9.84 ft/s)
Sound Velocity: ±0.2 % o.r. -
Measuring range
Gas: 0.3 m/s to 40 m/s
-
Max. process pressure
0.7 to 101 bar a (10.15 to 1464.88 psi a)
-
Medium temperature range
-50 to 150 °C (-58 to +302°F)
-50 to 100 °C (-58 to +212°F) with integrated pressure cell -
Ambient temperature range
-40 to 60 °C(-40 to +140 °F)
Optional: -50 to 60 °C(-58 to +140 °F) -
Sensor housing material
Stainless Steel, 1.4404(316/316L), 1.4408/1.4409 (CF3M)
-
Transmitter housing material
AlSi10Mg, coated; 1.4409 (CF3M) similar to 316L
Polycarbonate -
Degree of protection
Compact version: IP66/67, type 4X enclosure.
Optional: External WLAN antenna: IP67 -
Display/Operation
4-line backlit display with Touch Control (operation from outside)
Configuration via local display and operating tools possible
Remote display available -
Outputs
3 outputs:
4-20 mA HART (active/passive)
4-20 mA (active/passive)
Pulse/frequency/switch output (active/passive)
Double pulse output (active/passive)
Relay output -
Inputs
Status input
4-20 mA input -
Digital communication
HART, Modbus RS485
-
Power supply
24V DC
100 to 230 V AC
AC 100 to 230 V / DC 24 V (non harzardous area) -
Hazardous area approvals
ATEX, IECEx, cCSAus, JPN, EAC, UK Ex, KC
-
Product safety
CE, C-tick
-
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a -
Pressure approvals and certificates
PED, CRN
-
Material certificates
3.1 material
NACE MR0175/MR0103
Field of application
For a wide range of gas applications Prosonic Flow G provides reliable flow measurement, even with wet gas and changing gas properties and compositions. A pressure-rated sensor housing with rupture disc limits safety risks. The compact transmitter offers high flexibility in terms of operation and system integration: access from one side, remote display and improved connectivity options. Heartbeat Technology enables compliance and process safety at all times.
The measuring principle is unaffected by gas compositionAccurate measurement of natural and process gas in the chemical as well as oil and gas industries.Device properties:Direct measurement: flow, pressure & temperatureWetted parts: titanium / 316LMaximum measuring accuracy: 0.5 %Compact dual-compartment housing with up to 3 I/OsBacklit display with touch control and WLAN accessRemote display available
Benefits
Flexible device with user-definable gas mixtures for demanding measuring tasks
Maximum reliability even with humid or wet gas – sensor design insensitive to condensateHigh-performance process control – real-time pressure- and temperature-compensated values
Efficient solution – multivariable, no pressure loss
Full access to process and diagnostic information – numerous, freely combinable I/Os
Reduced complexity and variety – freely configurable I/O functionality
Integrated verification – Heartbeat Technology
Proline Promass X 300 Coriolis flowmeter
Highest capacity four-tube flowmeter with a compact, easily accessible transmitter
-
Measuring principle
Coriolis
-
Product headline
Highest capacity four-tube flowmeter with a compact, easily accessible transmitter.
For highest flow rates and outstanding performance in on/offshore oil and gas applications. -
Sensor features
Increased profit – single installation point providing premium accuracy for large quantities. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs.
Nominal diameter: DN 300 to 400 (12 to 16″). Four-tube system with low pressure drop. Complete exterior design made of 1.4435 (316L). -
Transmitter features
Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology.
Compact dual-compartment housing with up to 3 I/Os. -
Nominal diameter range
DN 300 to 400 (12 to 16″)
-
Wetted materials
Measuring tube: 1.4404 (316/316L)
Connection: 1.4404 (316/316L) -
Measured variables
Mass flow, density, temperature, volume flow, corrected volume flow (API tables), reference density, concentration
-
Max. measurement error
Mass flow (liquid): ±0.10 % (standard), 0.05 % (option)
Volume flow (liquid): ±0.10 %
Mass flow (gas): ±0.35 %
Density (liquid): ±0.0005 g/cm³ -
Measuring range
0 to 4100 t/h (0 to 4520 tn. sh./h)
-
Max. process pressure
PN 100, Class 600
-
Medium temperature range
–50 to +180 °C (–58 to +356 °F)
-
Ambient temperature range
Standard: –40 to +60 °C (–40 to +140 °F)
Option: –50 to +60 °C (–58 to +140 °F) -
Sensor housing material
1.4404 (316L), highest corrosion resistance
-
Transmitter housing material
AlSi10Mg, coated; 1.4409 (CF3M) similar to 316L
-
Degree of protection
IP66/67, type 4X enclosure
-
Display/Operation
4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available” -
Outputs
3 outputs:
4-20 mA HART (active/passive)
4-20 mA WirelessHART
4-20 mA (active/passive)
Pulse/frequency/switch output (active/passive)
Double pulse output (active/passive)
Relay output -
Inputs
Status input
4-20 mA input -
Digital communication
HART, PROFIBUS DP, PROFIBUS PA, FOUNDATION Fieldbus, Modbus RS485, PROFINET, PROFINET over Ethernet-APL, Ethernet/IP, OPC-UA
-
Power supply
DC 24 V
AC 100 to 230 V
AC 100 to 230 V / DC 24 V (non-hazardous area) -
Hazardous area approvals
ATEX, IECEx, cCSAus, NEPSI, INMETRO, EAC, UK Ex, KC
-
Product safety
CE, C-tick, EAC marking
-
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)
MI-005 Liquids other than water (Hydrocarbons), LPG, cryogenic
MI-002, PTB -
Marine approvals and certificates
LR approval, DNV approval, ABS approval, BV approval, CCS approval
-
Pressure approvals and certificates
PED, CRN
-
Material certificates
3.1 material
NACE MR0175/MR0103, PMI; welding test acc. to EN ISO, ASME, NORSOK
-
Measuring principle
Coriolis
-
Product headline
Proline Promass P 100 Coriolis flowmeter
The specialist for life sciences with an ultra-compact transmitter
-
Measuring principle
Coriolis
-
Product headline
Specialist for life sciences with an ultra-compact transmitter.
Dedicated to applications under sterile conditions in the life sciences industry. -
Sensor features
Highest process quality – fully compliant to industry requirements. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs.
ASME BPE, 3‐A and EHEDG conform & low delta ferrite. Electropolished measuring tube in 1.4435 (316L). Fast recovery from CIP/SIP. -
Transmitter features
Space‐saving transmitter – full functionality on the smallest footprint. Time‐saving local operation without additional software and hardware – integrated web server. Integrated verification – Heartbeat Technology.
Robust, ultra-compact transmitter housing. Highest degree of protection: IP69. Local display available. -
Nominal diameter range
DN 8 to 50 (⅜ to 2″)
-
Wetted materials
Measuring tube: 1.4435 (316L)
Connection: 1.4435 (316L); 1.4404 (316/316L) -
Measured variables
Mass flow, density, temperature, volume flow, corrected volume flow, reference density, concentration
-
Max. measurement error
Mass flow (liquid): ±0.1 %
Volume flow (liquid): ±0.1 %
Density (liquid): ±0.0005 g/cm³ -
Measuring range
0 to 70 000 kg/h (0 to 2570 lb/min)
-
Max. process pressure
PN 63, Class 300, 40K
-
Medium temperature range
Standard: –50 to +150 °C (–58 to +302 °F)
Option: –50 to +205 °C (–58 to +401 °F) -
Ambient temperature range
Standard: –40 to +60 °C (–40 to +140 °F)
Option: –50 to +60 °C (–58 to +140 °F) -
Sensor housing material
1.4301 (304), corrosion resistant
-
Transmitter housing material
Compact: AlSi10Mg, coated
Compact/ultra‐compact: 1.4301 (304) -
Degree of protection
Standard: IP66/67, type 4X enclosure
Option: IP69 -
Display/Operation
4‐line backlit display available (no local operation)
Configuration via web browser and operating tools possible -
Outputs
4‐20 mA HART (active)
Pulse/frequency/switch output (passive) -
Inputs
None
-
Digital communication
HART, Modbus RS485, EtherNet/IP, PROFIBUS DP, PROFINET
-
Power supply
DC 20 to 30 V
-
Hazardous area approvals
ATEX, IECEx, cCSAus, INMETRO, NEPSI, EAC
-
Product safety
CE, C-Tick, EAC marking
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Verification: Heartbeat Technology complies with requirements for traceable verification according to ISO 9001:2008, chapter 7.6. a (TUV attestation) -
Pressure approvals and certificates
PED, CRN
-
Material certificates
3.1 material
-
Hygienic approvals and certificates
3-A, EHEDG, ASME BPE, ISPE, cGMP
-
Measuring principle
Coriolis
-
Hygienic approvals and certificates
cGMP
-
Measuring principle
Coriolis
-
Product headline
Specialist for life sciences with an ultra-compact transmitter.
Dedicated to applications under sterile conditions in the life sciences industry. -
Sensor features
Highest process quality – fully compliant to industry requirements. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs.
ASME BPE, 3‐A and EHEDG conform & low delta ferrite. Electropolished measuring tube in 1.4435 (316L). Fast recovery from CIP/SIP. -
Transmitter features
Space‐saving transmitter – full functionality on the smallest footprint. Time‐saving local operation without additional software and hardware – integrated web server. I
Proline Prowirl F 200 vortex flowmeter
Versatile flowmeter with detection of wet steam conditions and best-in-class accuracy
-
Measuring principle
Vortex
-
Product headline
Versatile flowmeter with detection of wet steam conditions and best-in-class accuracy.
Easy energy management – integrated temperature and pressure measurement for steam and gases.
Suitable for a wide range of applications; optimized for steam applications. -
Sensor features
Space-saving engineering – inlet run compensation. Same accuracy down to Re 10 000 – most linear Vortex meter body. Long-term stability – robust drift-free capacitive sensor.
Wet steam capabilities for DN 25 to 300 (1 to 12″). Flexible positioning of pressure cell. Industrial siphon design for pressure measurement. -
Transmitter features
Convenient device wiring – separate connection compartment. Safe operation – no need to open the device due to display with touch control, background lighting. Integrated verification – Heartbeat Technology.
Display module with data transfer function. Robust dual-compartment housing. Plant safety: worldwide approvals (SIL, Haz. area). -
Nominal diameter range
DN 15 to 300 (½ to 12″)
-
Wetted materials
Measuring tube: 1.4408 (CF3M); CX2MW similar to Alloy C22, 2.4602
DSC sensor: 1.4404 (316/316L); UNS N06022 similar to Alloy C22, 2.4602
Process connection: 1.4404/F316/F316L); 2.4602 -
Measured variables
Volume flow, mass flow, corrected volume flow, energy flow, heat flow difference, temperature
-
Max. measurement error
Volume flow (liquid): ±0.75 %
Volume flow (optional): ±0.65 %
Volume flow (steam, gas): ±1.00 %
Mass flow (saturated steam): ±1.7% (temperature compensated); ±1.5% (temperature/pressure compensated)
Mass flow (superheated steam, gas): ±1.5 (temperature/pressure compensated); ±1.7% (temperature compensated + external pressure compensation)
Mass flow (liquid): ±0.85% -
Measuring range
Liquid: 0.076 to 2100 m³/h (0.045 to 1300 ft³/min)
depending on medium: water with 1 bar a, 20 °C (14.5 psi a, 68° F)
Steam, gas: 0.39 to 28000 m³/h (0.23 to 17000 ft³/min)
depending on medium: steam with 180 °C, 10 bar a (356 °F, 145 psi a); air with 25 °C, 4.4 bar a (77 °F, 63.8 psi a) -
Max. process pressure
PN 100, Class 600, 20K
-
Medium temperature range
Standard: –40 to +260 °C (–40 to +500 °F)
High/low temperature (option): –200 to +400 °C (–328 to +752 °F) -
Ambient temperature range
Compact version (standard): –40 to +80 °C (–40 to +176 °F)
Compact version (option): –50 to +80 °C (–58 to +176 °F)
Remote version (standard): –40 to +85 °C (–40 to +185 °F)
Remote version (option): –50 to +85 °C (–58 to +185 °F) -
Sensor housing material
Sensor connection housing: AlSi10Mg, coated; 1.4408 (CF3M)
-
Transmitter housing material
AlSi10Mg, coated; 1.4404 (316L)
-
Degree of protection
Compact version: IP66/67, type 4X enclosure
Sensor remote version: IP66/67, type 4X enclosure
Transmitter remote version: IP66/67, type 4X enclosure -
Display/Operation
4‐line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available -
Outputs
4‐20 mA HART (passive)
4‐20 mA (passive)
Pulse/frequency/switch output (passive) -
Inputs
Current Input 4‐20 mA (passive)
-
Digital communication
HART, PROFIBUS PA, FOUNDATION Fieldbus
-
Power supply
DC 12 to 35 V (4‐20 mA HART with/without pulse/frequency/switch output)
DC 12 to 30 V (4‐20 mA HART, 4‐20 mA)
DC 12 to 35 V (4‐20 mA HART, pulse/frequency/switch output, 4‐20 mA input)
DC 9 to 32 V (PROFIBUS PA, pulse/frequency/switch output) -
Hazardous area approvals
ATEX, IECEx, cCSAus, JPN, EAC
-
Product safety
CE, C-TICK, EAC, UK Ex
-
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation) -
Marine approvals and certificates
Proline t-mass F 300 thermal mass flowmeter
Inline flowmeter with long-term stability and a compact, easily accessible transmitter
-
Measuring principle
Thermal
-
Product headline
Inline flowmeter with long-term stability and a compact, easily accessible transmitter.
Flexible, convenient programming based on 21 standard gases or freely definable gas mixtures thereof.
Measurement of utility and process gases as well as gas mixtures in small line sizes. -
Sensor features
High level of process control – premium measurement accuracy and repeatability. Reliable monitoring – detection of process disturbances and reverse flow. Easy maintenance – removable sensor.
Inline version with DN 15 to 100 (½ to 4″). Bidirectional measurement; high measuring performance. Patented drift-free sensor with SIL 2. -
Transmitter features
Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology.
Compact dual-compartment housing with up to 3 I/Os. Backlit display with touch control and WLAN access. Remote display available. -
Nominal diameter range
DN 15 to DN 100 (1/2″ to 4″)
-
Wetted materials
Measuring tubes
DN 15 to 50 (½ to 2″): stainless cast steel, CF3M/1.4408
DN 65 to 100 (2½ to 4″): stainless steel, 1.4404 (316/316L)
Process connections
Flange connections
Stainless steel, 1.4404 (F316/F316L)
Threaded connections
Stainless steel, 1.4404 (316/316L)
Sensing element
Unidirectional
Stainless steel, 1.4404 (316/316L)
Alloy C22, 2.4602 (UNS N06022);
Bidirectional
Stainless steel, 1.4404 (316/316L)
Reverse flow detection
Stainless steel, 1.4404 (316/316L) -
Measured variables
Massflow, temperature, standard volume flow, volume flow, Free air delivery, velocity, heat flow, energy flow, density
-
Max. measurement error
Gas: 1.0% o.r. (10 to 100% o.f.s.), 0.1% o.f.s. (1 to 10% o.f.s.)
-
Measuring range
0.5 to 3750 kg/h (1.1 to 8250 lb/h)
-
Max. process pressure
PN40 / Cl. 300 / 20K
-
Medium temperature range
-40 °C to +180°C (-40 °F to +356 °F)
-
Ambient temperature range
-40 to 60°C (-40 to 140°F)
Optional:
Transmitter: -50 to 60°C (-50 to 140°F),
Sensor: -60 to 60°C (-60 to 140°F) -
Transmitter housing material
Aluminium, AlSi10Mg, coated
Polycarbonate -
Degree of protection
IP66/67, Type 4X enclosure
-
Display/Operation
4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available -
Outputs
3 outputs:
4-20 mA HART (active/passive)
4-20 mA (active/passive)
Pulse/frequency/switch output (active/passive)
Relay output -
Inputs
Status input
4-20 mA input -
Digital communication
HART, Modbus RS485
-
Power supply
DC 24V
AC 100 to 240V -
Hazardous area approvals
ATEX, cCSAus, IECEx, NEPSI, JPN, UK Ex, EAC, KC
-
Product safety
CE, C-tick
-
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a -
Pressure approvals and certificates
PED, CRN
-
Material certificates
3.1 material
NACE MR0175/MR0103
Field of application
The patented sensor design of t-mass F provides unprecedented measurement stability in thermal mass flow measurement. It compensates in real time for changes of process conditions: temperature, pressure, flow direction and gas type. Its compact transmitter offers high flexibility in terms of operation and system integration: access from one side, remote display and improved connectivity options. Heartbeat Technology enables measurement reliability and compliant verification.
Measuring principle is characterized by a high operable flow range and direct mass flow measurementMeasur
Proline Promag W 400 electromagnetic flowmeter
Versatile standard flowmeter for the water and wastewater industry
-
Measuring principle
Electromagnetic
-
Product headline
Versatile standard flowmeter for the water and wastewater industry.
Reliable measurement at constant accuracy with 0 x DN inlet run and no pressure loss.
Ideal for water measurement, e.g. drinking water, utility water and industrial/municipal wastewater. -
Sensor features
Flexible engineering – sensor with fixed or lap-joint process connections. Application fitness – EN ISO 12944 corrosion protection for underground or underwater installation. Improved plant availability – sensor compliant with industry-specific requirements.
International drinking water approvals. Degree of protection IP68 (Type 6P enclosure). Approved for custody transfer to MI-001/OIML R49. -
Transmitter features
Safe operation – no need to open the device due to display with touch control, background lighting. Time‐saving local operation without additional software and hardware – integrated web server. Integrated verification – Heartbeat Technology.
Transmitter housing made of durable polycarbonate or aluminium. WLAN access. Integrated data logger: measured values monitoring. -
Nominal diameter range
DN 25 to 3000(1 to 120″)
-
Wetted materials
Liner material hard rubber: 0 to +80 °C (+32 to +176 °F)
Liner material polyurethane: –20 to +50 °C (–4 to +122 °F)
Liner material PTFE: –20 to +90 °C (–4 to +194 °F)
Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum -
Measured variables
Volume flow, conductivity, mass flow
-
Max. measurement error
Volume flow (standard): ±0.5 % o.r. ± 1 mm/s (0.04 in/s)
Volume flow (option): ±0.2 % o.r. ± 2 mm/s (0.08 in/s), Flat Spec -
Measuring range
0.5 m3/h to 263000 m3/h (2.5gal/min to 1665 Mgal/d)
-
Max. process pressure
PN 40, Class 300, 20K
-
Medium temperature range
Liner material hard rubber: 0 to +80 °C (+32 to +176 °F)
Liner material polyurethane: –20 to +50 °C (–4 to +122 °F)
Liner material PTFE: –20 to +90 °C (–4 to +194 °F) -
Ambient temperature range
Liner material hard rubber: 0 to +80 °C (+32 to +176 °F)
Liner material polyurethane: –20 to +50 °C (–4 to +122 °F) -
Sensor housing material
DN 25 to 300 (1 to 12″): AlSi10Mg, coated
DN 25 to 2000 (1 to 78″): Carbon steel with protective varnish
Sensor connection housing (standard): AlSi10Mg, coated
Sensor connection housing (option): Polycarbonate -
Transmitter housing material
Polycarbonat; AlSi10Mg, coated
-
Degree of protection
Compact version: IP66/67, type 4X enclosure
Sensor remote version (standard): IP66/67, type 4X enclosure
Sensor remote version (option): IP68, type 6P enclosure, with protective varnish
according to EN ISO 12944 C5‐M/Im1/Im2/Im3
Transmitter remote version: IP66/67, Type 4X enclosure -
Display/Operation
4‐line backlit display with touch control (operation from outside)
Configuration via local display, web browser and operating tools possible -
Outputs
3 ouputs:
0‐20 mA/4‐20 mA HART (active)
Pulse/frequency/switch output (passive)
Pulse/frequency output (passive)
Switch output (passive) -
Inputs
Status input
-
Digital communication
HART, PROFIBUS DP, EtherNet/IP, Modbus RS485
-
Power supply
AC 100…240 V / AC/DC 24 V
-
Hazardous area approvals
cCSAus
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR
Custody transfer according to MI‐001 or OIML R49 (optional class I in combination with 0DN inlet run) -
Pressure approvals and certificates
CRN
-
Material certificates
3.1 material
-
Hygienic approvals and certificates
Drinking water approval: ACS, KTW/W270, NSF 61, WRAS BS 6920
Field of application
With its international approvals for custody transfer as well as drinking water, Promag W serves the broadest variety of applications. It is available as both compact or remote version. Promag W 400 saves time and costs thanks to the broad functionality of its transmitter optimized for Water & Wastewater. In addition, Heartbeat Technology enables measurement reliability and compliant verification.
The bidirectional measuring principle is virtually independent of pressure, density, temperature and viscosityIdeal for water measurement , e.g. drinking water, utility water and industrial/municipal wastewaterDevice propertiesInternational drinking water approvalsDegree of protection IP68 (Type 6P enclosure)Approved for custody transfer to MI-001/OIML R49Transmitter housing made of durable polycarbonate or aluminiumWLAN accessIntegrated data logger: measured values monitoring
Benefits
Reliable measurement at constant accuracy with 0 x DN inlet run and no pressure lossFlexible engineering – sensor with fixed or lap-joint process connectionsApplication fitness – EN ISO 12944 corrosion protection for underground or underwater installationImproved plant availability – sensor compliant with industry-specific requirementsSafe operation – no need to open the deviceTime-saving local operation without additional software and hardware – integrated web serverIntegrated verification and build-up detection – Heartbeat Technology
Proline Promag W 300 electromagnetic flowmeter
Specialist for demanding water and wastewater applications with a compact, easily accessible transmitter
-
Measuring principle
Electromagnetic
-
Product headline
Specialist for demanding water and wastewater applications with a compact, easily accessible transmitter.
Reliable measurement at constant accuracy with 0 x DN inlet run and no pressure loss.
Dedicated to the measurement of industrial or municipal water and wastewater . -
Sensor features
Flexible engineering – sensor with fixed or lap-joint process connections. Long-term operation – robust and completely welded sensor. Improved plant availability – sensor compliant with industry-specific requirements.
International drinking water approvals. -
Transmitter features
Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology.
Compact dual-compartment housing with up to 3 I/Os. Backlit display with touch control and WLAN access. Remote display available. -
Nominal diameter range
DN 25 to 3000(1 to 120″)
-
Wetted materials
Liner: Polyurethane; Hard rubber, PTFE
Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum -
Measured variables
Volume flow, conductivity, mass flow
-
Max. measurement error
Volume flow (standard): ±0.5 % o.r. ± 1 mm/s (0.04 in/s)
Volume flow (option): ±0.2 % o.r. ± 2 mm/s (0.08 in/s), Flat Spec -
Measuring range
0.5 m3/h to 263000 m3/h (2.5gal/min to 1665 Mgal/d)
-
Max. process pressure
PN 40, Class 300, 20K
-
Medium temperature range
Liner material hard rubber: 0 to +80 °C (+32 to +176 °F)
Liner material polyurethane: –20 to +50 °C (–4 to +122 °F)
Liner material PTFE: –20 to +90 °C (–4 to +194 °F) -
Ambient temperature range
Flange material carbon steel: –10 to +60 °C (+14 to +140 °F)
Flange material stainless steel: –40 to +60 °C (–40 to +140 °F) -
Sensor housing material
DN 25 to 300 (1 to 12″): AlSi10Mg, coated
DN 25 to 2400 (1 to 90″): Carbon steel with protective varnish
Sensor connection housing (standard): AlSi10Mg, coated
Sensor connection housing (option): Polycarbonate; 1.4409 (CF3M) similar to 316L -
Transmitter housing material
AlSi10Mg, coated; 1.4409 (CF3M) similar to 316L; stainless steel for hygenic transmitter design
-
Display/Operation
4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available -
Outputs
3 outputs:
4-20 mA HART (active/passive)
4-20 mA WirelessHART
4-20 mA (active/passive)
Pulse/frequency/switch output (active/passive)
Double pulse output (active/passive)
Relay output -
Inputs
Status input 4-20 mA input
-
Digital communication
HART, PROFIBUS DP, PROFIBUS PA, FOUNDATION Fieldbus, Modbus RS485, PROFINET, PROFINET over Ethernet-APL, Ethernet/IP, OPC-UA
-
Power supply
DC 24 V
AC 100 to 230 V
AC 100 to 230 V / DC 24 V (non-hazardous area) -
Hazardous area approvals
ATEX, IECEx, cCSAus, INMETRO, NEPSI, EAC; JPN, UK Ex, KC
-
Product safety
CE, C-tick, EAC marking
-
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation) -
Marine approvals and certificates
LR approval, DNV approval, ABS approval, BV approval
-
Pressure approvals and certificates
CRN, PED
-
Material certificates
3.1 material
-
Hygienic approvals and certificates
ACS, KTW/W270, NSF 61, WRAS BS 6920
Field of application
The premium device for water and wastewater measurement Promag W 300 was designed for reliable use in hazardous areas and under harsh conditions. Its compact transmitter offers high flexibility in terms of operation and system integration: access from one side, remote display and improved connectivity options. Heartbeat Technology enables measurement reliability and compliant verification.
The bidirectional measuring principle is virtually independent of pressure, density, temperature and viscosityDedicated to the measurement of industrial or municipal water and wastewaterDevice propertiesInternational drinking water approvalsDegree of protection IP68 (Type 6P enclosure)Compact dual-compartment housing with up to 3 I/OsBacklit display with touch control and WLAN accessRemote display available
Benefits
Reliable measurement at constant accuracy with 0 x DN inlet run and no pressure lossFlexible engineering – sensor with fixed or lap-joint process connectionsLong-term operation – robust and completely welded sensorImproved plant availability – sensor compliant with industry-specific requirementsFull access to process and diagnostic information – numerous, freely combinable I/Os and EthernetReduced complexity and variety – freely configurable I/O functionalityIntegrated verification – Heartbeat Technology
Proline Promag H 100 electromagnetic flowmeter
The specialist for hygienic applications with an ultra-compact transmitter
-
Measuring principle
Electromagnetic
-
Product headline
Specialist for hygienic applications with an ultra-compact transmitter.
Multivariable measurement of flow, temperature and conductivity.
Dedicated to demanding applications in the food and beverage as well as in life sciences industries. -
Sensor features
Flexible installation concept – numerous hygienic process connections. Energy‐saving flow measurement – no pressure loss due to cross section constriction. Maintenance‐free – no moving parts.
Integrated temperature measurement. Sensor housing made of stainless steel (3-A, EHEDG). Wetted materials CIP, SIP cleanable. -
Transmitter features
Space‐saving transmitter – full functionality on the smallest footprint. Time‐saving local operation without additional software and hardware – integrated web server. Integrated verification – Heartbeat Technology.
Robust, ultra-compact transmitter housing. High degree of protection: IP69. Local display available. -
Nominal diameter range
DN 2 to 150 (¹⁄₁₂ to 6″)
-
Wetted materials
Liner: PFA
Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum; Platinum
Process Connections: stainless steel, 1.4404 (F316L); PVDF; PVC adhesive sleeve
Seals: O-ring seal (EPDM, FKM, Kalrez), aseptic molded seal (EPDM, FKM, silicone)
Grounding Rings: stainless steel, 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); tantalum -
Measured variables
Volume flow, temperature, conductivity, mass flow, corrected volume flow, corrected
conductivity -
Max. measurement error
Volume flow (standard): ±0.5 % o.r. ± 1 mm/s (0.04 in/s)
Volume flow (option) ±0.2 % o.r. ± 2 mm/s (0.08 in/s) -
Measuring range
0.06 dm³/min to 600 m³/h (0.015 to 2650 gal/min)
-
Max. process pressure
PN 40, Class 150, 20K
-
Medium temperature range
–20 to +150 °C (–4 to +302 °F)
-
Ambient temperature range
–40 to +60 °C (–40 to +140 °F)
-
Sensor housing material
1.4301 (304), corrosion resistant
-
Transmitter housing material
Compact: AlSi10Mg, coated
Compact/ultra‐compact: 1.4301 (304) -
Degree of protection
Standard: IP66/67, type 4X enclosure
Option: IP69 -
Display/Operation
4‐line backlit display available (no local operation)
Configuration via web browser and operating tools possible -
Outputs
4‐20 mA HART (active)
Pulse/frequency/switch output (passive) -
Inputs
None
-
Digital communication
HART, PROFIBUS DP, Modbus RS485, EtherNet/IP, PROFINET
-
Power supply
DC 20 to 30 V
-
Hazardous area approvals
ATEX, IECEx, cCSAus, INMETRO, EAC
-
Product safety
CE, C-tick
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Verification: Heartbeat Technology complies with requirements for traceable verification according to ISO 9001:2008, chapter 7.6. a (TUV attestation) -
Marine approvals and certificates
LR approval, DNV approval, ABS approval, BV approval
-
Pressure approvals and certificates
PED, CRN
-
Material certificates
3.1 material
-
Hygienic approvals and certificates
Sanitary approval: EHEDG, 3-A, liner and seals acc. to FDA, cGMP
Field of application
Promag H is the preferred sensor for hygienic applications with highest requirements in the food and beverage and life sciences industries. The ultra-compact transmitter provides full performance on the smallest footprint and enables seamless system integration, making Promag H 100 the preferred choice for skid builders, equipment manufacturers and system integrators. Heartbeat Technology enables compliance and process safety at all times.
The bidirectional measuring principle is virtually independent of pressure, density, temperature and viscosityDedicated to demanding applications in the food and beverage as well as in life sciences industriesDevice propertiesIntegrated temperature measurementSensor housing made of stainless steel (3-A, EHEDG)Wetted materials CIP, SIP cleanableRobust, ultra-compact transmitter housingHigh ingress protection: IP69Local display available
Benefits
Multivariable measurement for flow, temperature and conductivityFlexible installation concept – numerous hygienic process connectionsEnergy-saving flow measurement – no pressure loss due to cross-section constrictionSpace-saving transmitter – full functionality on the smallest footprintTime-saving local operation without additional software and hardware – integrated web serverIntegrated verification – Heartbeat TechnologyMaintenance-free – no m
iTEMP TMT82 HART® 7 temperature transmitter
Dual channel, SIL compliant temperature transmitter for use in hazardous areas
-
Measuring principle
Head transmitter
-
Input
2 x RTD, TC, Ohm, mV
-
Output
1 x analog 4…20 mA / HART
-
Auxiliary power supply
11…42 V DC (head transmitter)
12…42 V DC (Din rail device)
11/12…30 V DC (Ex ia) -
Communication
HART-protocol
-
Installation
Terminal head form B, DIN-Rail housing or field mount housing
-
Accuracy
(Pt100, -50…200 °C) <= 0,1 K
(Pt100, -58…392 °F) <= 0,18 °F -
Galvanic isolation
yes
-
Certification
ATEX II1G Ex ia IIC T4/T5/T6
ATEX II3G Ex ic IIC T6 Gc
ATEX II2D Ex tb IIIC Db
ATEX II3D Ex tc IIIC Dc
ATEX II3G Ex nA IIC T6 Gc
ATEX II1G Ex ia IIC T6 Ga, II3D
ATEX II3G Ex nA IIC T6 Gc, II3D
ATEX II2G Ex d T6, II2D Ex tb IIIC
CSA C/US General Purpose
CSA IS, NI I / 1+2/A-D
CSA XP, NI, DIP I, II, III/1+2/A-G
FM IS,NI I / 1+2/A-D
FM XP, NI, DIP I, II, III/1+2/A-G
EAC Ex ia IIC T6 Ga
EAC Ex d IIC T6 Gb
EAC Ex ia IIC T6, II3D
IECEx Ex ia IIC T4/T5/T6
IECEx Ex tb IIIC Db
IECEx Ex nA II T4/T5/T6
IECEx Ex d T6 Gb, Ex tb IIIC Db
INMETRO Ex ia [ia Ga] IIC T6 Gb
INMETRO Ex d T6 Gb, Ex tb IIIC Db
NEPSI Ex ia IIC T6 Ga
NEPSI Ex d IIC T6 Gb
NEPSI Ex nA IIC T6 Gc
TIIS Ex ia IIC T6
TIIS Ex nA II T6
UK II1G Ex ia IIC T6 Ga
UK II2D Ex tb IIIC Db
UK II3D Ex tc IIIC Dc
UK II1G Ex ia IIC T6 Ga, II3D Ex ia IIIC Dc
UK II3G Ex nA IIC Gc, II3D Ex tc IIIC Dc
UK II2G Ex db IIC T6 Gb, II2D Ex tb IIIC Db
ATEX IECEx II1G Ex ia IIC T6 Ga
FM+CSA IS, NI I/1+2/A-D
Field of application
The iTEMP TMT82 is a highly reliable, accurate and long-term stable temperature transmitter for industrial processes requiring high availability. The versatile 2-channel device converts universal sensor inputs such as thermocouples or RTD into stable, scalable 4 to 20 mA output signals and HART® communication. With its built-in smart sensor monitoring capability and diagnostic status information according to NAMUR NE 107, the SIL 2, SC3 certified transmitter increases process safety and uptime.
Safe operation in hazardous areas; International approvals such as FM, CSA, ATEX, EAC, NEPSI, IECEx, DNV GL 2-channel universal signal inputs: RTD, thermocouple, resistance (Ω) and voltage transmitters (mV), HART® 7 protocolSuited for use in marine applicationsInstallation in explosion proof field housings; Protection class up to IP 68Fits in flat-face terminal heads as per DIN EN 50446Cabinet or DIN rail mounting; width: 17.5 mm
Benefits
Robust temperature transmitter delivering precise, long-term stable measurements for high plant availabilitySIL certification up to SIL2, SC3 according to IEC 61508:2010Versatile mounting and housing options:terminal head form B, DIN rail, field housing with separate terminal compartment, integrated display (TID10)Transmitter-sensor matching for highest measurement accuracySensor and hardware failure detection; diagnostic status informationaccording to NAMUR NE 107Time-saving installation: Tool-free wiring with push-in terminal technology
Intrinsically Safe Temperature Sensor: iTHERM ModuLine TM131
Trend-setting, highly modular and intrinsically safe RTD or TC thermometer – for a wide range of industrial applications
-
Measuring principle
Resistance Temperature Detector
-
Characteristic / Application
metric style
modular temperature assembly
universal range of application
suitable for hazardous areas
with neck or QuickNeck
incl. thermowell / protection tube (metal) or to be installed into thermowell
can be used with StrongSens, QuickSens insert
for fast response times -
Thermowell / protection tube
welded protection tube or to be used with thermowell
-
Insert / probe
mineral insulated (MI), flexible
-
Outer diameter protection tube / Insert
Thermowell:
9×1,25 mm
11×2 mm
12×2,5 mm
14×2 mm
16×3,5
1/4″ SCH80
1/2″ SCH80
1/2″ SCH40 -
Max. immersion length on request
up to 4.500,0 mm (177”)
-
Material protection tube/ thermowell
316 (1.4401)
316L (1.4404)
316Ti (1.4571)
Alloy 600 (2.4816)
Alloy C276 (2.4819)
Alloy 446 (1.4749)
Alloy 321 (1.4541)
Sheath PTFE
Sheath Tantal -
Process connection
Male thread:
G3/8, G1/2″, G3/4″, G1″
NPT1/2″, NPT3/4″, NPT 1″
M14x1.5, M18x1.5, M20x1.5, M27x2, M33x2
R1/2″, R3/4″
Female thread:
M20x1.5, M24x1.5
NPT1/2″
Cap-nut:
M20x1.5, G1/2″, G3/4″
Compression fitting:
NPT1/2″, G1/2″, G1″
Flange:
DN15 PN40 B1, C
DN25 PN20, PN40, PN100 B1, B2, C
DN40 PN40 B1
DN50 PN40 B1
ASME 1″ 150 RF
ASME 1″ 300 RF
ASME 1 1/2″ 150 RF
ASME 2″ 150 RF
ASME 2″ 300 RF -
Tip shape
straight
reduced
tapered
optimized for quick response times -
Surface roughness Ra
< 0.76 μm (0.03 μin)
-
Operating temperature range
PT100 TF StrongSens:
-50 °C …500 °C
(-58 °F …932 °F)
PT100 QuickSensTF:
-50 °C …200 °C
(-58 °F …392 °F)
PT100 WW:
-200 °C …600 °C
(-328 °F …1.112 °F)
PT100 TF:
-50 °C …400 °C
(-58 °F …752 °F)
Typ K:
max. 1.100 °C
(max. 2.012 °F)
Typ J:
max. 800 °C
(max. 1.472 °F)
Typ N:
max. 1.100 °C
(max. 2.012 °F) -
Max. process pressure (static)
depending on the configuration up to 100 bar
-
Accuracy
Class AA acc. to IEC 60751
Class A acc. to IEC 60751
Class B acc. to IEC 60751
Class special or standard acc. to ASTM E230
Class 1 or 2 acc. to IEC 60584-2 -
Response time
fastest response time with thermowell t90 starting at below 10 s
depending on configuration -
Integration head transmitter
yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)
-
Ex – approvals
ATEX
ATEX IECEx
NEPSI
IECEx
EAC Ex
CSA C/US
INMETRO -
Certification
SIL, MID, DNV/GL
Field of application
Intrinsically safe temperature sensor (RTD / TC assembly) with manufactured thermowell or for use with existing thermowell. An optional head transmitter with all common communication protocols – Bluetooth® technology inclusive – and special features like iTHERM QuickNeck for tool-free, easy recalibration, vibration-resistant and fast-response sensor and thermowell technology (iTHERM StrongSens, QuickSens) as well as hazardous area approvals ensure enhanced measurement accuracy and reliability.
Universal range of applicationMeasuring range: –200 to +1.100 °C (–328 to +2.012 °F)Pressure range up to 100 bar (1450 psi)Degree of protection: up to IP 68Vibration-resistant sensor elements up to 60gImproved ease of maintenance (sensor replacement without process shutdown), easy and safe recalibration of the measuring point
Benefits
Second process barrier with failure indication offering valuable health status information iTHERM QuickSens: fastest response times 1.5 s for optimum process control iTHERM StrongSens: unsurpassed vibration resistance (> 60g) for ultimate plant safety iTHERM QuickNeck: cost and time savings thanks to simple, tool-free recalibrationBluetooth® connectivity (optional) International certification: explosion protection according to ATEX, IECEx, CSA C/US and NEPSI
Field Xpert SMT70
Universal, high-performance tablet PC for device configuration
-
Measuring principle
Device configurator
-
Function
The Field Xpert SMT70 tablet PC for universal device configuration supports protocols such as HART, PROFIBUS DP/PA, FOUNDATION Fieldbus, Modbus, and Endress+Hauser service protocols (CDI, ISS, IPC and PCP). The devices can be connected directly via a suitable interface e.g. a modem (point-to-point) or a bus system (point-to-bus).
-
Input
Camera:
• 5 MP camera on the back with auto focus and LED flash with 1500 mcd
• 2 MP camera on the front
Scanner:
• 1D/2D barcode scanner -
CPU
Intel Core i5-5200U (2,2…2,7 GHz)
-
Memory
• 8 GB with DDR3L adapter
• 256 GB SSD M2-SATA robust version
• Micro SD card slot, support for SDHC/SDXC -
Power supply
Replaceable, rechargeable lithium polymer battery, 11.1 VDC,
4100 mAh (45 Wh)/7800 mAh (86 Wh) (optional) -
Dimensions (WxHxD)
298.5 mm x 192 mm x 20 mm
(11.4 in x 7.48 in x 0.78 in) -
Screen size
• 11.6″ LCD display, with Gorilla Glass protector
• Capacitive multi-touch screen
• Optical bonding, sunlight readable, LCD (730 Nit)
• Ambient light sensor -
Resolution
Field Xpert SMT50
Universal, high-performance tablet PC for device configuration
-
Measuring principle
Device configurator
-
Function
The Field Xpert SMT50 tablet PC for universal device configuration supports various protocols, the Endress+Hauser service protocols and connection to Endress+Hauser Bluetooth field devices. Field devices can be connected directly via suitable interface, e.g. a modem (point-to-point) or a bus system (point-to-bus) or wireless (WLAN/Bluetooth).
-
Input
2 MP camera on the front
5 MP camera or better, on the rear -
CPU
Intel® Skylake Core m3-6Y30 or better
-
Memory
Working memory: 4 GB DDR3L
Memory 128 GB M.2 SSD -
Power supply
Replaceable, rechargeable lithium polymer battery, 7,4 VDC, 6300 mAh (45 Wh)
-
Dimensions (WxHxD)
11.75 in ⋅ 8.5 in ⋅ 0.9 in
-
Screen size
12.2″”
Touch Screen: 10-point capacitive (Corning Gorilla Glass 3)
Brightness: 280 cd/m² -
Resolution
1920 × 1200 pixels
-
Communication interface
2x USB 2.0 + 1x USB 3.0
802.11 a/b/g/n | 2.4GHz + 5.8GHz Dual Band Wi-Fi
Bluetooth 4.0 (BLE) Class 1.5 | Transmission distance: 10 meters
Integrated 4G module + U-blox NEO-7 -
Operation
Operating system: Windows 10 LTSC
-
Configuration
Touch-enabled device configuration software optimized for Field Xpert that is characterized by easy, fast and intuitive operation. The Field Xpert device library has more than 3000 preinstalled device and communication drivers. They can be used to operate practically all HART and FOUNDATION Fieldbus devices (FieldComm Group libraries). Furthermore all Endress+Hauser field device drivers are installed. Further device drivers (DTMs) can be installed additionally. The generic HART DTM and PROFIBUS profile DTMs also enable operation of all the important basic functionalities of the relevant field devices.
-
Optional
4G LTE WWAN + 12 channel GPS
-
Certificates
Degree of protection:
IP65
Vibration resistance:
MIL-STD 810G 514.6 Procedure I Cat. 24, Fig. 514.6E-1
Shock resistance:
MIL-STD-810G
Interference conformance to:
EN 55022: 2010+ AC: 2011
EN 55022 Class B
EN 61000-3-2
EN 61000-3-3
Interference immunity:
EN 55024: 2010
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-8
IEC 61000-4-11
Approvals:
CE, FCC”
Field of application
The Field Xpert SMT50 tablet PC for universal device configuration supports various protocols, the Endress+Hauser service protocols and connection to Endress+Hauser Bluetooth field devices. Field devices can be connected directly via suitable interface, e.g. a modem (point-to-point) or a bus system (point-to-bus) or wireless (WLAN/Bluetooth). The Field Xpert software package is fast, easy and intuitive to use. Its device library has more than 3000 pre-installed device and communication drivers.
The Field Xpert SMT50 tablet PC for device configuration enables mobile plant asset management in non-hazardous areas. It is suitable for commissioning and maintenance. It manages Endress+Hauser and third-party instruments with a digital communication interface and documents the progress of the work. The SMT50 is designed as a complete solution. It comes with a pre-installed driver library and is an easy-to-use, touch-enabled tool for managing field devices throughout their entire life cycle.
Benefits
Robust tablet with large 12″ display and a battery runtime of up to 8 hours.Touch-enabled device configuration software for optimized asset management.Direct link to Netilion, Endress+Hauser’s IIoT ecosystem.Pre-installed device driver libraries for all of the important industrial protocols.Fast connection to devices with a single click thanks to automatic hardware detection. Online application software and DTM/Drivers update service offers new functionality and maximum security.Integrated Heartbeat Verification incl. PDF documentation.Connection to Endress+Hauser Bluetooth and WLAN devices – perfect for difficult-to-access instruments.
RMA42 Process transmitter with control unit
Universal transmitter, loop power supply, barrier and limit switch in one device!
-
Measuring principle
Process transmitter
-
Measuring principle
Process transmitter
-
Function
Universal process and temperature transmitter with control unit
-
Loop power supply
24 V intrinsically safe loop power supply
-
Power Supply
Wide range power supply 24 to 230 V AC/DC (-20 % / +10 %) 50/60 Hz
-
Input
2 x universal (current, voltage, R, RTD, TC, resistance)
-
Output
2 x Analog (current, voltage)
-
Software functions
Internal software for calculations, linearization, limit monitoring
Monitoring of sensor wires according NAMUR NE43
Application package differential pressure measurement -
Certificates
ATEX
FM
NEPSI
CSA
UKCA
UR C/US
marine approval
KTA
German WHG overfill protection -
SIL
2
-
Display
LCD
7 segment
Multicolour
Bargraph
TAG unit
Colour change in case of failure -
Interface
Not defined
-
Dimensions (WxHxD)
45 x 115 x 118 mm (1.77″ x 4.53″ x 4.65″)
-
Measuring principle
Process transmitter
-
Function
Universal process and temperature transmitter with control unit
-
Certification
ATEX
FM
NEPSI
CSA
CSA GP
UR C/US
Marine approval
KTA
German WHG overfill protection -
Input
2 x universal (current, voltage, R, RTD, TC, resistance)
-
Output
2 x Analog (current, voltage)
-
Relay output
2 x relay
1 x open collector -
Auxiliary power supply / Loop power supply
24 V intrinsically safe loop power supply
-
Dimensions (wxhxd)
45 x 115 x 118 mm
(1.77″ x 4.53″ x 4.65″) -
Software
Internal software for calculations, linearization, limit monitoring
Monitoring of sensor wires according NAMUR NE43
Application package differential pressure measurement -
Operation
On site via 3 buttons
PC configuration via FieldCare -
Display
LCD
7 segment
Multicolour
Bargraph
TAG unit
Colour change in case of failure -
Operating Voltage
20-250V AC/DC wide range power supply
-
Number of poles
.
Field of application
Analog measured values are indicated, evaluated and calculated by the process transmitter. With the integrated loop power supply two-wire sensors can be powered. The universal inputs enable the connection of current, voltage, RTDs and TCs. Limit values can be monitored and relays can be switched. Via analog outputs process signals can be forwarded. The integrated application package “differential pressure” allows the quick and easy commissioning for differential pressure applications.
The RMA42 process transmitter exhibits a display, analog inputs for process and status signals, analog and digital outputs, and an interface for configuration. Connected sensors can be powered by the integrated LPS. All measured and calculated values are available as a signal source for display, outputs, relays and the interface.Plant and apparatus engineeringControl rooms and cabinetsProcess control, recording and supervisionSignal adjustment and conversionOverfill protection (WHG)
Benefits
SIL 2 approval (optional);
Marine approval1 or 2 calculated values and linearization tableMin./max. value saved5-digit 7-segment LC display, backlit1 or 2 universal inputs2 relays (optional)Digital status output (Open Collector)
RIA15 Loop-powered indicator for 4 to 20 mA or HART signals
For the indication of the 4 to 20 mA signal or the HART® values in the field or in the panel!
-
Measuring principle
Loop powered indicators
-
Measuring principle
Loop-powered field indicator
-
Function
Display
-
Display
LCD 17 mm (0.67″)
5-digit
7-segment bargraph
Plain text display for unit/TAG
Activatable backlight -
Power Supply
Loop powered
-
Input
1 x analogue 4…20 mA/HART®
-
Output
Not defined
-
Output
Not defined
-
Software functions
Not defined
-
Certificates
ATEX II2(1)G Ex ib [ia Ga] IIC T6 Gb
CSA C/US IS CI.I Div.1 Gr.A-D CSA C/US NI CI.I Div.2 Gr.A-D
CSA Ex ia, CI. I, Div. 1, Gr. ABCD
FM IS, CI. I, Div. 1, Gr. ABCD
EAC 1Ex ib IIC T6 Gb
IECEx Ex ib [ia Ga] IIC T6 Gb
JPN Ex ib IIC Gb
NEPSI Ex ib IIC Gb
UK II2(1)G Ex ib [ia Ga] IIC T6 Gb
ATEX IECEx II2(1)G Ex ib [ia Ga] IIC T6
marine certificate
UL C/US -
SIL
Not defined
-
Panel mounting
field
panel mounting
Field of application
The process indicator is available with 4 to 20 mA or with HART® protocol. In the HART® version up to 4 measured values of one sensor can be indicated in alternation. For this the indicator can be set as primary or secondary HART® master and actively inquire the values from the sensor. The small voltage drop of < 1V (< 1.9 V with HART®) is a huge advantage especially in Ex-applications, where in most cases only a reduced voltage is available.
Display of 4 to 20 mA measured values or HART® process variables in all industriesPanel-mount housing or field housing as local displayScalable display valueWith the “Level” or “Analysis” options, basic settings can be made on selected sensors and available measured values and units can be preset quickly and easily.
Benefits
Display of 4 to 20 mA measured values or optionally up to four of a sensor’s HART® process variables in all industries; Use as primary or secondary HART® master5-digit measured value display with 17 mm (0.67″) character height with dimension and bargraph; backlight can be activated Panel-mount housing (IP65) or field housing (IP66; Ex approval) as local display; Minimal installation depthNo external power supply required; Voltage drop ≤1 V (HART® ≤1.9 V) Easy 3-key operation for configuring the deviceInternational approvals; marine approvalSIL interference freeness acc. to EN 60508
EngyCal RH33 BTU meter
Custody transfer BTU meter for recording heat quantities of water
-
Measuring principle
Energy manager
-
Measuring principle
Energy manager
-
Function
Monitoring and billing of energy in heating and cooling applications as well as combined heating-and cooling circles.
Typical applications can be found in industries, district heating and building automation. -
Calculations
heat quantity and heat quantity difference
-
Number of applications
1
-
Data storage
yes
-
Calculation standards
IAPWS-97
-
Communication
web server
USB
Ethernet
Modbus RTU/TCP Slave
M-Bus -
Power supply
Not defined
-
Loop power supply
Low voltage power supply:
100 bis 230 V AC (-15% / +10%)
Small voltage power supply:
24 V DC (-50% / +75%)
24 V AC (±50%) -
Protection class
IP65
-
Input
1x Puls/Analog
2x RTD/Analog
Loop power supply 24V DC (+/-16%) -
Output
1x 4…20mA
2x digital (Open Collector) -
Dimensions (WxHxD)
144 x 144 x 103.1 mm (5.67″ x 5.67″ x 4.06″)
-
Operation
3 button on site or via FieldCare
read out of historical / logged data via Field Data Manager Software (SQL database and visualization interface)
selectable language -
Display
160 x 80 Dot-Matrix LCD with white backlit
colour change in case of alarm event
active display area 70 x 34 mm -
Software functions
Calculated Values:
Energy, volume, density, enthalpy & enthalpy difference, DP-Flow-Compensation, mass, temperature difference
Counters:
Volume, mass, energy, counter in case of failure
Optional:
tariff 1, tariff 2 or seperated heating energy, cooling energy, bilance energy -
Certificates
CE
CSA GP
MID 004 (custody transfer) according EN1434 (water/other liquids)
OIML R75
-
Measuring principle
Energy manager
-
Function
Monitoring and billing of energy in heating and cooling applications as well as combined heating-and cooling circles.
Typical applications can be found in industries, district heating and building automation. -
Certification
CE, UL listed, CSA GP, MID 004 (custody transfer) according EN1434 (water/other liquids) and OIML R75
-
Input
1x Puls/Analog,
2x RTD/Analog,
Loop power supply 24V DC (+/-16%) -
Output
1x 4…20mA,
2x digital (Open Collector) -
Relay output
2x
-
Auxiliary power supply
Low voltage power supply: 100 bis 230 V AC (-15% / +10%); Small voltage power supply: 24 V DC (-50% / +75%), 24 V AC (±50%)
-
Dimensions (wxhxd)
144 x 144 x 103.1 mm
(5.67″ x 5.67″ x 4.06″) -
Software
Calculated Values:
Energy, volume, density, enthalpy & enthalpy difference,
DP-Flow-Compensation,
mass,
temperature difference;
Counters: volume, mass, energy, counter in case of failure
Optional: tariff 1, tariff 2 or seperated heating energy, cooling energy, bilance energy -
Operation
3 button on site or via FieldCare; read out of historical / logged data via Field Data Manager Software (SQL database and visualization interface), selectable language
-
Display
160 x 80 Dot-Matrix LCD with white backlit, colour change in case of alarm event, active display area 70 x 34 mm
-
Others
electronic temperature sensor matching via CvD coefficients, logbook measured values, custody transfer logbook, event logbook, limit monitoring
-
Calculations
IAPWS-IF97
Field of application
The BTU meter RH33 is used in applications with liquid energy carriers.It calculates the thermal energy of water acc. to EN1434, glycol/water mixtures or other fluids such as thermal oils. Temperature sensor matching using calibrated temperature sensors is done in the device. Potential for cost savings can be shown by using the software available as an accessory.The device has a custody transfer approval and allows bi-directional measurement, e.g. charging/discharging of a heat accumulator.
Recording and billing the quantities of heat and cold in:Heating circuitsCooling circuitsCombined heating and cooling circuitsTypical application areas in industry, long-distance heat networks and building automation.
Benefits
Transparent energy consumption helps you save on energy costsCalibrated, electronically paired temperature sensors ensure the highest accuracy and enable replacement of individual temperature sensors even for certified devices in the field (without reapproval!)Tariff counter for requirements-based billingDetailed data logging of current and counter values and of error messages, off-limit conditions and changes to operating parametersStandard models are suitable for connecting and supplying all common volume flow transmitters and temperature measuring pointsRemote readout via Ethernet and
Cerabar PMP71B – pressure transmitter
Smart pressure transmitter – its health can be verified without process interruption
-
Measuring principle
Absolute and gauge pressure
-
Characteristic
Pressure transmitter with metallic process diaphragm.
Available with diaphragm seals.
Device health verification without process interruptions.
Intuitive and clear wizards for commissioning, proof tests and verification. -
Supply voltage
4…20 mA HART
Ex d, Ex e, non-Ex:
10.5…35V DC :
Ex i: 10.5…30V DC -
Reference Accuracy
Standard:
up to 0.05 %
Platinum:
up to 0.025 % -
Long term stability
0.05 % of URL/ year
0.07 % of URL/ 5 years
0.10 % of URL/ 10 years -
Process temperature
Standard:
-40°C…+125°C
(-40°F…+257°F)
Diaphragm seal:
-70°C…+400°C
(-94°F…+752°F) -
Ambient temperature
-60°C…+80°C
(-76°F…+185°F) -
Measuring cell
400 mbar…700 bar
(6 psi…10,500 psi) -
Smallest calibratable span
5 mbar (0.075 psi)
-
Vacuum resistance
5 mbar (0.725 psi)
-
Max. Turn down
100:1
-
Max. overpressure limit
2,800 bar (40,600 psi)
-
Process connection
Thread:
G1/2…G2, R1/2, MNPT3/4…MNPT2,
FNPT1/2, R1/2
Flange:
DN25…DN100,
NPS 1″…4″
Pancake: 2″…4″, DN50…DN100 -
Material process membrane
316L, AlloyC,
Tantal, Monel,
PTFE,
Gold -
Material gasket
PTFE
-
Fill fluid
Deltabar PMD75B – differential pressure transmitter
Smart pressure transmitter which detects process anomalies like plugged impulse lines
-
Measuring principle
Differential pressure
-
Characteristic
Differential pressure transmitter which detects process anomalies like plugged impulse lines.
Intuitive and clear wizards for commissioning, proof tests and verification. -
Supply voltage
Ex d, Ex e, non-Ex:
10.5…35V DC
Ex i: 10.5…30V DC -
Reference Accuracy
Standard:
up to 0.05 %
Platinum:
up to 0.035 % -
Long term stability
0.025 % of URL/ year
0.05 % of URL/ 5 years
0.10 % of URL/ 10 years -
Process temperature
-40°C…+110°C
(-40°F…+230°F) -
Ambient temperature
-40°C…+85°C
(-40°F…+185°F) -
Measuring cell
10 mbar…250 bar
(0.15 psi…3750 psi) -
Smallest calibratable span
0.25 mbar (0.00375 psi)
-
Vacuum resistance
25 mbar (0.0375 psi)
-
Max. Turn down
100:1
-
Max. overpressure limit
1600 bar (23,200 psi)
-
Process connection
NPT1/4-18
RC1/4″ -
Material process membrane
316L, AlloyC,
Tantal,
Monel,
Gold -
Material gasket
PTFE, FKM, EPDM, FFKM
-
Fill fluid
Silicone oil,
Inert oil -
Material housing
Aluminium, 316L
-
Communication
4…20 mA HART
PROFINET with Ethernet-APL
Bluetooth® wireless
technology and app (optional) -
Certificates / Approvals
ATEX, CSA, IECEx, JPN Ex, INMETRO, NEPSI, UK Ex
-
Safety approvals
SIL
-
Design approvals
EN10204-3.1,
NACE MR0103,
NACE MR0175 -
Drinking water approvals
NSF, KTW
-
Specialities
Heartbeat Technology,
Bluetooth® Operation and maintenance SmartBlue App,
LED Module,
RFID TAG for easy identification,
Plug and play functionalities,
HistoROM
-
Measuring principle
Differential pressure
-
Characteristic / Application
Differential pressure transmitter which detects process anomalies like plugged impulse lines.
Intuitive and clear wizards for commissioning, proof tests and verification. -
Specialities
Heartbeat Technology,
Bluetooth® Operation and maintenance with SmartBlue App,
LED Module,
RFID TAG for easy identification,
Plug and play functionalities
HistoROM -
Supply / Communication
Ex d, Ex e, non-Ex:
10.5…35V DC
Ex i:
10.5…30V DC -
Accuracy
Standard:
up to 0.05 %
Platinum:
up to 0.035 % -
Long term stability
0.025 % of URL/ year
0.05 % of URL/ 5 years
0.10 % of URL/ 10 years -
Ambient temperature
-54°C…+85°C
(-65°F…+185°F) -
Process temperature
-40°C…+110°C
(-40°F…+230°F)
Raman Rxn2 analyzer
Bridge your application from the laboratory to the process environment
-
Measuring principle
Raman spectroscopy
-
Spectral coverage
Starter:
300-3300 cm-1 (785 nm)
Base Model:
150-4350 cm-1 (532 nm)
150-3425 cm-1 (785 nm)
200-2400 cm-1 (1000 nm)
Hybrid: 175-1890 cm-1 (785 nm) -
Spectral resolution
Starter:
10 cm-1 (average)
Base model (average):
5 cm-1 (532 nm)
4 cm-1 (785 nm)
5 cm-1 (1000 nm)
Hybrid:
4 cm-1 (785 nm) average -
Channels
Starter:
Single channel
Base Model:
Up to four channels
Hybrid:
Up to two channels -
Temperature
Operating:
15 to 30 °C
59 to 86 °F
Storage:
-15 to 50 °C
5 to 122 °F -
Relative humidity
20 to 80 %, non-condensing
-
Input voltage
100-240 V, 50-60 Hz, ±10%
-
Power consumption (W)
400 (max)
250 (typical start-up)
120 (typical running) -
Warm up time (minutes)
120
-
Unit dimensions (width x height x depth)
Benchtop model:
279 x 483 x 592 mm
10.99 x 19.02 x 23.31 inch
Cart model:
685 x 1022 (to tabletop) x 753 mm
26.97 x 40.24 (to tabletop) x 29.65 inch -
Weight
Base model:
32 kg / 71 lbs
Cart model:
93 kg / 205 lbs -
Sampling probe compatibility
Starter:
Raman Rxn-10 (with accessory optics)
Base model:
Raman Rxn-10 (with accessory optics), Rxn-40, Rxn-41, Rxn-45, Rxn-46
Hybrid:
Channel 1 – Raman Rxn-20 (with accessory optics)
Channel 2 – Raman Rxn-10 (with accessory optics), Rxn-40, Rxn-41, Rxn-45, Rxn-46 -
Automation interface
OPC
Modbus
HTTPS
(contact us for other options) -
Installation options
Benchtop (standard) or mobile wheeled cart
-
Hazardous area certifications
ATEX, CSA, IECEx
-
Measuring principle
Raman spectroscopy
-
Spectral coverage
Starter:
300-3300 cm-1 (785 nm)
Base Model:
150-4350 cm-1 (532 nm)
150-3425 cm-1 (785 nm)
200-2400 cm-1 (1000 nm)
Hybrid: 175-1890 cm-1 (785 nm) -
Spectral resolution
Starter:
10 cm-1 (average)
Base model (average):
5 cm-1 (532 nm)
4 cm-1 (785 nm)
5 cm-1 (1000 nm)
Hybrid:
4 cm-1 (785 nm) average -
Channels
Starter:
Single channel
Base Model:
Up to four channels
Hybrid:
Up to two channels -
Temperature
Raman flow assembly
Precise composition measurement of high-speed, low volume samples for downstream optimization
-
Measuring principle
Raman Spectroscopy
-
Temperature
Micro flow bench:
Operating:
0 to 40 °C / 32 to 104 °F
Storage:
-10 to 50 °C / 14 to 122 °F
Micro flow cell:
Operating:
0 to 40 °C / 32 to 104 °F
Storage:
-20 to 60 °C / -4 to 140 °F -
Relative humidity
Micro flow cell:
Storage:
< 90 % Non-Condensing -
Unit dimensions (width x height x depth)
Micro flow bench:
254 x 102 x 66 mm
10.0 x 4.0 x 2.6 inch
Micro flow cell:
46 x 25 x 11 mm
1.8 x 1.0 x 0.4 inch -
Sampling probe compatibility
Raman Rxn-10 in tandem
-
Sample interface
Micro flow bench:
Micro flow cell -
Pressure
Micro flow cell:
Max pressure: 4.8 barg / 70 psig -
Wetted materials
Micro flow cell:
Body: 316L stainless steel
Window: Proprietary material, optimized for bioprocesses
Surface finish: Ra 15 with Electropolish to ASME BPE SF4 finish
Adhesive: USP Class VI and ISO 10993 compatible -
Sterilization method
Micro flow cell:
CIP/SIP -
Process connection
Micro flow cell:
5/16-24 flat bottom ferrules
Up to 3/16 inch OD tubing -
Fluid Path Thickness
Micro flow bench:
1.0 to 5.0 mm (flow cell driven)
Micro flow cell:
2.5 mm -
Flow Conditions (Flowcell)
Micro flow cell:
Laminar -
Flow Cell Maximum Volumetric Flowrate (mL/min)
Micro flow cell:
15 -
Amplification Available or Amplification Factor?
Micro flow bench:
Yes
> 3.0 (TYP)
> 2.5 (MIN)
Field of application
The Raman flow assembly brings in-line Raman spectroscopy to downstream bioprocessing. Comprised of a flow cell and associated flow bench connected to an Rxn-10 probe, this system meets the speed demands of downstream by optimizing Raman collection for low-turbidity, low volume samples. Ideally suited for development-scale flow rates, the Raman flow assembly provides real-time monitoring, rapid process optimization, and process control for chromatography and other downstream unit operations.
Downstream: aggregation, protein crystallization, formula stability, product CQA, protein concentration, buffer excipients Midstream: analyte monitoring in permeate
Benefits
Brings Raman-based process insights and control to downstream bioprocessing unit operations Achieves lower limits of detection through enhancement of the Raman signal Provides continuous, real-time measurements with high sensitivity for development-scale flow rates Easily connects to an Rxn-10 probe paired with a Raman Rxn embedded analyzer, and ensures transferability when used with the Raman flow assembly calibration and verification kitWith the flow bench, provides flexibility to swap out flow cells any time without impacting your process or sterility
Raman bio multi optic and bio sleeve
Maintain sterility and spectral quality while reducing maintenance and cleaning burdens
-
Measuring principle
Raman spectroscopy
-
Temperature
-30 to 150 °C / -22 to 302 °F
-
Sampling probe compatibility
Raman Rxn-10
-
Pressure
Bio Sleeve:
Max pressure: 13.8 barg / 200 psig -
Wetted materials
Bio sleeve:
Body: 316L stainless steel
Window: Proprietary material, optimized for bioprocesses
Surface finish: Ra 15 with Electropolish to ASME BPE SF4 finish
Adhesive: USP Class VI and ISO 10993 compatible -
Length
120 or 220 mm
-
Diameter
Bio sleeve:
12 mm -
Sterilization method
Bio sleeve:
Autoclave, gamma irradiation -
Process connection
Bio Sleeve:
PG13.5
Field of application
The bio multi optic and bio sleeve duo delivers reliable, real-time bioprocess measurements with maximum sterility and efficiency. The non-contact multi optic features a patented self-aligning lens for top spectra quality and reproducibility. It pairs with the Rxn-10 probe and connects to an autoclavable (or gamma sterilizable), disposable sleeve via an easy snap-in-place design. The system reduces maintenance and cross-contamination risk for reusable or single-use benchtop bioreactor setups.
Benchtop bioreactors to measure glucose, lactate, amino acids, cell density, titer and more Benchtop fermenters to measure glycerol, methanol, ethanol, sorbitol, biomass, and more Use with flow assembly CYA680 for select downstream bioprocessing applications
Benefits
Patented self-focusing lens with automatic alignment for optimal spectral quality and reproducibility Improved ease of use with less probe service and maintenance Simplified workflow for greater process efficiency – swap out probes mid-batch, use multiple sleeves in the same reactor for simplified redundancy, and calibrate / verify probes at any time Autoclavable, gamma sterilizable, and compatible with standard cleaning protocols (bio sleeve), with reduced risk of cross-contamination (via non-contact multi optic) 120 mm version is compatible with Endress+Hauser flow assembly CYA680
Raman non-contact optics
Protect your samples with contact-free measurement
-
Measuring principle
Raman spectroscopy
-
Sampling probe compatibility
Raman Rxn-10 probe
-
Sample interface
Temperature: ambient
-
Pressure
Ambient
-
Length
Varies based on model
-
Diameter
Varies based on model
-
Working distance
NCO-0.4-NIR: 10 mm / 0.40 inch
NCO-0.5-VIS: 12.5 mm / 0.50 inch
NCO-1.3-VIS: 33 mm / 1.30 inch
NCO-2.5-VIS: 64 mm / 2.52 inch
NCO-3.0-NIR: 75 mm / 2.96 inch
NCO-5.5-VIS/NIR: 140 mm / 5.52 inch
Field of application
Paired with the Rxn-10 probehead, Raman non-contact optics provide contact-free Raman measurements of samples either directly or through sight glass or translucent packaging. Available with a range of working distances, these highly versatile optics are the perfect choice for use with solids or turbid media. They are also well suited for delicate or corrosive samples when sample contamination or damage to optical components is a concern.
Pharmaceutical: Drug product manufacturingChemical: Polymer solids (pellets, films, or powders)Food & Beverage: Meat or fish qualityRaw material identificationFormulation optimization
Benefits
Range of working distances for remote measurements either directly or through sight glass and translucent packagingHighly versatile – measuring films to pellets to powdersAccurate measurement of static or moving samplesContact-free analysis of delicate or corrosive samples
Raman bIO-Optics
Continuous, inline, multi-component bioprocess measurement
-
Measuring principle
Raman spectroscopy
-
Sampling probe compatibility
Raman Rxn-10 probe
-
Sample interface
Temperature: –30 to 150 °C / -22 to 302 °F
-
Pressure
Max pressure: 13.8 barg / 200 psig
-
Wetted materials
Body: 316L stainless steel
Window: Proprietary material, optimized for bioprocesses
Process connection: PG13.5
Surface finish: Ra 15 with electropolish
Adhesive: USP Class VI and ISO 10993 compatible -
Length
120, 220, 320, or 420 mm
4.73, 8.67, 12.60, 16.54 inch -
Diameter
12 mm / 0.48 inch
-
Sterilization method
Autoclave
-
Process connection
PG13.5
Field of application
Raman bIO-Optics are versatile immersion optics used in conjunction with the Rxn-10 probehead and compatible with standard bioreactor ports. The fixed focus design provides long-term measurement stability along with superior signal performance, essential for transferable, high performance Raman-based bioprocess analysis. Available in industry standard lengths, Raman bIO-Optics are ideally suited for both glass and plastic benchtop bioreactor/fermenter applications requiring headplate entry.
Cell culture: glucose, lactate, glutamine, amino acids, cell density, titerFermentation: glycerol, methanol, ethanol, sorbitol, biomassDownstream: aggregation, protein crystallization, formula stability, product CQA, protein concentration, buffer excipients
Benefits
Immersion optics for multi-component bioprocess measurementCompatible with standard bioreactor portsFixed-focus designAvailable in 12 mm diameter with Pg 13.5 threaded connector120, 220, 320, or 420 mm lengthsAutoclavable
Raman immersion optic
Unlock the power of in situ laboratory analysis of liquids, slurries, and semi-solids
-
Measuring principle
Raman spectroscopy
-
Sampling probe compatibility
Raman Rxn-10 probe
-
Sample interface
Temperature range:
-30 to 120 °C / -22 to 248 °F (316L stainless steel)
-30 to 280 °C / -22 to 536 °F (C276 alloy)
-30 to 300 °C / -22 to 572 °F (Grade 2 Titanium) -
Pressure
Min pressure:
Full vacuum (0 bara) but not ultrahigh vacuum (UHV) where outgassing maybe a concern
Max pressure 1/4 inch IO:
68.5 barg/ 993 psig (316L stainless steel)
74 barg/ 1073 psig (C276 alloy)
29 barg/ 420 psig (Grade 2 titanium)
Max pressure 1/2 inch IO:
68.5 barg/ 993 psig (316L stainless steel)
74 barg/ 1073 psig (C276 alloy)
29 barg/ 420 psig (Grade 2 titanium) -
Wetted materials
Metal: C276 alloy as standard (316L stainless steel or Grade 2 titanium upon request)
Window: high-purity sapphire, proprietary compression fit non-brazed design -
Length
Immersion Optic shaft length:
152, 305, or 457 mm
6, 12, or 18 inch
Outer diameter: 12.7 mm / 0.5 inch
Immersion Optic shaft length:
152 or 203 mm
6 or 8 inch
Outer diameter: 6.35 mm / 0.25 inch -
Working distance
Short (S): 0 mm / 0 inch
Long (L): 3 mm / 0.12 inch -
Operating range
VIS or NIR
Field of application
The fixed focus design of the Raman immersion optic provides long-term measurement stability along with superior signal performance which is essential for transferable, high-performance Raman measurements. Compatible with the Rxn-10 probehead, the immersion optic uses no movable parts so it is always in focus. The Raman immersion optic is well-suited for use in reaction vessels, laboratory reactors, or process streams, and its wettable components resist corrosion in harsh environments.
Pharmaceutical: drug substance unit operations, reaction analysis, crystallization, end-point detection, solvent swapsChemical: material identification, reaction analysis, polymerization, cross-linking, blendingFood & Beverage: blending, purification, natural and synthetic componentsDevelopment laboratory
Benefits
Ideal for quick liquids, slurries & semi-solids measurements in the labSuperior signal performanceAvailable in 1/4” and 1/2” diametersNo user-adjustable partsWettable components resist corrosion in harsh environments (including acidic solutions)
Raman electro-optical (EO) fiber optic cables
Innovative Raman fiber optics designed for maximum laser safety and easy installation
-
Measuring principle
Raman spectroscopy
-
Temperature
Range:
-40 to 70 °C / -40 to 158 °F -
Fiber optic cable
Length: 5000 mm / 196.86 inch (custom lengths available)
Bend radius: 152.4 mm / 6 inch
Field of application
Our Raman embedded analyzers and probes are connected by electro-optical (EO) fiber optic cables with integrated low-voltage electrical interlock and optical fiber cable components bundled into one CSA-rated assembly. Rated for indoor/outdoor use, the EO connector is designed to protect the connection from external influences, avoid breakage, and ensure a secure connection. EO fiber cables are available in several size lengths with extension cables to simplify installation and remote analysis.
Used in all Raman Rxn embedded analyzer installationsSupports remote analysis of both lab and process measurementsUsed in a variety of industries: Life Sciences, Chemical, Oil & Gas, Food and Beverage
Benefits
Simple plug and lock designIntegrated safety interlock to prevent laser exposureCertified for hazardous area environmentsIndoor/outdoor rated cable – flame retardant, fungus resistant, internal strength memberAvailable in different fiber lengths to meet your installation requirements
JT33 TDLAS gas analyzer
Reliable H2S measurement for increased quality, process control, and asset integrity
-
Measuring principle
TDLAS
-
Channels
1
-
Analyte and measurement ranges
H2S (Hydrogen sulfide):
0 to 10 ppmv
0 to 500 ppmv
other ranges by request -
Measured variables
Concentration
Cell pressure
Cell temperature -
Ambient temperature range
JT33 TDLAS gas analyzer system:
–20 to 50 °C (–4 to 122 °F)
–10 to 60 °C (14 to 140 °F) -
Operating pressure range
Application dependent
800 to 1200 mbara (standard)
800 to 1700 mbara (optional) -
Analyzer wetted materials
316L stainless steel
FKM O-Rings
Glass
PCTFE/PTFE -
Power supply
For controller (spectrometer):
AC 100 to 240 V tolerance ±10 %, 50/60 Hz, 10W
DC 24 V tolerance ±20 %, 10W
UM = AC 250 V
For sample conditioning system electronics:
AC 100 to 240 V ±10 %, 50/60 Hz, 275W
UM = AC 250 V -
Communication
I/O1: Modbus RTU/RS485 or Modbus TCP/Ethernet
I/O2 and 3: Relay output or Universal I/O (UIO); UIO can be configured as analog input/output (4-20 mA) or digital/status output
Service web server interface: Ethernet RJ45 -
Housing materials
Electronics: Coated copper-free aluminum or Cast stainless steel
Sample system enclosure: 304 stainless steel or 316 stainless steel
Sample system panel: Anodized aluminum
Enclosure window (optional): Polycarbonate -
Hazardous area approvals
ATEX/IECEx/UKEx Zone 1
CSA Class I, Division 1
CSA Class I, Zone 1 -
Degree of protection
IP66, Type 4X
-
Product safety
CE
RCM
FCC
CRN
J22 TDLAS gas analyzer
Proven technology for accurate and reliable measurement of H2O in natural gas
-
Measuring principle
TDLAS
-
Product Headline
Best-in-class gas analyzer for moisture (H2O) that is exceptionally reliable and tailored for the natural gas industry. The sensor measures gas using a tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the gas without coming into physical contact with the stream. Heartbeat Technology ensures compliance and process safety at all times.
-
Channels
1
-
Analyte and Measurement ranges
H2O (Moisture):
0 to 50 ppmv (minimum)
0 to 6000 ppmv (maximum) -
Measured Variables
Concentration
Dew Point
Cell Pressure
Cell Temperature -
Ambient Temperature range
-20 to 60°C (-4 to 140°F)
-
Operating Pressure range
Inlet Pressure: 140 to 310 kPa (20 to 45 psig)
Sample Cell: 800 to 1200 mbara or 800 to 1700 mbara (optional) -
Analyzer Wetted materials
316L stainless steel
FKM O-Rings
Glass -
Power supply
Without heater: 24 VDC ±20% or 100 to 240 VAC ±10%, 50/60 Hz, 10W
With optional heater: 100 to 240 VAC ±10%, 50/60 Hz, 80W -
Communication
Service webserver interface: Ethernet RJ45
I/O1: Modbus RS485 or Modbus TCP
I/O2 and 3: Relay output or Universal I/O (UIO); UIO can be configured as analog input/output (4-20 mA) or digital/status output -
Housing materials
Electronics: Copper-free aluminum
Sample System Enclosure: 304 stainless steel or 316 stainless steel
Sample System Panel: Anodized aluminum -
Hazardous area approvals
ATEX / IECEx /UKEx Zone 1
PESO / KTL / CML Zone 1
INMETRO Zone 1
CSA Class I, Division 1
CSA Class I, Zone 1 -
Degree of protection
IP66, Type 4X
-
Product safety
CE
RCM
FCC
CRN
Field of application
The J22 gas analyzer uses patented tunable diode laser absorption spectroscopy (TDLAS) technology to provide accurate measurement of H2O in natural gas. Featuring the reliable diagnostic capabilities of Heartbeat Technology, the J22 analyzer determines the concentration of a gas without coming into physical contact with streams. It allows pipeline operators and suppliers to meet quality specifications, prevent pipeline corrosion, and stop hydrate formation to ensure safety and asset integrity.
For use in natural gas for quality measurements in production, storage, transportation, and distribution Monitoring and control in the natural gas pipeline, processing plant, compressor station, gathering site, and distributionIdeal for custody transfer measurements to ensure gas quality Measuring at the inlet and outlet of the dehydration process to ensure dry gas product quality in H2 production and CCUS applications Measurable ranges of up to 0-6000 ppm (0-284 lb/mmscf) H2O
Benefits
Ensures highest availability for production, transmission, storage and distribution of natural gasUser-friendly interface with intuitive menu and web server software Integrated diagnostic and verification with Heartbeat Technology Automatically-stored historical data and spectrum logging NIST-traceable calibration providing superior accuracy and repeatabilityRobust design for easy installation, commissioning, and repairField serviceable components and modules for minimal downtime
Commubox FXA195 USB/HART modem
World’s only intrinsically safe HART/USB interface for Device & FieldCare and FieldXpert software
-
Measuring principle
Communication Interface
-
Function
Connects intrinsically safe Smart transmitters with HART protocol to the USB port of a personal computer or laptop. This makes the remote operation of the transmitters possible with FieldCare.
-
Input
USB signal as per USB specification.
-
Output
Connects HART field devices with the USB interface of a laptop
-
Certificates
ATEX
FM
CSA
Field of application
The Commubox FXA195 modem connects HART logged, intrinsically safe measuring devices to the USB interface of a Windows system.
This makes the remote operation of the transmitters possible with the aid of the Endress+Hauser FieldCare software. Power is supplied to the Commubox by means of the USB interface.
Benefits
Remote commissioning of field devices directly from the control room or anywhere in the current loopFull support for FieldCare softwareFor connection to intrinsically safe circuits (4…20 mA)Reactionless in accordance with IEC 61508 for connection to 4…20 mA SIL 2 loopsSwitchable communication resistorAdapter cable for the FXA195 USB/HART Modem supplied (4mm/2mm)
Liquiphant FTL51B – digital, simple and safe
Point level switch for all liquid media
-
Measuring principle
Vibration Liquids
-
Characteristic / Application
Compact vibronic device with an optional pipe extension up to 6 m (19.7 ft);
Modular housing concept;
Wide range of certificates -
Specialities
Heartbeat Technology,
Bluetooth® Operation and maintenance SmartBlue App,
LED Module,
RFID TAG for easy identification,
Second line of defense,
Ambient temperature down to -50°C (-60°C optional),
Temperature spacer for high process temperature,
Plug and play functionalities -
Supply / Communication
19…253V AC / 2-wire,
10…55V / DC-PNP 3-wire,
9…20V DC with relay DPDT,
19…253V AC or 19…55V DC with relay DPDT,
2-wire NAMUR,
2-wire PFM,
4-20mA HART -
Ambient temperature
-60 °C…+70 °C
(-51 °F…+158 °F) -
Process temperature
-50 °C…+150 °C
(-58 °F…+302 °F) -
Process pressure / max. overpressure limit
Vacuum…100 bar
Vacuum…1450 psi -
Min. density of medium
0.5 g/cm³
(0.4 g/cm³ optional) -
Main wetted parts
316L
Alloy -
Process connection
Threads:
G3/4A, G1A, R3/4″, R1, NPT3/4, NPT1
Flanges:
DN25…DN100
ASME 1″…4″
JIS 25A…100A -
Process connection hygienic
Tri-Clamp ISO2852
-
Sensor length
Up to 6 m (19.7 ft)
-
Communication
2-wire direct load
HART
DC-PNP 3-wire
Relays
NAMUR
PFM
Bluetooth® wireless technology -
Certificates / Approvals
ATEX, IEC Ex, CSA, NEPSI, KC, JPN Ex, INMETRO, UK Ex
-
Safety approvals
Overfill protection WHG
SIL2/ SIL 3 according to IEC 61508 -
Design approvals
EN 10204-3.1
NACE MR0175, MR0103
ASME B31.3 Process Piping
AD2000
CRN
AD 2000 -
Marine approval
ABS/ LR/ BV/ DNV
-
Options
Heavy duty stainless steel housing
-
Components
FTL325P Interface PFM
FTL325N Interface NAMUR
-
Measuring principle
Vibration Density
-
Product Headline
Vibronic point level switch with optional extension tube for liquid applications in hazardous areas .
Reliable measurement values not affected by: changing media properties, gas bubbles, foam, build-up, flow or turbulences.
Micropilot FMR62B – 80 GHz radar sensor
Smart Safety for your process
-
Measuring principle
Level radar
-
Characteristic / Application
Premium device for continuous non-contact
level measurements, where aggressive media
are used; For vessels with many internals -
Specialities
Heartbeat Technology;
SIL 2/3 according to IEC 61508;
Bluetooth® commissioning;
Operation and maintenance with SmartBlue App;
HistoROM;
RFID TAG for easy identification;
Very narrow beam angle -
Supply / Communication
2-wire (HART/ PROFIBUS PA)
PROFINET with Ethernet-APL
Bluetooth® wireless technology -
Frequency
W-Band (~80 GHz)
-
Accuracy
+/-1 mm (0.04 in)
-
Ambient temperature
-40 … +85 °C (-40 … +185 °F)
-
Process temperature
-196 … +450 °C (-321 … +842 °F)
-
Process pressure / max. overpressure limit
Vacuum…160 bar (2320 psi)
-
Main wetted parts
316L or PTFE (clad 316L flanges)
-
Process connection
Flange:
DN50 … DN150
ASME 2″ … 8″ -
Max. measurement distance
80 m (262 ft)
-
Communication
HART/ PROFIBUS PA
PROFINET with Ethernet-APL
Bluetooth® wireless
technology and app (optional) -
Certificates / Approvals
IEC Ex, ATEX, CSA C/US, EAC Ex, JPN Ex, KC Ex,
INMETRO Ex, NEPSI Ex, UK Ex -
Safety approvals
Overfill protection WHG;
SIL 2/ SIL 3 -
Design approvals
Acceptance test certificate 3.1, EN10204;
NACE MR0175 / ISO 15156;
NACE MR0103 / ISO 17945;
EU Food Contact Material (EC) 1935/2004;
US Food Contact Material FDA CFR 21;
Declaration of Conformity ASME B31.3;
Conformity to cGMP;
AD2000;
PMI test;
CRN approval;
Helium leak test;
Liquid penetrant test AD2000-HP5-3(PT);
Liquid penetrant testing ISO23277-1 (PT);
Liquid penetrant testing ASME VIII-1 (PT);
Welding documentation (according to ISO or ASME) -
Marine approval
ABS, LR, BV, RINA, DNV
-
Options
Display,
Customized settings,
App ‘SmartBlue’ for commissioning and operation via Bluetooth
Gas-tight implementation;
Overvoltage protection;
PWIS free -
Application limits
Strong turbulences and foam,
long stilling wells (>12 m) and bypass -> FMR5x,FMP5x
Field of application
Micropilot FMR62B is part of the new generation of 80 GHz radar sensors. It is used for continuous non-contact level measurement in challenging liquid applications. It masters process requirements like high temperatures, high process pressure as well as changing media. Testing concepts for SIL-locking and proof tests reduces the risk of systematic errors. Heartbeat Technology offers verification and monitoring functions to identify anomalies for increased safety and productivity.
The Micropilot FMR62B is suitable for free-space applications up to 80m and offers high beam focusing for vessels with many internal fittings. Process connections: FlangesTemperature: –196 to +450 °C (-320 to +842°F)Pressure: –1 bis +160 bar (-14.5 to 2320.6 psi)Maximum measuring range: 80 m (262.5 ft)Accuracy: ±1 mm (0.04 in)W-band: 80 GHzInternational explosion protection certificates, overfill prevention WHG, SIL, marine approval, 5- or 10-point factory calibration certificate
Benefits
Increased productivity thanks to process monitoring, verification, and diagnostics in the running process with Heartbeat Technology Safety by design: Device development according to IEC 61508 for SIL 2 and SIL 3 applications (e.g. in homogeneous redundancy) Reduction of systematic errors through guided SIL-locking, verification and proof testing Easy handling through intuitive operation and wizards, for commissioning and verifications Easier troubleshooting through recommended actions shown on the display
Digital pH sensor Memosens CPS11E
Memosens 2.0 pH electrode for standard applications in process and water & wastewater industries
-
Measuring principle
Potentiometric
-
Application
Long-term monitoring and limit control in processes with stable process conditions
Water and wastewater treatment -
Characteristic
Digital pH electrode for standard applications in process and environmental technology with dirt-repellent PTFE ring junction and built-in temperature sensor
-
Measurement range
Application A
• pH: 1 to 12
Application B
• pH: 0 to 14
Application F
• pH: 0 to 10 -
Measuring principle
Gel compact electrode with PTFE ring junction
-
Design
All shaft lengths with temperature sensor
Advanced gel technology -
Material
Sensor shaft: Glass to suit process
pH membrane glass: Type A, B, F
Metal lead: Ag/AgCl
Aperture: Ring-shaped PTFE junction, sterilizable
O-ring: FKM
Process coupling: PPS fibre-glass reinforced
Nameplate: ceramic metal oxide -
Dimension
Diameter: 12 mm (0.47 inch)
Shaft length: 120, 225, 360 and 425 mm
(4.72, 8.86, 14.17 and 16.73 inch) -
Process temperature
Application A: –15 to 80 °C (5 to 176 °F)
Application B: 0 to 135 °C (32 to 275 °F)
Application F: 0 to 70 °C (32 to 158 °F) -
Process pressure
Applications A and B:
0.8 to 17 bar (11.6 to 246.5 psi) absolute
Application F:
0.8 to 7 bar (11.6 to 101.5 psi) absolute -
Temperature sensor
NTC 30K
-
Ex certification
With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in
hazardous areas Zone 0, Zone 1 and Zone 2 -
Connection
Inductive, digital connection head with Memosens 2.0 technology
-
Ingress protection
IP68
-
Additional certifications
Additional certifications
Field of application
Memosens CPS11E is the digital all-rounder for stable processes. It measures reliably even in extreme pH ranges or hazardous areas. Thanks to Memosens 2.0 digital technology, the pH sensor resists moisture and enables lab calibration, increasing process safety and simplifying operation. Its extended storage of calibration and process data provides the perfect basis for predictive maintenance. Designed for a long operating life and low maintenance, Memosens CPS11E offers you best value for money.
Long-term monitoring and limit monitoring of processes with stable process conditions: – Chemical industry- Pulp & paper industry- Power plants (e.g. flue gas washers, boiler feed water)- Incineration plantsWater & wastewater treatment:- Drinking water- Cooling water- Well water- Wastewater treatment plantsWith ATEX, IECEx, CSA C/US, NEPSI, Japan, INMETRO approvals for use in hazardous areas, zones 0, 1 and 2.
Benefits
IIoT ready: Memosens 2.0 offers extended storage of calibration and process data, enabling better trend identification and providing a future-proof basis for predictive maintenance and enhanced IIoT services. Low operating costs: Lab calibration and quick sensor exchange in the process result in minimized process downtime and longer sensor lifetime. Robust, low maintenance electrode: Long poison diffusion path or optimized ion trap prevent poisoning of the electrode reference. Large, dirt-repellent PTFE junction protects from soiling by the medium. Process glass is suitable for the full pH range and pressure-stable up to 17 bar (246.5 psi) absolute. Optional F-glass allows application in media containing hydrofluoric acid such as etching baths in semiconductor production. Improved optional salt storage ensures reliable measurement in low conductivity applications such as boiler feed water. Maximum process integrity through non-contact, inductive signal transmission.
Flow assembly Flowfit CPA25
Universal assembly for the water and wastewater industry and industrial auxiliary processes.
-
Measuring principle
Potentiometric
-
Application
Water, Wastewater, Utilities
-
Installation
Flow assembly
-
Characteristic
Piping, bypass, branch line
-
Design
‘- Three PG13.5 sensor ports (shaft length 120 mm)
– calibration direct in the holder possible with a removable calibration cap -
Material
Holder: Polypropylene PP
Sealing: EPDM or FKM available -
Process temperature
0 to 80°C
(32 to 176°F) -
Process pressure
max. 6 bar at 20°C
(87 psi at 68°F)
-
Measuring principle
Amperometric oxygen measurement
-
Application
Water, Wastewater, Utilities
-
Installation
Flow assembly
-
Characteristic
Piping, bypass, branch line
-
Design
‘- Three PG13.5 sensor ports (shaft length 120 mm)
– calibration direct in the holder possible with a removable calibration cap -
Material
Holder: Polypropylene PP
Sealing: EPDM or FKM available -
Process temperature
0 to 80°C
(32 to 176°F) -
Process pressure
max. 6 bar at 20°C
(87 psi at 68°F)
-
Measuring principle
Conductive
-
Application
Water, Wastewater, Utilities
-
Installation
Flow assembly
-
Characteristic
Piping, bypass, branch line
-
Design
‘- Three PG13.5 sensor ports (shaft length 120 mm)
– calibration direct in the holder possible with a removable calibration cap -
Material
Holder: Polypropylene PP
Sealing: EPDM or FKM available -
Process temperature
0 to 80°C
(32 to 176°F) -
Process pressure
max. 6 bar at 20°C
(87 psi at 68°F)
4-channel transmitter Liquiline CM444
Expandable multiparameter field device for all industries
-
Measuring principle
Potentiometric
-
Application
Modular 4 wire controller
Multichannel: 1 to 4 digital channels for Memosens protocol
Multiparameter: pH, redox, ISFET, conductivity inductive and conductive, dissolved oxygen, nitrate and turbidity
4 sensors connectable in any combination
Functionality modular extendable at any time
Plug and play for modules and sensors -
Characteristic
Robust plastic housing
All non-ex applications -
Design
2 to 4x Memosens input, 2 to 8x current output
Alarmrelay, 4x relay, SD card for software update and copy & paste of setup -
Material
Housing base: PC-FR
Housing seal: EPDM -
Dimension
237 x 194 x 162 mm
9.33 x 7.64 x 6.38 inch -
Process temperature
-20 to 55 °C (0 to 130 °F)
-
Ingress protection
IP66 / IP 67
-
Input
1 to 4x Memosens digital input
2x 0/4 to 20mA Input optional
2 to 4x Digital input optional -
Output / communication
2 to 8x 0/4 to 20 mA current outputs, alarmrelay,
4x relay, ProfibusDP, Modbus RS485, Modbus TCP, Ethernet -
Additional certifications
EAC, cCSAus
-
Measuring principle
Sensor ORP / Redox
-
Application
Modular 4 wire controller
Multichannel: 1 to 4 digital channels for Memosens protocol
Multiparameter: pH, redox, ISFET, conductivity inductive and conductive, dissolved oxygen, nitrate and turbidity
4 sensors connectable in any combination
Functionality modular extendable at any time
Plug and play for modules and sensors -
Characteristic
Robust plastic housing
All non-ex applications -
Design
2 to 4x Memosens input, 2 to 8x current output
Alarmrelay, 4x relay, SD card for software update and copy & paste of setup -
Material
Housing base: PC-FR
Housing seal: EPDM -
Dimension
237 x 194 x 162 mm
9.33 x 7.64 x 6.38 inch -
Process temperature
-20 to 55 °C (0 to 130 °F)
-
Ingress protection
IP66 / IP 67
-
Input
1 to 4x Memosens digital input
2x 0/4 to 20mA Input optional
2 to 4x Digital input optional -
Output / communication
2 to 8x 0/4 to 20 mA current outputs, alarmrelay,
4x relay, ProfibusDP, Modbus RS485, Modbus TCP, Ethernet -
Additional certifications
EAC, cCSAus
-
Measuring principle
Conductive
-
Application
Modular 4 wire controller
Multichannel: 1 to 4 digital channels for Memosens protocol
Multiparameter: pH, redox, ISFET, conductivity inductive and conductive, dissolved oxygen, nitrate and turbidity
4 sensors connectable in any combination
Functionality modular extendable at any time
Plug and play for modules and sensors -
Characteristic
Robust plastic housing
All non-ex applications -