Measuring principle
TDLAS
-
Product Headline
An economical gas analyzer for moisture (H2O) in the natural gas industry. The sensor measures gas using tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the gas without coming into physical contact with the stream. Option of simplified sample conditioning to reduce complexity of the analyzer system.
-
Channels
1
-
Analyte and Measurement ranges
H2O (Moisture):
5 to 422 ppmv (minimum)
5 to 2110 ppmv (maximum) -
Measured Variables
Concentration
Dew Point
Cell Pressure
Cell Temperature -
Ambient Temperature range
-20 to 50°C (-4 to 122°F)
-
Operating Pressure range
Inlet Pressure: 140 to 350 kPa (20 to 50 psig)
Sample Cell: 700 to 1400 mbara -
Analyzer Wetted materials
316L stainless steel
FKM O-Rings
Glass -
Power supply
100 to 240 VAC, 50 to 60 Hz
or
9 to 16 VDC or 18 to 32 VDC – optional
1 Amp maximum at 120 VAC
1.6 Amps at 24VDC, 3.2 Amps at 12 VDC -
Communication
Analog Output: 1 or 2 4-20mA Isolated, 1200 ohms at 24 VDC max load
Serial: RS232C
Protocol: Modbus Gould RTU or Daniel RTU or ASCII
Alarms: 2, General Fault and Concentration Alarms via Modbus and Analog Output(s) -
Housing materials
Electronics: 304 stainless steel (Class I Division 2)
Sample System Panel: anodized aluminum -
Hazardous area approvals
CSA Class I, Division 2
CSA Class I, Zone 2 -
Degree of protection
Type 3R (Class I Division 2)
Field of application
Using patented tunable diode laser spectroscopy (TDLAS) technology, the SS500 analyzer is an economical choice for measuring H2O concentration in natural gas without coming into physical contact with the stream. The result is a sensor that does not suffer from contamination or drift due to vapor impurities such as glycol, methanol, or amines.
Natural gas quality measurements in production, storage, transportation, and distribution Monitoring for natural gas pipelines, processing plants, compressor stations, gathering sites, and distribution quality Suitable for use as a check-measurement deviceMeasurable ranges of up to 0-250 lb/mmscf (0-5275 ppm) H2O
Benefits
Economical real-time moisture measurements Virtually maintenance-free Reliable in harsh environmentsFast and accurate with no wet-up or dry-down delays No contamination or drift due to vapor impurities such as glycol, methanol, amines, hydrogen sulfide, or mercaptansNIST-traceable calibration with analog and serial outputs for remote monitoringCSA certified for CSA Class 1, Division 2 or Class 1, Division 1