  • Measuring principle

    Inductive

  • Application

    Food and beverage industries

  • Characteristic

    Specifically intended for use in beverage industry applications and especially suitable for:
    Phase separation, control of CIP, concentration control, product monitoring, leakage monitoring

  • Measurement range

    200 µS/cm to 1000 mS/cm
    Cell constant k: 11.0 1/cm

  • Measuring principle

    Inductive conductivity measurement

  • Design

    Hygienic design with FDA certified virgin PEEK

  • Material

    Sensor: PEEK
    Process connection 1.4435 (316L)
    Seal: EPDM

  • Process temperature

    -10 °C to 110 °C (14 °F to 230 °F)
    Sterilization: max. 130 °C at 6 bar abs up to 60 min
    (Max. 266 °F at 87 psi up to 60 min)

  • Process pressure

    13 bar abs up to 50 °C (188.5 psi up to 122 °F)
    7.75 bar abs at 110 °C (112 psi at 230 °F)
    6.0 bar abs at 130 °C for max. 60 minutes
    (87 psi at 266 °F for max. 60 minutes)

  • Temperature sensor

    Pt 1000 (class B)

  • Connection

    Dairy pipe, aseptic-fitting DN40 and DN50, Clamp 2″
    ISO 2852, SMS-2″, Varivent N DN 40-125
    Thread G1 1/2 SS, G1 1/2 PVC , Union nut 2 1/4 Inch PVC

  • Ingress protection

    IP69

  • Output / communication

    0/4…20mA
    IO-Link (option)

  • Additional certifications

    3A, FDA, EHEDG
    EC Regulation No. 1935/2004

Field of application
Smartec CLD18 is the cost-effective solution for all applications in beverage plants where small tubes are used and where the toroidal measurement is used as a switch function. The combination of a transmitter and a sensor is interference-free, easy-to-use and features a hygienic design that protects your products and processes from contamination. Smartec CLD18 ensures the Overall Equipment Effectiveness (OEE) of your plant thanks to fast and reliable CIP and optional IO-Link protocol.

The hygienic Smartec CLD18 measures toroidal conductivity in beverage applications:Control of CIP processes in the return channel Concentration control in remaking of CIP cleaning agentsProduct monitoring in pipe systems, bottling plantsBeer – Push water interphaseBeer – Water cut for brewing linesLeakage monitoringIt features the following protocols and interfaces:0/4…20mAIO-Link (optional)

Benefits
Unique hygienic design avoids contamination3-A approvalCompliant with EG 2023/2006 and 1935/2004Suitable for cleaning-in-place (CIP)Smart and easy integration into your plant infrastructure thanks to optional IO-Link protocolCompact design permits installation in small tubes without restricting the flow