Digital ORP sensor Memosens CPS12E
Memosens 2.0 ORP electrode for standard applications in process and water & wastewater industries
-
Measuring principle
Sensor ORP / Redox
-
Application
Long-term monitoring and limit control in processes with stable process conditions
Water treatment -
Characteristic
Digital ORP sensor for standard applications in process
technology and environmental engineering with dirt-repellent PTFE junction and integrated temperature sensor -
Measurement range
-1500 mV to +1500 mV
-
Measuring principle
Gel compact electrode with teflon junction and double gel reference
Measuring part as gold-pin or platinum ring -
Design
All shaft lengths with temperature sensor
Advanced gel technology -
Material
Sensor shaft: Glass to suit process
ORP measuring element: Platinum or gold
Metal lead: Ag/AgCl
Aperture: Ring-shaped PTFE diaphragm, sterilizable
O-ring: FKM
Process coupling: PPS fibre-glass reinforced
Nameplate: ceramic metal oxide -
Dimension
Diameter: 12 mm (0.47 inch)
Shaft lengths: 120, 225 and 360 mm
(4.72, 8.86 and 14.17 inch) -
Process temperature
-15 to 135 °C
(5 to 275 °F) -
Process pressure
0.8 to 17 bar (11.6 to 246.5 psi) absolute
-
Temperature sensor
NTC 30K
-
Ex certification
With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in
hazardous areas Zone 0, Zone 1 and Zone 2 -
Connection
Inductive, digital connection head with Memosens 2.0 technology
-
Ingress protection
IP68
-
Additional certifications
Additional certifications
Field of application
Memosens CPS12E is the digital all-round ORP sensor. It measures reliably even in harsh applications or hazardous areas. Designed for low maintenance and a long operating life, the ORP sensor offers you best value for money. Memosens CPS12E features Memosens 2.0 digital technology that provides the perfect basis for predictive maintenance thanks to extended storage of calibration and process data. The plug & play principle of Memosens allows for easy operation and more process uptime.
Long-term monitoring and limit monitoring of processes with stable process conditions: – Chemical industry- Pulp & paper industry- Power plants- Incineration plants Water & wastewater treatment:- Drinking water- Cooling water- Well water- Wastewater treatment plants
With ATEX, IECEx, CSA C/US, NEPSI, Japan, INMETRO approvals for use in hazardous areas, zones 0, 1 and 2.
Benefits
Robust, low maintenance ORP sensor: Large, dirt-repellent PTFE junction protects from soiling by the medium. Long poison diffusion path prevents poisoning of the electrode reference. Suitable for harsh applications: Gold pin or platinum cap cover the full range of oxidizing or reducing media. Shaft glass ensures high chemical stability. The sensor is pressure stable up to 17 bar (246.5 psi) absolute. Memosens 2.0 technology enables the sensor to store more process and calibration data. It lays the basis for IIoT services and predictive maintenance. Extremely reliable measurement with to non-contact, inductive signal transmission results in increased process safety. Lab calibration and fast sensor replacement on site allow for minimized process downtimes, reducing operating costs.
Digital pH sensor Memosens CPL51E
Robust Memosens 2.0 pH sensor for laboratory measurement and random sampling in the field
-
Measuring principle
Potentiometric
-
Application
Measurements in drinking water, wastewater and process water
-
Characteristic
Digital Memosens 2.0 pH sensor for laboratory measurements and
random sampling in the field
Robust pH sensor with plastic shaft -
Measurement range
pH 0 to 14 (1 to 12 application range)
-
Material
Glass and polymer
-
Dimension
Diameter: 12 mm (0.47 inch)
Shaft length: 120 mm (4.72 inch) -
Process temperature
-5 to 80 °C (23 to 176 °F)
-
Process pressure
1 bar, not intended for continues measurment in process
-
Temperature sensor
NTC 30kΩ
-
Connection
Inductive, digital Connection head with Memosens technology
-
Ingress protection
IP68
Field of application
With its robust plastic shaft and contactless inductive Memosens cable connection, Memosens CPL51E is made for random sampling directly in the field or in the laboratory. The sensor comes pre-calibrated and is directly ready for use. Ideally to be used in combination with the mobile handheld Liquiline Mobile CML18 without any additional settings.
Memosens CPL51E is suited for the following applications in drinking water, wastewater and process water: Check of on-line measuring loopsAnalysis of grab samples (on-site as well as in the laboratory)pH measurement at process points where no online measurement is installed
Benefits
The inductive Memosens sensor connection is insensitive to moisture and corrosion. Suitable for use in the laboratory or in the field to check your running process. A physical protection keeps the glass membrane safe in both environments. Automatic temperature compensation thanks to integrated temperature sensor. Fast response time even in cold ionweak water. Reference system with open aperture minimizes the risk of clogging. Supports the requirements of quality management with tamper-proof stored information such as serial number or calibration history
Manual or automatic retractable assembly Cleanfit CPA875
Sterilizable assembly for the life sciences and food & beverage industries
-
Measuring principle
Potentiometric
-
Application
Food and beverages, biotechnology, life sciences, special chemicals
-
Installation
Retractable process assembly
-
Characteristic
Open and closed tanks, piping
-
Design
Sterile and hygienic stile
Manual or pneumatic -
Material
Holder: stainless steel 1.4435
Seals: EPDM-FDA, FKM-FDA, FFKM-FDA -
Dimension
Immersion depth: 16.1 to 81.4 mm (0.63 to 3.20 inch), depending on process adaption
-
Process temperature
-10 to 140 °C (14 to 284 °F)
-
Process pressure
Pneumatic drive: 16 bar up to 140 °C (232 psi up to 284 °F)
Manual drive: 8 bar up to 140 °C (116 psi up to 284 °F) -
Connection
Aseptic DN25 clampable, Aseptic DN50 screw-in, Clamp ISO 2852, ASME BPE, Varivent N
-
Measuring principle
Sensor ORP / Redox
-
Application
Food and beverages, biotechnology, life sciences, special chemicals
-
Installation
Retractable process assembly
-
Characteristic
Open and closed tanks, piping
-
Design
Sterile and hygienic stile
Manual or pneumatic -
Material
Holder: stainless steel 1.4435
Seals: EPDM-FDA, FKM-FDA, FFKM-FDA -
Dimension
Immersion depth: 16.1 to 81.4 mm (0.63 to 3.20 inch), depending on process adaption
-
Process temperature
-10 to 140 °C (14 to 284 °F)
-
Process pressure
Pneumatic drive: 16 bar up to 140 °C (232 psi up to 284 °F)
Manual drive: 8 bar up to 140 °C (116 psi up to 284 °F) -
Connection
Aseptic DN25 clampable, Aseptic DN50 screw-in, Clamp ISO 2852, ASME BPE, Varivent N
Field of application
Completely sterilizable, the Cleanfit CPA875 retractable assembly guarantees maximum protection from contamination making it ideal for use in life science and food & beverage applications. Fully certified (EHEDG and ASME BPE) and made of FDA-recommended materials, the CPA875 is the right choice where hygiene counts!
Cleanfit CPA875 is a retractable assembly for pH, ORP, oxygen and NIR sensors with a diameter of 12mm and 225/360mm length.
It has been developed for highest safety in:
Food & beverages
Biotechnology
Life sciences
Special chemicals
Benefits
Highest process safety: The patented design of the dynamic sealing preventscontamination. Simple operation: Remove, sterilize and calibrate the sensor while the process is running – aseptic and hygienically safe.No cross-contamination: Sterile design according to EHEDG and ASME BPE as well as sealings according to FDA and USP class VI.Flexible: Individual adaptation to your process requirements based on variety of process connections.Optional double service chamber: Sealing between sterile and non-sterile part moves dynamically with the sensor guide – just like in a syringe. The sterile space and the process medium remain untouched.
Manual or automatic retractable assembly Cleanfit CPA871
Flexible process assembly for the water & wastewater and chemical industries
-
Measuring principle
Potentiometric
-
Application
Water and wastewater including sea water, chemical industry, oil and gas, electricity and energy, hazardous areas, primaries and metals
-
Installation
Retractable process assembly
-
Characteristic
Open and closed tanks, piping
-
Design
Robust and safe style
Manual or pneumatic -
Material
Seals: EPDM, FPM (Viton) or FFKM
Immersion tube, process connection, service chamber: Stainless steel 1.4404 Ra < 0.76, PEEK,
Alloy C22 Ra <0.76, PVDF, conductive PVDF -
Dimension
Immersion depth: 32.2 to 188.6 mm (1.27 to 7.40 inch), depending on process adaption
-
Process temperature
-10 to 140 °C (14 to 284 °F) for all materials except PVDF and conductive PVDF
-10 to 100 / 90 °C (14 to 212 / 194 °F) for PVDF and conductive PVDF materials -
Process pressure
Stainless steel, Alloy C22, PEEK:
16 bar up to 140 °C (232 psi up to 284 °F)
PVDF, conductive PVDF:
Basic version: 16 bar up to 100 °C (232 psi up to 212 °F)
Immersion chamber version:
4 bar to 90 °C (58 psi up to 194 °F) -
Connection
Clamp 2″, ISO2852, ASME BPE-2012, Clamp 2½”, Flange DN 40, DN50, DIN65, DN80, EN1092-1, ASME B16.5, 10K50, JIS B2220, 10K80, Thread NPT 1½”, Thread ISO 228 G1¼
-
Measuring principle
Sensor ORP / Redox
-
Application
Water and wastewater including sea water, chemical industry, oil and gas, electricity and energy, hazardous areas, mining, minerals & metals
-
Installation
Retractable process assembly
-
Characteristic
Open and closed tanks, piping
-
Design
Robust and safe style
Manual or pneumatic -
Material
Seals: EPDM, FPM (Viton) or FFKM
Immersion tube, process connection, service chamber: Stainless steel 1.4404 Ra < 0.76, PEEK,
Alloy C22 Ra <0.76, PVDF, conductive PVDF -
Dimension
Immersion depth: 32.2 to 188.6 mm (1.27 to 7.40 inch), depending on process adaption
-
Process temperature
-10 to 140 °C (14 to 284 °F) for all materials except PVDF and conductive PVDF
-10 to 100 / 90 °C (14 to 212 / 194 °F) for PVDF and conductive PVDF materials -
Process pressure
Stainless steel, Alloy C22, PEEK:
16 bar up to 140 °C (232 psi up to 284 °F)
PVDF, conductive PVDF:
Basic version: 16 bar up to 100 °C (232 psi up to 212 °F)
Immersion chamber version:
4 bar to 90 °C (58 psi up to 194 °F) -
Connection
Clamp 2″, ISO2852, ASME BPE-2012, Clamp 2½”, Flange DN 40, DN50, DIN65, DN80, EN1092-1, ASME B16.5, 10K50, JIS B2220, 10K80, Thread NPT 1½”, Thread ISO 228 G1¼
-
Measuring principle
Amperometric oxygen measurement
-
Application
Water and wastewater including sea water, chemical industry, oil and gas, electricity and energy, hazardous areas, mining, minerals & metals
-
Installation
Retractable process assembly
-
Characteristic
Open and closed tanks, piping
-
Design
Robust and safe style
Manual or pneumatic -
Material
Seals: EPDM, FPM (Viton) or FFKM
Immersion tube, process connection, service chamber: Stainless steel 1.4404 Ra < 0.76, PEEK,
Alloy C22 Ra <0.76, PVDF, conductive PVDF -
Dimension
Immersion depth: 32.2 to 188.6 mm (1.27 to 7.40 inch), depending on process adaption
-
Process temperature
Digital oxygen sensor Oxymax COS61D
Memosens optical oxygen sensor for water, wastewater and utilities
-
Measuring principle
Optical oxygen measurement
-
Application
Aeration tank, river monitoring, water treatment, fish farming
-
Characteristic
Digital, optical measurement of dissolved oxygen based on fluorescence quenching
Measurement possible in still water -
Measurement range
0 to 20 mg/l
0 to 200 %SAT
0 to 400 hPa -
Measuring principle
Oxygen-sensitive molecules (marker) are integrated in an optical active layer (fluorescence layer). The fluorescence layer surface is in contact with the medium. The sensor optics are directed at the back of the fluorescence layer. The sensor optics transmit green light pulses to the fluorescence layer. The markers respond (fluoresce) with red light pulses. The duration and intensity of the response signals depend directly on the oxygen contents or partial pressure.
-
Design
Calibration data saved in sensor
High degree of EMC protection -
Material
Sensor shaft: stainless steel 1.4435
Membrane cap: POM -
Dimension
Diameter: 40 mm (1.57 inch)
Shaft length: 186 mm (7.32 inch) -
Process temperature
-5 to 60 °C
(20 to 140 °F) -
Process pressure
Max. 10 bar abs
(Max. 145 psi) -
Temperature sensor
NTC 30K
-
Connection
Process connection: G1, NPT 3/4″
Cable connection: fixed cable or TOP68 plug-in-head
Field of application
Oxymax COS61D is a high-performance, digital oxygen sensor that offers fast, accurate and drift-free measurement. It supports your process with low maintenance, high availability and easy handling. The sensor’s long-term stable fluorescence layer is exclusively oxygen-selective (interference-free), ensuring consistently reliable measurement. Thanks to Memosens digital technology, the Oxymax COS61D combines maximum process and data integrity, and facilitates simple lab calibration.
Oxymax COS61D measures the dissolved oxygen concentration in:
Wastewater treatment plants:- O2 control in the aeration basin- Process water treatment and monitoring
Water works:- Status monitoring of drinking water (oxygen enrichment, corrosion protection, etc.)- Water quality monitoring in rivers, lakes or seas
Utilities of all industries:- O2 control in the biological treatment- Process water treatment and monitoring
Fish farming:- O2 control for optimum growth conditions
Benefits
Minimum maintenance, maximum availability Fast, drift-free measurement for precise aeration control and process monitoring Long-term stability for increased process safetyOutstanding performance in all aeration processes (SBR, Anamox, etc.)Chemical-free: no electrolyte handling
Digital oxygen sensor Memosens COS51E
Memosens 2.0 amperometric oxygen sensor for the wastewater industry and utilities
-
Measuring principle
Amperometric oxygen measurement
-
Application
Typical applications are:
– Wastewater treatment plants: oxygen control in aeration basin, process water treatment and monitoring
.
– Water plants: status monitoring of drinking water, water quality monitoring in rivers, lakes or seas
.
– Utilities of all industries: Oxygen control in biological treatment, process water treatment and monitoring. -
Characteristic
Digital amperometric 3-electrode longterm-stable sensor to measure dissolved oxygen
-
Measurement range
0.01 to 100 mg/l
0.00 to 1000 %SAT
0 to 2000 hPa -
Measuring principle
Measuring principle
-
Design
Design
-
Material
Sensor shaft: POM
Membrane cap: POM
Working electrode: Gold
Counter and reference electrode: silver/silver halide
Membrane: ETFE (COS51-TN), FEP (COS51-TF) -
Dimension
Diameter: 40mm (1.57 inch)
Shaft length: 146 mm (5.74 inch) -
Process temperature
-5 to 60 °C
(20 to 140 °F) -
Process pressure
Max. 5 bar abs
(Max. 72.5 psi abs) -
Temperature sensor
NTC 30k
-
Ex certification
With ATEX, IECEx, CSA C/US, NEPSI, JapanEx and INMETRO approval for use in Zone 0, Zone 1 and Zone 2 hazardous areas. With CSA
C/US approval also in Class I Division 1 hazardous areas in the
Gas Ex area. Additionally suitable for Class I Division 2. -
Connection
Inductive, contactless connection head with Memosens 2.0 technology
-
Ingress protection
IP 68
Field of application
Memosens COS51E is perfectly suited for all wastewater and utility applications including hazardous areas. The sensor is highly accurate and moisture-resistant ensuring the safety of your processes. COS51E features Memosens 2.0 digital technology. It is able to store more calibration and process data and thus provides the perfect basis for predictive maintenance and IIoT services. Designed to be low-maintenance with a long operating life, the sensor offers outstanding value for money.
Memosens COS51E measures the dissolved oxygen concentration in:
Wastewater treatment plants:- O2 control in the aeration basin- Process water treatment and monitoring
Utilities of all industries:- O2 control in the biological treatment- Process water treatment and monitoring Water works:- Status monitoring of drinking water (oxygen enrichment, corrosion protection, etc.)- Water quality monitoring in rivers, lakes or seas
The sensor is suitable for hazardous area use (ATEX, FM, CSA).
Benefits
Wide measuring range: The amperometric measuring principle and cathode design make the sensor suitable for all wastewater applications from industrial to municipal plants. High accuracy: Thanks to its amperometric, three-electrode design, the sensor provides long-term stable measurement with low drift. Increased plant availability: Memosens technology and Liquiline transmitters enable plug & play for fast commissioning and sensor exchange. Non-contact, inductive signal transmission ensures maximum process integrity.Memosens 2.0 digital technology makes the sensor ready for predictive maintenance and IIoT services since it offers extended storage of calibration and process data.
Multiparameter handheld Liquiline Mobile CML18
Mobile device for pH/ORP, conductivity, oxygen and temperature measurement in all industries
-
Measuring principle
Potentiometric
-
Application
Multiparameter mobile device for field and laboratory use
-
Characteristic
Connectable to all pH, conductivity and dissolved oxygen Memosens sensors
-
Design
Small and handy – designed to fit in your pocket
-
Material
Housing: PBT
Display: PMMA
Buttons, cap: TPE
M12 connector: CuZn, nickel-plated -
Dimension
84 x 54 x 47 mm (h x w x d)
3.31 x 2.13 x 1.85 inch (h x w x d) -
Ingress protection
IP 66
-
Input
One channel instrument with inductive Memosens coupling
M12 connector for using CYK10 and CYK20 Memosens cable -
Output / communication
Bluetooth
-
Measuring principle
Conductive
-
Application
Multiparameter mobile device for field and laboratory use
-
Characteristic
Connectable to all pH, conductivity and dissolved oxygen Memosens sensors
-
Design
Small and handy – designed to fit in your pocket
-
Material
Housing: PBT
Display: PMMA
Buttons, cap: TPE
M12 connector: CuZn, nickel-plated -
Dimension
84 x 54 x 47 mm (h x w x d)
3.31 x 2.13 x 1.85 inch (h x w x d) -
Ingress protection
IP 66
-
Input
One channel instrument with inductive Memosens coupling
M12 connector for using CYK10 and CYK20 Memosens cable
1-channel transmitter Liquiline Compact CM82
Multiparameter field device for hazardous and non-hazardous applications in all industries
-
Measuring principle
Potentiometric
-
Application
All sectors, supporting all sensors with the blue Memosens plug-in head, enviromental sector
-
Installation
Connection of digital sensors with Memosens-Protocol
-
Characteristic
Liquiline CM82 is a two-wire transmitter for all areas of process engineering
-
Design
Liquiline CM82 is optionally equipped with Bluetooth for comfortable remote commissioning via Smartphone/Tablets
-
Material
Housing, cover: Peek 151
Strain relief: EPDM
Axial ring Peek 450 G -
Dimension
Diameter: 18.95 mm (0.75 inch)
Length: 114.7 mm (4.52 inch) -
Connection
Connection of digital sensors with Memosens-Protocol
-
Ingress protection
IP67, IP68, NEMA Type 6
-
Input
One channel transmitter
-
Output / communication
4 to 20 mA, HART
-
Additional certifications
CE, CSA C/US GP
-
Measuring principle
Potentiometric
-
Application
All sectors, supporting all sensors with the blue Memosens plug-in head, enviromental sector
-
Installation
Connection of digital sensors with Memosens-Protocol
-
Characteristic
Liquiline CM82 is a two-wire transmitter for all areas of process engineering
-
1-channel transmitter Liquiline Compact CM72
Single-parameter field device for hazardous and non-hazardous applications in all industries
-
Measuring principle
Potentiometric
-
Application
All sectors, supporting all sensors with the blue Memosens plug-in head, enviromental sector
-
Installation
Connection of digital sensors with Memosens-Protocol
-
Characteristic
Liquiline CM72 is a fixed configured two-wire transmitter for all areas of process engineering
-
Measurement range
pH Sensor: 0 to 14 pH
-
Design
Peek housing
-
Material
Housing, cover: Peek 151
Strain relief: EPDM
Axial ring: Peek 450 G -
Dimension
Diameter: 18.95 mm (0.75 inch)
Length: 114.7 mm (4.52 inch) -
Connection
Connection of digital sensors with Memosens-Protocol
-
Ingress protection
IP67, IP68, NEMA Type 6
-
Input
One channel transmitter
-
Output / communication
4 to 20 mA
-
Additional certifications
CE, CSA C/US GP
-
Measuring principle
Potentiometric
-
Application
All sectors, supporting all sensors with the blue Memosens plug-in head, enviromental sector
-
Installation
Connection of digital sensors with Memosens-Protocol
-
Characteristic
Liquiline CM72 is a fixed configured two-wire transmitter for all areas of process engineering
-
Measurement range
ORP sensors: -1500 mV to 1500 mV
-
Design
Peek housing
-
Material
Housing, cover: Peek 151
Strain relief: EPDM
Axial ring: Peek 450 G -
Dimension
Diameter: 18.95 mm (0.75 inch)
Length: 114.7 mm (4.52 inch) -
Connection
Connection of digital sensors with Memosens-Protocol
-
Ingress protection
IP67, IP68, NEMA Type 6
-
Input
One channel transmitter
1-/2-channel transmitter Liquiline CM442R
Compact multiparameter DIN-rail and cabinet device for all industries
-
Measuring principle
Potentiometric
-
Application
Modular 4 wire controller
Multichannel: 1/2 digital channels for Memosens protocol
Multiparameter: pH, redox, ISFET, conductivity inductive and conductive, dissolved oxygen, nitrate and turbidity
Two sensors connectable in any combination
Functionality modular extendable at any time
Plug and play for modules and sensors -
Characteristic
Robust plastic housing
All non-ex applications -
Design
2x Memosens input, 2 to 4x current output, alarmrelay
2x relay, SD card for software update and copy & paste of setup -
Material
DIN rail housing: PC-FR
Display seal: EPDM -
Dimension
99 x 100 x 97 mm
3.9 x 3.94 x 3.82 inch -
Process temperature
0 to 60 °C (32 to 140 °F)
-
Ingress protection
Transmitter: IP20
Optional Display: IP66 -
Input
1 to 2x Memosens digital input
2x 0/4 to 20mA Input optional
2x Digital input optional -
Output / communication
2 to 8x 0/4 to 20 mA current outputs
Alarmrelay, 2x relay, ProfibusDP, Modbus RS485,
Modbus TCP, Ethernet -
Additional certifications
EAC, cCSAus
-
Measuring principle
Sensor ORP / Redox
-
Application
Modular 4 wire controller
Multichannel: 1/2 digital channels for Memosens protocol
Multiparameter: pH, redox, ISFET, conductivity inductive and conductive, dissolved oxygen, nitrate and turbidity
Two sensors connectable in any combination
Functionality modular extendable at any time
Plug and play for modules and sensors -
Characteristic
Robust plastic housing
All non-ex applications -
Design
2x Memosens input, 2 to 4x current output, alarmrelay
2x relay, SD card for software update and copy & paste of setup -
Material
DIN rail housing: PC-FR
Display seal: EPDM -
Dimension
99 x 100 x 97 mm
3.9 x 3.94 x 3.82 inch -
Process temperature
0 to 60 °C (32 to 140 °F)
-
Ingress protection
Transmitter: IP20
Optional Display: IP66 -
Input
1 to 2x Memosens digital input
2x 0/4 to 20mA Input optional
2x Digital input optional -
Output / communication
2 to 8x 0/4 to 20 mA current outputs
Alarmrelay, 2x relay, ProfibusDP, Modbus RS485,
Modbus TCP, Ethernet -
Additional certifications
EAC, cCSAus
-
Measuring principle
Conductive
-
Application
Modular 4 wire controller
Multichannel: 1/2 digital channels for Memosens protocol
Multiparameter: pH, redox, ISFET, conductivity inductive and conductive, dissolved oxygen, nitrate and turbidity
Two sensors connectable in any combination
Functionality modular extendable at any time
Plug and play for modules and sensors -
Characteristic
Robust plastic housing
All non-ex applications -
1-/2-channel transmitter Liquiline CM442
Expandable multiparameter field device for all industries
-
Measuring principle
Potentiometric
-
Application
Modular 4 wire controller
Multichannel: 1/2 digital channels for Memosens protocol
Multiparameter: pH, redox, ISFET, conductivity inductive and conductive, dissolved oxygen, nitrate and turbidity
Two sensors connectable in any combination
Functionality modular extendable at any time
Plug and play for modules and sensors -
Characteristic
Robust plastic housing
All non-ex applications -
Design
2x Memosens input, 2 to 4x current output, Alarmrelay,
2x relay, SD card for software update and copy & paste of setup -
Material
Housing base: PC-FR
Housing seal: EPDM -
Dimension
237 x 194 x 162 mm
9.33 x 7.64 x 6.38 inch -
Process temperature
-20 to 60 °C (0 to 140 °F)
-
Ingress protection
IP66 / IP67
-
Input
1 to 2x Memosens digital input
2x 0/4 to 20mA input optional
2x Digital input optional -
Output / communication
2 to 8x 0/4 to 20 mA current outputs
Alarmrelay, 2x relay, ProfibusDP, Modbus RS485, Modbus TCP, Ethernet -
Additional certifications
EAC, cCSAus, MCERTS
-
Measuring principle
Sensor ORP / Redox
-
Application
Modular 4 wire controller
Multichannel: 1/2 digital channels for Memosens protocol
Multiparameter: pH, redox, ISFET, conductivity inductive and conductive, dissolved oxygen, nitrate and turbidity
Two sensors connectable in any combination
Functionality modular extendable at any time
Plug and play for modules and sensors -
Characteristic
Robust plastic housing
All non-ex applications -
Design
2x Memosens input, 2 to 4x current output, Alarmrelay,
2x relay, SD card for software update and copy & paste of setup -
Material
Housing base: PC-FR
Housing seal: EPDM -
Dimension
237 x 194 x 162 mm
9.33 x 7.64 x 6.38 inch -
Process temperature
-20 to 60 °C (0 to 140 °F)
-
Ingress protection
IP66 / IP67
-
Input
1 to 2x Memosens digital input
-
Output / communication
2 to 8x 0/4 to 20 mA current outputs
Alarmrelay, 2x relay -
Additional certifications
EAC, cCSAus, MCERTS
-
Measuring principle
Conductive
-
Application
Modular 4 wire controller
Multichannel: 1/2 digital channels for Memosens protocol
Multiparameter: pH, redox, ISFET, conductivity inductive and conductive, dissolved oxygen, nitrate and turbidity
Two sensors connectable in any combination
Functionality modular extendable at any time
Plug and play for modules and sensors -
Characteristic
Robust plastic housing
All non-ex applications -
Design
Digital 4-electrode conductivity sensor Memosens CLS82E
Memosens 2.0 contacting conductivity sensor for hygienic applications in life sciences, food & bev
-
Measuring principle
Conductive
-
Application
Phase seperation, chromotography, fermentation
CIP monitoring in small pipes
Ultrafiltration
Final rinse -
Characteristic
Conductivity sensor with a broad measurement range for all hygienic applications with advanced requirements to process safety
-
Measurement range
1 µS/cm to 500 mS/cm
-
Measuring principle
Hygienic 4-electrode conductivity sensor
-
Design
Certified hygienic design with FDA compliant materials, 3-A and EHEDG certified, available as 120 mm sensor with PG13.5 and in many other standard process connections
-
Material
Process connection: stainless steel 1.4435 (316L)
Sensor element: platinum and ceramic -
Dimension
Sensor diameter: 12 mm (0.47 inch)
Electrode length: depending on process connection -
Process temperature
-5 to 120 °C (23 to 248 °F)
Sterilization: max. 140 °C at 6 bar for max. 45 min
(Max. 284 °F at 87 psi for max. 45 min) -
Process pressure
17 bar abs at 20 °C (247 psi at 68 °F)
9 bar abs at 120 °C (131 psi at 248 °F) -
Temperature sensor
Pt1000
-
Ex certification
ATEX, NEPSI, CSA, IECE, EAC Ex, INMETRO
-
Connection
Clamp 1,5″, Clamp 2″, Varivent F DN25,Varivent N DN40-DN125, BioControl DN25, PG13,5, DN25 standard/40mm, DN25 B.Braun port/65mm, Threat G1″, Thread NPT 1″
Sensor connection: Inductive, digital connection head with Memosens 2.0 technology -
Ingress protection
IP 68 / NEMA Type 6P
-
Additional certifications
Material certification 3.1
3-A and EHEDG certified, hygienic design
ASME BPE
Conformity to cGMP derived requirements
ASME BPE
FDA approved
Regulation (EC) No. 1935/2004
China Food
Field of application
Memosens CLS82E is a high-end, hygienic sensor for applications with widely varying conductivity values. It enables reliable monitoring of core processes as well as final rinse with only one sensor. FDA compliance and a small, compact design ensure the perfect fit to your processes. With Memosens 2.0 digital technology, CLS82E allows for extended storage of process and sensor data facilitating lab calibration and predictive maintenance and providing the perfect basis for enhanced IIoT services.
Memosens CLS82E measures conductivity in applications that need to cover a wide measuring range as in:
Phase separation
Downstream processing
FermentationChromatographyUltrafiltration
Final rinse in pharma applications
With its international explosion protection approvals, such as ATEX, FM, CSA and NEPSI, the digital sensor is suitable for hazardous area use.
Benefits
Hygienic design and materials certified and approved by EHEDG, 3-A and FDA ensure compliance with GMP requirements. The sensor is sterilizable and autoclavable. Absolute loop safety thanks to Memosens and unique detection of build-up on electrodes. The broad measuring range enables monitoring of core processes and final rinse with one sensor, saving costs. Its compact design makes the sensor suitable for small pipe diameters and narrow, space-limited installations. Quality certificate stating the individual cell constant allows precise adjustment of the measurement. Non-contact, inductive signal transmission ensures high process and data integrity. IIoT ready: Memosens 2.0 offers extended storage of calibration and process data, enabling better trend identification and providing a future-proof basis for predictive maintenance and enhanced IIoT services.
Digital conductivity sensor Memosens CLS21E
Memosens 2.0 contacting conductivity sensor for surface and drinking water monitoring
-
Measuring principle
Conductive
-
Application
Surface water + water
Waste water
Process industry
Media separation
Industrial water -
Characteristic
Digital 2-electrode conductivity sensor
-
Measurement range
k=1: 10 µS/cm to 20 mS/cm
-
Measuring principle
Conductive conductivity cell with graphite electrodes
-
Design
2-electrode conductivity cell with parallel arranged electrodes
-
Material
Sensor shaft: PES
Electrodes: graphite -
Dimension
Diameter: 24 mm (0.94 inch)
Length: 61 mm (2.40 inch) -
Process temperature
-20 to 135 °C at 3.5 bar abs
(-4 to 275 °F at 50 psi) -
Process pressure
17 bar abs at 20 °C
(246 psi at 68 °F) -
Temperature sensor
Pt1000
-
Ex certification
ATEX, CSA, NEPSI, EAC Ex, IECEx, INMETRO
-
Connection
Process connection: DN25, DN40, G1, NPT 1″
Sensor connection: Inductive, digital connection head with Memosens 2.0 technology -
Ingress protection
IP68
Field of application
Memosens CLS21E is a robust conductivity sensor for all water applications with medium conductivities. It provides you with precise and reliable measured values for optimum quality monitoring. The sensor is corrosion-free and features high mechanical stability ensuring a long operating life. Thanks to Memosens 2.0 digital technology, CLS21E combines maximum process and data integrity with simple operation. It facilitates predictive maintenance and provides the perfect basis for IIoT services.
Memosens CLS21E measures conductivity in all water applications, such as:
Potable water treatment
Desalination plants
Industrial water treatment
River/lake monitoring
With its international explosion protection approvals, such as ATEX, FM, CSA and NEPSI, the digital sensor is suitable for hazardous area use.
Benefits
Reliable and accurate measured values at medium conductivities ensure precise monitoring of water quality. Robust design for long durability and low maintenance: The sensor is corrosion-free and features high mechanical and chemical stability. Quality certificate stating the individual cell constant allows accurate measurement. Non-contact inductive signal transmission ensure maximum process and data integrity. Memosens 2.0 technology enables the sensor to store more process and calibration data. It lays the basis for IIoT services and predictive maintenance.
Digital conductivity sensor Memosens CLS16E
Memosens 2.0 contacting conductivity sensor for hygienic applications in life sciences and food
-
Measuring principle
Conductive
-
Application
Pure and ultrapure water
Pharmaceutical industry
Final rinse
WFI -
Characteristic
Hygienic 2-electrode conductivity sensor
-
Measurement range
k=0,1: 0,04 to 500 µS/cm
-
Measuring principle
Conductive conductivity cell with electropolished stainless steel electrodes
-
Design
Hygienic 2-electrode conductivity sensor with coaxially arranged electrodes, electropolished
-
Material
n contact with medium
Sensor depending on order execution:
Electropolished, stainless steel 1.4435 (AISI 316L)
PEEK
Sealing depending on order execution:
Form seal ISOLAST (FFKM)
Form seal EPDM -
Dimension
Electrode diameter: 17 mm (0.66 inch)
Electrode length: 54 mm (2.13 inch) -
Process temperature
-5 to 120 °C (23 to 248 °F)
For Sterilizsation: max. 150 °C at 5 bar
(Max. 302 °F at 73 psi) -
Process pressure
13 bar at 20 °C (188 psi at 68 °F)
9 bar at 120 °C (130 psi at 248 °F) -
Temperature sensor
Pt1000
-
Ex certification
ATEX, NEPSI, CSA, IECEx, INMETRO, EAC Ex
-
Connection
Process connection: Varivent, Neumo, BioControl, Clamp
Sensor connection: Inductive, digital connection head with Memosens 2.0 technology -
Ingress protection
IP68
-
Additional certifications
Material certification 3.1
EHEDG certified, hygienic design
CoC ASME BPE
Conformity to cGMP derived
Field of application
Memosens CLS16E is a high-end conductivity sensor with certified hygienic design conforming to FDA and USP Class VI. It measures with highest precision delivering reliable data for optimum process and product quality. The sensor is robust and offers a long operating life thanks to its replaceable seal. With Memosens 2.0 technology, CLS16E is able to store more calibration and process data, facilitating predictive maintenance and providing the perfect basis for IIoT services.
Memosens CLS16E measures conductivity in pure and ultrapure water applications for the monitoring and control of:
Water for injection (WFI)
Electro-deionizing
Ion exchangers
Reverse osmosis
Distillation
With its international explosion protection approvals, such as ATEX, FM, CSA and NEPSI, the digital sensor is suitable for hazardous area use.
Benefits
Certified with quality certificate and EN 10204 3.1 and complying with EHEDG, USP Class VI, ASME and FDA, Memosens CLS16E is perfectly suited for hygienic applications in the life sciences and food industries.The sensor is sterilizable and autoclavable, resists cleaning in place (CIP) and sterilization in place (SIP). Stainless steel ensures robust, corrosion-free operation and the replaceable seal enables a long operating life. Highest precision and measuring accuracy even at high temperatures and under pressure provide you with reliable data for optimum process and product quality. Non-contact, inductive signal transmission ensures high process integrity. IIoT ready: Memosens 2.0 offers extended storage of calibration and process data, enabling better trend identification and providing a future-proof basis for predictive maintenance and enhanced IIoT services.
Digital conductivity sensor Condumax CLS16D
Memosens conductive conductivity sensor for hygienic applications in pure and ultrapure water
-
Measuring principle
Conductive
-
Application
Pure and ultrapure water, pharmaceutical industry
-
Characteristic
Hygienic 2-electrode conductivity sensor
-
Measurement range
k=0,1: 0,04 to 500 µS/cm
-
Measuring principle
Conductivity cell
-
Design
Hygienic 2-electrode conductivity sensor with coaxially arranged electrodes, electropolished stainless steel and FDA sealing material; sterilizable
-
Material
Isolation: PEEK
Electrode: polished stainless steel 1.4435
Sealing: Isolast (FFKM) FDA conform -
Dimension
Electrode diameter: 17 mm (0.66 inch)
Electrode length: 54 mm (2.13 inch) -
Process temperature
-5 to 120 °C (23 to 248 °F)
For Sterilizsation: max. 150 °C at 5 bar
(Max. 302 °F at 73 psi) -
Process pressure
13 bar at 20 °C (188 psi at 68 °F)
9 bar at 120 °C (130 psi at 248 °F) -
Temperature sensor
NTC 30K
-
Connection
Process connection: Varivent, Neumo, Clamp
Cable: Memosens connector -
Ingress protection
IP68
-
Additional certifications
Material certification 3.1
EHEDG certified, hygienic design, CoC ASME BPE
Conformity to cGMP derived requirements
ASME BPE
FDA approved
Regulation (EC) No.
1935/2004
China Food
Field of application
Condumax CLS16D is a high-end conductivity sensor with certified hygienic design. It measures with highest precision providing you with reliable data for optimum process and product quality. Numerous certificates and various hygienic process connections ensure the perfect fit for your process. Condumax CLS16D is not the latest Memosens generation. To get information on the new Memosens CLS16E sensor with extended functionality, click here.
Condumax CLS16D measures conductivity in pure and ultrapure water applications for the monitoring and control of:
Ion exchangers
Reverse osmosis
Distillation
Electro-deionizing
Water for injection (WFI)
With its international explosion protection approvals, such as ATEX, FM, CSA and NEPSI, the digital sensor is suitable for hazardous area use.
Benefits
Highest precision and measuring accuracyHygienic design certified by EHEDGCertified biocompatibility according to USP class VI and FDA compliance of all wetted materialsSterilizable and autoclavableQuality certificate stating the individual cell constantLogging of sensor-specific data for easy traceability and predictive maintenanceConvenient maintenance strategy with Memobase Plus for easy calibration and sensor management
Digital conductivity sensor Memosens CLS15E
Memosens 2.0 contacting conductivity sensor for standard applications in pure and ultrapure water
-
Measuring principle
Conductive
-
Application
Measurement in pure and ultrapure water range
Monitoring of ion exchangers
Reverse osmosis
Distillation and chip cleaning -
Characteristic
Digital 2-electrode conductivity sensor
-
Measurement range
k=0,01: 0.04 to 20 µS/cm
k=0,1: 0.10 to 200 µS/cm -
Measuring principle
Conductive conductivity cell with electropolished stainless steel electrodes
-
Design
2-electrode conductivity cell with coaxially arranged electrodes, electropolished
-
Material
Electrode: stainless steel 1.4435
Electrode shaft: PES -
Dimension
Electrode diameter: 16 mm (0.63 inch)
Electrode length: appr. 55 mm (2.17 inch) -
Process temperature
Threaded with fixed cable:
-20 to 100 °C (-4 to 212 °F)
Threaded with plug-in head:
-20 to 120 °C (-4 to 248 °F)
Sterilization: max. 140 °C (284 °F) for 30 minutes -
Process pressure
13 bar at 20 °C (188 psi at 68 °F) absolute
1 bar at 120 °C (14 psi at 248 °F) absolute -
Temperature sensor
Pt1000
-
Ex certification
ATEX, NEPSI, CSA, IECEx, INMETRO, EAC Ex
-
Connection
Process: 1/2″ and 3/4″ NPT, Clamp 1 1/2″
Sensor connection: Inductive, digital connection head with Memosens 2.0 technology -
Ingress protection
IP68
-
Additional certifications
Calibration certification of the cell constante and temperature, material certification 3.1
Field of application
Memosens CLS15E is perfectly suited for conductivity measurement in applications with low measuring ranges such as boiler feedwater and chip cleaning. The contacting conductivity sensor performs reliably and accurately even in hazardous areas. CLS15E features Memosens 2.0 digital technology, offering extended storage of calibration, adjustment and process data. It facilitates predictive maintenance and provides the perfect basis for IIoT services.
Memosens CLS15E measures conductivity in pure and ultrapure water applications for the monitoring and control of:
Boiler feedwater
Ion exchangers
Reverse osmosis
Cooling water
Distillation
Chip cleaning
The digital sensor can be used in power plants for:
Degassed conductivity
pH measurement with differential conductivity
With its international explosion protection approvals, such as ATEX and FM, the sensor is suitable for hazardous area use.
Benefits
Designed for low maintenance and a long operating life, the sensor offers best value for money. The sensor is sterilizable and autoclavable, allowing for installation in sterile plants. Thanks to its electrode geometry, Memosens CLS15E provides reliable and accurate measured values at low conductivities.A quality certificate stating the individual cell constant enables perfect adjustment of the measuring point. IIoT ready: Memosens 2.0 offers extended storage of calibration and process data, enabling better trend identification and providing a future-proof basis for predictive maintenance and enhanced IIoT services. Non-contact inductive signal transmission ensures maximum process safety.
Compact conductivity device Smartec CLD18
Toroidal conductivity measuring system for hygienic applications
-
Measuring principle
Inductive
-
Application
Food and beverage industries
-
Characteristic
Specifically intended for use in beverage industry applications and especially suitable for:
Phase separation, control of CIP, concentration control, product monitoring, leakage monitoring -
Measurement range
200 µS/cm to 1000 mS/cm
Cell constant k: 11.0 1/cm -
Measuring principle
Inductive conductivity measurement
-
Design
Hygienic design with FDA certified virgin PEEK
-
Material
Sensor: PEEK
Process connection 1.4435 (316L)
Seal: EPDM -
Process temperature
-10 °C to 110 °C (14 °F to 230 °F)
Sterilization: max. 130 °C at 6 bar abs up to 60 min
(Max. 266 °F at 87 psi up to 60 min) -
Process pressure
13 bar abs up to 50 °C (188.5 psi up to 122 °F)
7.75 bar abs at 110 °C (112 psi at 230 °F)
6.0 bar abs at 130 °C for max. 60 minutes
(87 psi at 266 °F for max. 60 minutes) -
Temperature sensor
Pt 1000 (class B)
-
Connection
Dairy pipe, aseptic-fitting DN40 and DN50, Clamp 2″
ISO 2852, SMS-2″, Varivent N DN 40-125
Thread G1 1/2 SS, G1 1/2 PVC , Union nut 2 1/4 Inch PVC -
Ingress protection
IP69
-
Output / communication
0/4…20mA
IO-Link (option) -
Additional certifications
3A, FDA, EHEDG
EC Regulation No. 1935/2004
Field of application
Smartec CLD18 is the cost-effective solution for all applications in beverage plants where small tubes are used and where the toroidal measurement is used as a switch function. The combination of a transmitter and a sensor is interference-free, easy-to-use and features a hygienic design that protects your products and processes from contamination. Smartec CLD18 ensures the Overall Equipment Effectiveness (OEE) of your plant thanks to fast and reliable CIP and optional IO-Link protocol.
The hygienic Smartec CLD18 measures toroidal conductivity in beverage applications:Control of CIP processes in the return channel Concentration control in remaking of CIP cleaning agentsProduct monitoring in pipe systems, bottling plantsBeer – Push water interphaseBeer – Water cut for brewing linesLeakage monitoringIt features the following protocols and interfaces:0/4…20mAIO-Link (optional)
Benefits
Unique hygienic design avoids contamination3-A approvalCompliant with EG 2023/2006 and 1935/2004Suitable for cleaning-in-place (CIP)Smart and easy integration into your plant infrastructure thanks to optional IO-Link protocolCompact design permits installation in small tubes without restricting the flow
Digital free bromine sensor Memosens CCS55D
Memosens sensor for seawater and process water
-
Measuring principle
Free bromine
-
Application
Seawater – to ensure reliable disinfection
Process water – wide range of application due to large measuring range
Pool water – efficient dosing thanks to precise measurement -
Characteristic
Amperometric measurement of dissolved free bromine
-
Measurement range
Trace: 0 … 5 mg/l HOBr
Standard: 0 … 20 mg/l HOBr
High: 0 … 200 mg/l HOBr -
Measuring principle
Closed, membrane covered measuring cell
Reduction of free bromine at the working electrode -
Design
Closed amperometric 2-electrode measuring cell with PET membrane
-
Material
Sensor shaft: POM
Membrane: PET
Membrane cap: PVDF -
Dimension
Diameter: 25 mm (0.98 in)
Length: 161 mm (6.34 in) -
Process temperature
+0…55 °C (32 … 130°F), non-freezing
-
Process pressure
Max. 1 bar (max. 14.5 psi)
-
Temperature sensor
10k NTC integrated (Memosens)
-
Connection
Inductive, digital connection head with Memosens technology
Field of application
Memosens CCS55D is a low-maintenance sensor for free bromine. It provides precision measurement in seawater for desalination plants, process or cooling water, swimming pools and fish farming industry. The bromine sensor ensures safe disinfection with bromine even at low water volumes. It features fast response times guaranteeing efficient process control. Memosens CCS55D is not the latest Memosens generation. More information on the new Memosens CCS55E sensor with extended functionality here.
The Memosens CCS55D free bromine sensor is suitable for a wide range of water qualities:Seawater applications such as desalination plants – to ensure reliable disinfection and avoid overdosingFood industry – to provide accurate control of wastewater in fish farmingProcess and cooling water – where bromine disinfectants are used due to their reduced corrosion effectPool water, therapeutic baths – where bromine is used due to the salinity of the water and as alternative to chlorine
Benefits
The right sensor version for every application: From trace measurement up to free bromine concentrations of 200 mg/l. Fast response time (t90 < 20 s) provides an accurate process view and enables prompt reaction to process changes as well as efficient process control. No inactivation in bromine-free media ensures fast measurement after addition of bromine. Increased process safety: precise and long-term stable measurement ensures consistent process monitoring and allows for individually adapted disinfectant dosing. Safe disinfection in seawater applications since disinfection with chlorine forms bromine compounds when used in seawater. Chlorine monitoring would lead to measurement errors underestimating the disinfection effect. More process up-time thanks to fast sensor exchange: pre-calibrate the sensor in your lab and then swap it into your process with plug & play. Connection to the Liquiline multiparameter transmitter allows for easy combination with other relevant parameters of liquid analysis such as pH and ORP.
Digital total chlorine sensor Memosens CCS120D
Memosens disinfection sensor for wastewater and process water
-
Measuring principle
Total chlorine
-
Application
Wastewater treatment plants, disinfection monitoring and control in wastewater outlet, reuse of outlet water, utilities of all industries, all disinfection measuring points
-
Characteristic
Amperometric measurement of total chlorine
-
Measurement range
0.1 to 10 mg/l total chlorine
-
Measuring method
Total chlorine consists of free chlorine (HOCl, OCl-) and bound chlorine (chloramines)
All components are reduced at the working electrode
Almost pH independent -
Design
Closed, membrane covered measuring cell
-
Material
Sensor shaft: PVC
Membrane: PET
Membrane cap: PPE -
Dimension
Diameter: 25 mm (0.98 inch)
Length: 148 mm (5.83 inch) -
Process temperature
5 to 45 °C
(41 to 113 °F) -
Process pressure
1 to 4 bar (14.5 to 58 psi)
-
Temperature sensor
10k NTC integrated (Memosens)
-
Connection
Inductive, digital connection head with Memosens
Field of application
Total chlorine is the mixture of free and bound chlorine components and thus a good indicator of residual disinfectants in discharge water. The sensor Memosens CCS120D provides reliable monitoring of total chlorine limits in the outlet. It reduces your maintenance effort due to easy replacement of electrolyte and membrane cap. Memosens CCS120D is not the latest Memosens generation. More information on the new sensor Memosens CCS53E.
Memosens CCS120D measures total chlorine in:Wastewater treatment plants- Disinfection monitoring and control in wastewater outlet- Reuse of outlet waterUtilities of all industries- All disinfection measuring points
Benefits
Easy handling thanks to Memosens digital technology Enables predictive maintenance due to storage of sensor and process-specific data Low maintenance thanks to fast exchange of membrane cap and electrolyte Flexible installation in CYA112 immersion assembly or CCA250 flow assembly Easy combination with other relevant parameters of liquid analysis thanks to connection to the Liquiline multichannel transmitter
Digital nitrate or SAC sensor Viomax CAS51D
Memosens sensor for optical measurement of nitrate, SAC, CODeq, TOCeq, BODeq or DOCeq
-
Measuring principle
UV photometric
-
Characteristic
Insitu (immersion in process), extractive with sample and flow cell
Optical photometric sensor for nitrate and SAC -
Size
388 / 394 mm x 40 mm, depends on version
-
Design
40 mm sensor, stainless steel 316L
-
Process temperature
5 to 50°C (41 to 120 °F)
-
Ambient temperature
-20 to 60 °C (-4 to 140 °F)
-
Specials
G1, NPT3/4
IP68 (1m/3,3ft) water column with 1N KCl during 60d ays -
Application
Drinking water, biological waste wate process, WWTP effluent
Monitoring of surface water -
Measurement range
0.01 to 20 mg/l / 0.1 to 50 mg/l NO3-N
0.04 to 80 mg/l / 0.4 to 200 mg/l NO3
0 to 50 1/m / 0 to 250 1/m / 0 to 1000 1/m SAC
0 to 75 mg/l / 0 to 375 mg/l / 0 to 1500 mg/l COD/BOD (254nm)
0 to 30 mg/l / 0 to 150 mg/l / 0 to 600 mg/l TOC/DOC (254nm)
Field of application
Viomax CAS51D is a smart sensor that allows unattended operation in a wide range of process conditions in water, wastewater and utilities. It ensures continuous online measurements without any chemicals and it is extremely robust as well as low-maintenance thanks to its self-cleaning design. You can select the CAS51D as nitrate sensor or as sensor for SAC and several sumparameters. Thanks to Memosens digital technology, Viomax CAS51D offers high process and data integrity and is easy-to-use.
Viomax CAS51D measures nitrate, SAC or several sum parameters in drinking water, process water and wastewater applications as well as in the utilities sector:
Organic load or nitrate in inlet and outlet
Determination of CODeq, TOCeq, BODeq or DOCeq
Monitoring, control and optimization of treatment processes
Nitrate monitoring in the aeration basin and denitrification stage
Benefits
Cost-effective: Easy, chemical-free handling is safe, environmentally friendly and saves on operational expenditure.Extremely fast response time: Measured value processing in the sensor provides reliable real-time process information.Fast and easy commissioning: Sensor comes precalibrated and preconfigured ex factory.Enables unattended plant operation: Intelligent design and automatic air-cleaning functionality maximize availability and minimize maintenance.Perfect adaption to open basins: Fast and flexible sensor positioning with Flexdip CYH112 holder and CYA112 assembly system.Perfect adaption to pipes: Various flow cells allow easy plant integration.
Proline Prosonic Flow W 400 ultrasonic clamp-on flowmeter
Reliable, flexible, zero maintenance – noninvasive flow measurement, consistently accurate, even in limited spaces
-
Measuring principle
Ultrasonic flow
-
Product headline
Clamp-on meter with Heartbeat Technology and web server for the water and wastewater industry.
Short inlet run thanks to FlowDC.
Bidirectional measurement for water and wastewater as well as process water and hydropower plants. -
Sensor features
Low capital investment – cost-effectiveness increases with pipe diameter (up to DN 2000/80″). Long-term stable signal – maintenance-free permanent mounting from outside with coupling pads. Reliable measurement on various pipe materials – sensor for GRP and plastic pipes available.
Mounting without process interruption. Wide nominal diameter range: DN 15 to 4000 (½ to 160″). Medium temperature up to +130 °C (+266 °F). -
Transmitter features
Safe operation – no need to open the device due to display with touch control, background lighting. Full remote access – web server. Integrated diagnostics, verification and monitoring – Heartbeat Technology.
Transmitter housing made of durable polycarbonate or aluminium. Remote version for wall mounting. Integrated data logger: measured values monitoring. -
Nominal diameter range
DN15 to 4000 (1/2 to 160″)
-
Sensor materials
Clamp on system:
Sensor holder 1.4301 (304), 1.4404 (316L)
Sensor housing 1.4301 (304), 1.4404 (316L)
Strapping bands 1.4301 (304), 1.4404 (316L) -
Measured variables
Volume flow, Flow velocity, Sound velocity
-
Max. measurement error
Volume flow:
±3% o.r. for DN15
±2% o.r. for DN25 to 200
±2% o.r. above DN200 -
Measuring range
0 to 15 m/s (0 to 50 ft/s)
-
Max. process pressure
N/A
-
Medium temperature range
–40 to +130 °C (–40 to +266 °F)
-
Ambient temperature range
Sensor DN 15 to 65: -40 to 130°C
Sensor DN 50 to 4000: -40 to 130°C -
Sensor housing material
N/A
-
Transmitter housing material
Wall-mounted housing
Polycarbonat; AlSi10Mg, coated -
Degree of protection
Transmitter: IP66/67, Type 4X enclosure
Sensor: IP66/67, Type 4X enclosure, IP68, Type 6P enclosure (optional) -
Display/Operation
4‐line backlit display with touch control (operation from outside)
Configuration via local display, web browser and operating tools possible -
Outputs
3 ouputs:
4‐20 mA/4‐20 mA HART (active)
Pulse/frequency/switch output (passive)
Pulse/frequency/switch output (passive) -
Inputs
Status input
-
Digital communication
HART
-
Power supply
AC 100 to 240 V / AC/DC 24 V
-
Approvals
cCSAus
-
Other approvals and certificates
Other approvals and certificates
-
Product safety
CE, C-tick marking, EAC-marking
-
Proline Prosonic Flow P 500 ultrasonic clamp-on flowmeter
Reliable, flexible, zero maintenance – noninvasive flow measurement, consistently accurate, even under challenging conditions
-
Measuring principle
Ultrasonic flow
-
Product headline
Clamp-on flowmeter for limited spaces in process industries with up to 3 I/Os.
Constant accuracy even when mounted with short inlet run thanks to FlowDC.
Bidirectional measurement of various fluids, e.g. liquid hydrocarbons and chemicals. -
Sensor features
High safety standards – SIL by design,
international hazardous area approvals. Long-term stable signal – maintenance-free
permanent mounting from outside with coupling pads. Reliable measurement on various pipe materials – sensor for GRP and plastic pipes available.
Mounting without process interruption. Wide nominal diameter range: DN 15 to 4000 (½ to 160″). Medium temperature: –40 to +550 °C (–40 to +1022 °F). -
Transmitter features
Full access to process and diagnostic information – numerous, freely combinable I/Os. Reduced complexity and variety –
freely configurable I/O functionality. Integrated verification – Heartbeat Technology.
Remote version with up to 3 I/Os. Backlit display with touch control and WLAN access. Standard cable between sensor and transmitter. -
Nominal diameter range
Dual channel, 1 or 2 paths : DN 15 to 4000 (1/2 to 160″)
-
Sensor materials
Clamp on system:
Sensor holder 1.4301/304, 1.4404/316L
Sensor housing 1.4301/304, 1.4404/316L
Strapping bands 1.4301/304, 1.4404/316L -
Measured variables
Volume flow, sound velocity, flow velocity, totalizer
-
Max. measurement error
Volume flow:
±3% o.r. for DN 15
±2% o.r. for DN 25 to DN 200
±2% o.r. above DN 200 -
Measuring range
0 to 15 m/s (0 to 50 ft/s)
-
Max. process pressure
N/A
-
Medium temperature range
DN 15 to 65 (½to 2½”): –40 to +150°C (–40 +302 °F)
DN 50 to 4000 (2 to 160″): –40 to +170°C (–40 to +338 °F)
DN 50 to 600 (2″ to 24″): 150°C to 550°C (+302 to +1022 °F) -
Ambient temperature range
–40 to +60 °C (–40 to +140 °F)
–50 to +60 °C (–58 to +140 °F) optional -
Sensor housing material
N/A
-
Transmitter housing material
AlSi10Mg, coated; 1.4409 (CF3M) similar to 316L
-
Degree of protection
Sensor remote version: IP68 type 6P
Transmitter remote version: IP66/67, Type 4X enclosure -
Display/Operation
4-line backlit display with touch control(operation from outside)
Optional: WLAN
Configuration via local display and operating tools possible -
Outputs
3 outputs:
4-20 mA HART (active/passive)
4-20 mA (active/passive)
Pulse/frequency/switch output (active/passive)
Double pulse output (active/passive)
Relay output -
Inputs
Status input
4-20 mA input -
Digital communication
HART, Modbus RS485
-
Proline Prosonic Flow I 400 ultrasonic flowmeter
Insertion flowmeter with Heartbeat Technology and web server for the water and wastewater industry
-
Measuring principle
Ultrasonic flow
-
Product headline
Insertion meter with Heartbeat Technology and web server for the water and wastewater industry.
Bidirectional flow measurement of water and wastewater. -
Sensor features
Low capital investment – cost-effectiveness increases with pipe diameter (up to DN 2000/80″). Long-term stable signal – maintenance-free permanent mounting from outside with coupling pads. Process transparency – diagnostic capability.
Supports two parallel measurement paths. For large diameters: DN 200 to 4000 (8 to 160″). Medium temperatur: –40 to +80 °C (–40 to +176 °F). -
Transmitter features
Safe operation – no need to open the device due to display with touch control, background lighting. Full remote access – web server. Integrated diagnostics, verification and monitoring – Heartbeat Technology.
Transmitter housing made of durable polycarbonate or aluminium. Remote version for wall mounting. Integrated data logger: measured values monitoring. -
Nominal diameter range
DN200 to 4000 (8 to 160″)
-
Wetted materials
Insertion system:
Sensor holder 1.4301 (304), 1.4404 (316L)
Sensor housing 1.4301 (304), 1.4404 (316L) -
Measured variables
Volume flow, Flow velocity, Sound velocity
-
Max. measurement error
Volume flow: ±2% o.r. above DN200
-
Measuring range
0 to 15 m/s (0 to 50 ft/s)
-
Max. process pressure
N/A
-
Medium temperature range
–40 to +80 °C (–40 to +176 °F)
-
Ambient temperature range
-20 to 60°C (-4 to 140°F)
-
Sensor housing material
N/A
-
Transmitter housing material
Wall-mounted housing:
Polycarbonat; AlSi10Mg, coated -
Degree of protection
Transmitter: IP66/67, Type 4X enclosure
Sensor: IP66/67, Type 4X enclosure, IP68, Type 6P enclosure (optional) -
Display/Operation
4‐line backlit display with touch control (operation from outside)
Configuration via local display, web browser and operating tools possible -
Outputs
3 ouputs:
4‐20 mA/4‐20 mA HART (active)
Pulse/frequency/switch output (passive)
Pulse/frequency/switch output (passive) -
Inputs
Status input
-
Digital communication
HART
-
Power supply
AC 100 to 240 V / AC/DC 24 V
-
Approvals
cCSAus
-
Other approvals and certificates
Other approvals and certificates
-
Product safety
CE, C-tick marking, EAC-marking
-
Metrological approvals and certificates
Flowmeter verification
Verification performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a -
Marine approvals and certificates
Marine approvals and certificates
Field of application
Prosonic Flow I 400 combines insertion sensor technology with the benefits of our Proline 400 transmitters, such as Heartbeat Technology for safe process monitoring and web server access for easy installation and operation. Welded directly into the pipe wall, Prosonic Flow I 400 is especially suitable for large pipelines such as water distribution lines and projects where a fixed device installation is necessary.
The measuring principle is independent of density and conductivityBidirectional flow measurement of water and wastewaterDevice properties:Supports two parallel measurement pathsFor large diameters: DN 200 to 4000 (8 to 160″)Medium temperatur: –40 to +80 °C (–40 to +176 °F)Transmitter housing made of durable polycarbonate or aluminiumRemote version for wall mountingIntegrated data logger: measured values monitoring
Benefits
Low capital investment – cost-effectiveness increases with pipe diameter (up to DN 4000/160″) Long-term stable signal – maintenance-free direct installation of the sensor into the medium Process transparency – diagnostic capability Safe operation – no need to open the device due to display with touch control, background lighting Full remote access – web server Integrated diagnostics, verification and monitoring – Heartbeat Technology
Proline Prosonic Flow G 500 ultrasonic flowmeter
Highly robust gas specialist for fluctuating process conditions as remote version with up to 4 I/Os
-
Measuring principle
Ultrasonic flow
-
Product headline
Highly robust gas specialist for fluctuating process conditions as remote version with up to 4 I/Os.
Flexible device with user-definable gas mixtures for demanding measuring tasks.
Accurate measurement of natural and process gas in the chemical as well as oil and gas industries. -
Sensor features
Maximum reliability even with humid or wet gas – sensor design insensitive to condensate. High-performance process control – real-time pressure- and temperature-compensated values. Efficient solution – multivariable, no pressure loss.
Direct measurement: flow, pressure & temperature. Wetted parts: titanium / 316L. Maximum measuring accuracy: 0.5 %. -
Transmitter features
Full access to process and diagnostic information – numerous, freely combinable I/Os. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology.
Remote version with up to 4 I/Os. Backlit display with touch control and WLAN access. Standard cable between sensor and transmitter. -
Nominal diameter range
DN 25 to 300 (1 to 12″)
-
Wetted materials
Measuring tube: 1.4408/1.4409 (CF3M)
Transducer: 1.4404 (316, 316L, Titan Grade 2 -
Measured variables
Volume flow, corrected volume flow, mass flow, flow velocity, speed of sound, pressure, temperature, density, dynamic viscosity, energy flow, Wobbe index, methane fraction, calorific value, molar mass
-
Max. measurement error
Volume flow (standard):
‐ ±1.0 % o.r. for 3 to 40 m/s (9.84 to 131.23 ft/s)
‐ ±2 % o.r. for 0.3 to 3 m/s (0.98 to 9.84 ft/s)
Volume flow (optional calibration):
‐ ±0.5 % o.r. for 3 to 40 m/s (9.84 to 131.23 ft/s)
‐ ±1.0 % o.r. for 0.3 to 3 m/s (0.98 to 9.84 ft/s)
Corrected volume flow (standard):
‐ ±1.5 % o.r. for 3 to 40 m/s (9.84 to 131.23 ft/s)
‐ ±2.5 % o.r. for 0.3 to 3 m/s (0.98 to 9.84 ft/s)
Corrected volume flow (optional calibration):
‐ ±1.0 % o.r. for 3 to 40 m/s (9.84 to 131.23 ft/s)
‐ ±1.5 % o.r. for 0.3 to 3 m/s (0.98 to 9.84 ft/s)
Sound Velocity: ±0.2 % o.r. -
Measuring range
Gas: 0.3 m/s to 40 m/s
-
Max. process pressure
0.7 to 101 bar a (10.15 to 1464.88 psi a)
-
Medium temperature range
-50 to 150 °C (-58 to +302°F)
-50 to 100 °C (-58 to +212°F) with integrated pressure cell -
Ambient temperature range
-40 to 60 °C(-40 to +140 °F)
Optional: -50 to 60 °C(-58 to +140 °F) -
Sensor housing material
Stainless Steel, 1.4404(316/316L), 1.4408/1.4409 (CF3M)
-
Transmitter housing material
AlSi10Mg, coated; 1.4409 (CF3M) similar to 316L
Polycarbonate -
Degree of protection
Sensor remote version: IP66/67, type 4X enclosure
Transmitter remote version: IP66/67, Type 4X enclosure -
Display/Operation
4-line backlit display with Touch Control (operation from outside)
Configuration via local display and operating tools possible -
Outputs
4 outputs:
4-20 mA HART (active/passive)
4-20 mA (active/passive)
Pulse/frequency/switch output (active/passive)
Double pulse output (active/passive)
Relay output -
Inputs
Status input
4-20 mA input -
Digital communication
HART, Modbus RS485
-
Power supply
AC 100 to 230 V / DC 24 V (non-hazardous area)
-
Hazardous area approvals
ATEX, IECEx, cCSAus, JPN, EAC, UK Ex, KC
-
Product safety
CE, C-tick
-
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a -
Pressure approvals and certificates
PED, CRN
-
Material certificates
3.1 material
NACE MR0175/MR0103
Field of application
For a wide range of gas applications Prosonic Flow G provides reliable flow measurement, even with wet gas and changing gas properties and compositions. A pressure-rated sensor housing with rupture disc limits safety risks. The innovative remote transmitter maximizes installation flexibility and operational safety in demanding environments. Heartbeat Technology ensures compliance and process safety at all times.
The measuring principle is unaffected by gas composition
Accurate measurement of natural and process gas in the chemical as well as oil and gas industries.
Device properties:
Direct measurement: flow, pressure & temperature
Wetted parts: titanium / 316L
Maximum measuring accuracy: 0.5 %
Remote version with up to 4 I/Os
Backlit display with touch control and WLAN access
Standard cable between sensor and transmitter
Benefits
Flexible device with user-definable gas mixtures for demanding measuring tasks
Maximum reliability even with humid or wet gas – sensor design insensitive to condensateHigh-performance process control – real-time pressure- and temperature-compensated valuesEfficient solution – multivariable, no pressure lossFull access to process and diagnostic information – numerous, freely combinable I/OsReduced complexity and variety – freely configurable I/O functionalityIntegrated verification – Heartbeat Technology
Prosonic Flow E Heat ultrasonic flowmeter
Industrial, certified heat flow sensor for improved energy consumption measurement
-
Measuring principle
Ultrasonic flow
-
Product headline
Industrial, certified heat flow sensor for improved energy consumption measurement.
Full compliance with custody transfer regulations.
Best choice for energy management of water (e.g. heating and cooling) across all industries. -
Sensor features
Long-term stability – reliable sensor with robust industrial design. Energy and cost savings – optimized sensor for fully insulated pipes. Dependable flow measurement – high turndown.
Accuracy Class 2 according to international approvals such as MI-004, EN 1434, OIML R75. Entire sensor housing made of stainless steel. Process temperatures up to 150 °C (302 °F). -
Transmitter features
Effortless, safe operation – no commissioning needed, no unauthorized device access due to locked pulse output. Simple process indication – direct reading of status information via color LEDs. Increased reliability – comprehensive diagnostics.
Certified pulse output. Cost-efficient, application-optimized transmitter. -
Nominal diameter range
DN 50 to 150 (2″to 6″)
-
Wetted materials
Measuring tube: 1.4301 (F304)
Process Connection: 1.4571; 1.4404 (F316L); 1.0038 (S235JR); 1,4306 (F304L); 1.4307 (F304L); A105 -
Measured variables
Flow velocity, sound velocity
-
Max. measurement error
MID 004 accuracy class II (2%)
-
Measuring range
0.025 to 5 m/s (0.02 to 16.4 ft/s)
0 to 6360 dm3/min (0 to 1680 gal/min) -
Max. process pressure
PN 25/ ASME Cl. 150
-
Medium temperature range
0 to 150° (17.8 to 302 °F)
-
Ambient temperature range
-20 to 60 °C (-4 to 140 °F)
-
Sensor housing material
Stainless steel
-
Transmitter housing material
Compact: AlSi10Mg, coated
-
Degree of protection
Standard: IP66/67, Type 4X enclosure
With opened housing: IP20, Type 1 enclosure -
Display/Operation
direct reading of status information via color LEDs
-
Outputs
Pulse/Frequency
-
Inputs
None
-
Digital communication
None
-
Power supply
DC 19.2 to 28.8 V
-
Hazardous area approvals
None
-
Product safety
CE, C-Tick, EAC marking
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
MI-004 Thermal Energy
OIML R75 Class 2 -
Pressure approvals and certificates
PED
-
Material certificates
3.1 material
Field of application
Prosonic Flow E Heat is the perfect heat flow sensor for enhanced energy management throughout all industries. It offers recognized custody transfer approvals for heating and cooling applications and is optimized for flexible industrial use.
The measuring principle is independent of pressure, density, temperature and conductivity
Best choice for energy management of water (e.g. heating and cooling) across all industries.
Device properties:
Accuracy Class 2 according to international approvals such as MI-004, EN 1434, OIML R75
Entire sensor housing made of stainless steel
Process temperatures up to 150 °C (302 °F)
Certified pulse output
Cost-efficient, application-optimized transmitter
Benefits
Full compliance with custody transfer regulationsLong-term stability – reliable sensor with robust industrial designEnergy and cost savings – optimized sensor for fully insulated pipesDependable flow measurement – high turndownEffortless, safe operation – no commissioning needed, no unauthorized device access due to locked pulse outputSimple process indication – direct reading of status information via color LEDsIncreased reliability – comprehensive diagnostics
Proline Prosonic Flow E 100 ultrasonic flowmeter
Economical Ultrasonic flowmeter with integrated temperature measurement
-
Measuring principle
Ultrasonic flow
-
Product headline
Economical Ultrasonic flowmeter with integrated temperature measurement.
Bidirectional measuring of demineralized water applications for Utilities, e.g. boiler condensate return lines. -
Sensor features
Long-term stability – reliable, robust sensor. Reducing further measuring point – multivariable device. Dependable flow measurement – high turndown (200:1).
Measurement accuracy up to ±0.5 % (flow) or according to EN 1434 Cl. 2, ±2.0 °C (± 3.6 °F) (temperature). Process temperatures up to 150 °C (302 °F). Entire meter housing made of stainless steel. -
Transmitter features
Time-saving local operation without additional software and hardware – integrated web server. Extended calibration intervals – integrated device verification due to Heartbeat Technology. Easy commissioning – brief parameter explanations.
4-20 mA HART, pulse/frequency output. Local display for reading and monitoring available. Robust transmitter housing. -
Nominal diameter range
DN 50 to 150 (2″to 6″)
-
Wetted materials
Measuring tube: 1.4301 (F304)
Process Connection: 1.4571; 1.4404 (F316L); 1.0038 (S235JR); 1,4306 (F304L); 1.4307 (F304L); A105 -
Measured variables
Volume flow, calculated mass flow, flow velocity, sound velocity, temperature
-
Max. measurement error
Volume flow:
– ± 0.5 % o.r. ± 0.02% o.f.s. for 0.5 to 10 m/s (1.6 to 33 ft/s)
– ± 0.07 % o.f.s. for flow velocities < 0.5 m/s (1.64 ft/s) -
Measuring range
0.025 to 5 m/s (0.02 to 16.4 ft/s)
0 to 6360 dm3/min (0 to 1680 gal/min) -
Max. process pressure
PN 25 / ASME Cl. 150
-
Medium temperature range
0 to 150° (17.8 to 302 °F)
-
Ambient temperature range
-20 to 60 °C (-4 to 140 °F)
-
Sensor housing material
Stainless steel
-
Transmitter housing material
Compact:
AlSi10Mg, coated
Stainless steel, 1.4301 -
Degree of protection
Standard: IP66/67, Type 4X enclosure
With opened housing: IP20, Type 1 enclosure -
Display/Operation
4‐line backlit display available (no local operation)
Configuration via web browser and operating tools possible -
Outputs
4‐20 mA HART (active)
Pulse/frequency/switch output (passive) -
Inputs
None
-
Power supply
DC 19.2 to 28.8 V
-
Hazardous area approvals
None
-
Product safety
CE, C-Tick, EAC marking
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)
Calibration according to EN 1434 Class 2 -
Pressure approvals and certificates
PED
-
Material certificates
3.1 material
Field of application
Measuring of demineralized water in utilities, for example in boiler condensate return lines, requires reliable devices. Prosonic Flow E 100 provides economical and accurate bidirectional flow measurement across all industries. With its integrated temperature measurement, Prosonic Flow E 100 proves to be a multivariable flowmeter for considerable additional value in your plant. Heartbeat Technology enables measurement reliability and extension of recalibration cycles.
The measuring principle is independent of pressure, density, temperature and conductivityBidirectional measuring of demineralized water applications for utilities, e.g. boiler condensate return linesDevice properties:Accuracy up to ±0.5 % (flow) or acc. to EN 1434 Cl. 2, ±2.0 °C (± 3.6 °F) (temp.)Process temperatures up to 150 °C (302 °F)Entire meter housing made of stainless steel4-20 mA HART, pulse/frequency outputLocal display for reading and monitoring availableRobust transmitter housing
Benefits
Long-term stability – reliable, robust sensorReducing further measuring point – multivariable deviceDependable flow measurement – high turndown (200:1)Time-saving local operation without additional software and hardware – integrated web serverExtended calibration intervals – integrated device verification due to Heartbeat TechnologyEasy commissioning – brief parameter explanations
Proline Prosonic Flow B 200 ultrasonic flowmeter
The device for accurate, reliable biogas measurement with loop-powered technology
-
Measuring principle
Ultrasonic flow
-
Product headline
The device for accurate, reliable biogas measurement with loop-powered technology.
Inline flowmeter designed for measurement of wet biogas and digester gas under fluctuating process conditions. -
Sensor features
Optimized for low pressure gas – specialized sensor design. No additional pressure loss – full-bore design. Process transparency – diagnostic capability.
Multivariable device: flow, temperature and methane. Medium temperature: 0 to 80 °C (32 to 176 °F). Process pressure: 0.7 to 11 bar a (10.2 to 159 psi a). -
Transmitter features
Convenient device wiring – separate connection compartment. Safe operation – no need to open the device due to display with touch control, background lighting. Integrated verification – Heartbeat Technology.
Loop-powered technology. Robust dual-compartment housing. Plant safety: worldwide approvals. -
Nominal diameter range
Single‐path version: DN 50 (2″), DN 80 (3″)
Two‐path version: DN 100 to 200 (4 to 8″) -
Wetted materials
Sensor: 1.4404/1.4435 (316L)
Transducer: HNBR
Temperature sensor (option): AFM 34 -
Measured variables
Volume flow, corrected volume flow, corrected methane volume flow, energy flow, methane fraction, calorific value, temperature
-
Max. measurement error
Volume flow (standard):
‐ ±1.5 % o.r. for 3 to 30 m/s (9.84 to 98.4 ft/s)
‐ ±3 % o.r. for 1 to 3 m/s (3.28 to 9.84 ft/s)
Volume flow (option):
‐ ±0.1 % o.f.s. for 0.3 to 1 m/s (0.98 to 3.28 ft/s)
‐ ±1.5 % o.r. for 1 to 30 m/s (3.28 to 98.4 ft/s) -
Measuring range
Standard: 1 to 30 m/s (3.28 to 98.4 ft/s)
Option: 0.3 to 30 m/s (0.98 to 98.4 ft/s) -
Max. process pressure
11 bar a (159 psi a)
-
Medium temperature range
0 to +80 °C (+32 to +176 °F)
-
Ambient temperature range
Flange material carbon steel: –10 to +60 °C (+14 to +140 °F)
Flange material stainless steel: –40 to +60 °C (–40 to +140 °F) -
Transmitter housing material
AlSi10Mg, coated; 1.4404 (316L)
-
Degree of protection
IP66/67, type 4X enclosure
-
Display/Operation
4‐line backlit display with touch control
(operation from outside)
Configuration via local display and operating tools possible -
Outputs
4‐20 mA HART (passive)
4‐20 mA (passive)
Pulse/frequency/switch output (passive) -
Inputs
Current input 4‐20 mA (passive)
-
Digital communication
HART
-
Power supply
DC 12 to 35 V (4‐20 mA HART with/without pulse/frequency/switch output)
DC 12 to 30 V (4‐20 mA HART, 4‐20 mA)
DC 12 to 35 V (4‐20 mA HART, pulse/frequency/switch output, 4‐20 mA input)
DC 9 to 32 V (PROFIBUS PA, pulse/frequency/switch Output) -
Hazardous area approvals
ATEX, IECEx, cCSAus, NEPSI, INMETRO, JPN
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
-
Pressure approvals and certificates
PED
-
Material certificates
3.1 material (wetted parts)
Field of application
Prosonic Flow B 200 is the specialist for accurate and reliable measurement of wet biogas and digester gas,even under fluctuating process conditions. With its unique integrated real-time methane fraction analysis, Prosonic Flow B 200 enables continuous measurement and monitoring of gas flow as well as gas quality. In addition to efficient process control and energy balancing, Heartbeat Technology enables compliance and process safety at all times.
The measuring principle is unaffected by gas compositionInline flowmeter designed for wet biogas and digester gas under fluctuating process conditions.Device properties:Multivariable device: flow, temperature and methaneMedium temperature: 0 to 80 °C (32 to 176 °F)Process pressure: 0.7 to 11 bar a (10.2 to 159 psi a)Loop-powered technologyRobust dual-compartment housingPlant safety: worldwide approvals
Benefits
Integrated real-time methane fraction measurementOptimized for low pressure gas – specialized sensor designNo additional pressure loss – full bore designProcess transparency – diagnostic capabilityConvenient device wiring – separate connection compartmentSafe operation – no need to open the device due to display with touch control, background lighting Integrated verification – Heartbeat Technology
Proline Promass X 500 Coriolis flowmeter
Highest capacity four-tube flowmeter, as remote version with up to 4 I/Os
-
Measuring principle
Coriolis
-
Product headline
Highest capacity four-tube flowmeter, as remote version with up to 4 I/Os.
For highest flow rates and outstanding performance in on/offshore oil and gas applications. -
Sensor features
Increased profit – single installation point providing premium accuracy for large quantities. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs.
Nominal diameter: DN 300 to 400 (12 to 16″). Four-tube system with low pressure drop. Complete exterior design made of 1.4435 (316L). -
Transmitter features
Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology.
Remote version with up to 4 I/Os. Backlit display with touch control and WLAN access. Standard cable between sensor and transmitter. -
Nominal diameter range
DN 300 to 400 (12 to 16″)
-
Wetted materials
Measuring tube: 1.4404 (316/316L)
Connection: 1.4404 (316/316L) -
Measured variables
Mass flow, density, temperature, volume flow, corrected volume flow (API tables), reference density, concentration
-
Max. measurement error
Mass flow (liquid): ±0.10 % (standard), 0.05 % (option)
Volume flow (liquid): ±0.10 %
Mass flow (gas): ±0.35 %
Density (liquid): ±0.0005 g/cm³ -
Measuring range
0 to 4100 t/h (0 to 4520 tn. sh./h)
-
Max. process pressure
PN 100, Class 600
-
Medium temperature range
–50 to +180 °C (–58 to +356 °F)
-
Ambient temperature range
Standard: –40 to +60 °C (–40 to +140 °F)
Option: –60 to +60 °C (–76 to +140 °F) -
Sensor housing material
1.4404 (316L), highest corrosion resistance
Sensor connection housing (standard): 1.4409 (CF3M), similar to 316L -
Transmitter housing material
AlSi10Mg, coated; 1.4409 (CF3M) similar to 316L; Polycarbonat
-
Degree of protection
Sensor remote version (standard): IP66/67, type 4X enclosure
Sensor remote version (option): IP69. Transmitter remote version: IP66/67, Type 4X enclosure -
Display/Operation
4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible -
Outputs
4 outputs:
4-20 mA HART (active/passive)
4-20 mA WirelessHART
4-20 mA (active/passive)
Pulse/frequency/switch output (active/passive)
Double pulse output (active/passive)
Relay output -
Inputs
Status input
4-20 mA input -
Digital communication
HART, PROFIBUS DP, PROFIBUS PA, FOUNDATION Fieldbus, Modbus RS485, PROFINET, PROFINET over Ethernet-APL, Ethernet/IP, OPC-UA
-
Power supply
DC 24 V
AC 100 to 230 V
AC 100 to 230 V / DC 24 V (non-hazardous area) -
Hazardous area approvals
ATEX, IECEx, cCSAus, NEPSI, INMETRO, EAC, KC
-
Product safety
CE, C-tick, EAC marking
-
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)
MI-005 Liquids other than water (Hydrocarbons), LPG, cryogenic
MI-002, PTB -
Marine approvals and certificates
LR approval, DNV approval, ABS approval, BV approval, CCS approval
-
Proline Promass S 500 Coriolis flowmeter
Easy-to-clean device with self-drainable single-tube system, as remote version with up to 4 I/Os
-
Measuring principle
Coriolis
-
Product headline
Easy-to-clean device with self-drainable single-tube system, as remote version with up to 4 I/Os.
Dedicated to applications requiring optimal cleanability under hygienic conditions. -
Sensor features
Increased process safety – easily cleanable and fully self-drainable tube design. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs.
Large range of hygienic process connections. 3-A and EHEDG conform. Fast recovery from CIP/SIP. -
Transmitter features
Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology.
Remote version with up to 4 I/Os; hygienic sensor connection housing with IP69. Backlit display with touch control and WLAN access. Standard cable between sensor and transmitter. -
Nominal diameter range
DN 8 to 50 (⅜ to 2″)
-
Wetted materials
Measuring tube: 1.4435 (316L)
Connection: 1.4435 (316L); 1.4404 (316/316L) -
Measured variables
Mass flow, density, temperature, volume flow, corrected volume flow, reference density, concentration
-
Max. measurement error
Mass flow (liquid): ±0.10 %
Volume flow (liquid): ±0.10 %
Density (liquid): ±0.0005 g/cm³ -
Measuring range
0 to 70 000 kg/h (0 to 2570 lb/min)
-
Max. process pressure
PN 40, Class 150, 20K
-
Medium temperature range
–50 to +150 °C (–58 to +302 °F)
-
Ambient temperature range
Standard: –40 to +60 °C (–40 to +140 °F)
Option: –50 to +60 °C (–58 to +140 °F) -
Sensor housing material
1.4301 (304), corrosion resistant
Sensor connection housing (standard): AlSi10Mg, coated
Sensor connection housing (option): 1.4301 (304); 1.4404 (316L) -
Transmitter housing material
AlSi10Mg, coated; Polycarbonat
-
Degree of protection
Sensor remote version (standard): IP66/67, type 4X enclosure
Sensor remote version (option): IP69. Transmitter remote version: IP66/67, Type 4X enclosure -
Display/Operation
4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible -
Outputs
4 outputs:
4-20 mA HART (active/passive)
4-20 mA WirelessHART
4-20 mA (active/passive)
Pulse/frequency/switch output (active/passive)
Double pulse output (active/passive)
Relay output -
Inputs
Status input
4-20 mA input -
Digital communication
HART, PROFIBUS DP, PROFIBUS PA, FOUNDATION Fieldbus, Modbus RS485, PROFINET, PROFINET over Ethernet-APL, Ethernet/IP, OPC-UA
-
Power supply
DC 24 V
AC 100 to 230 V
AC 100 to 230 V / DC 24 V (non-hazardous area) -
Hazardous area approvals
ATEX, IECEx, cCSAus, NEPSI, INMETRO, EAC, KC
-
Product safety
CE, C-tick, EAC marking
-
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation) -
Pressure approvals and certificates
PED, CRN
-
Material certificates
3.1 material
-
Hygienic approvals and certificates
3-A, EHEDG, cGMP
-
Measuring principle
Coriolis
-
Product headline
Proline Promass S 300 Coriolis flowmeter
Easy‐to‐clean, self‐drainable single‐tube system with a compact, easily accessible transmitter
-
Measuring principle
Coriolis
-
Product headline
Easy‐to‐clean, self‐drainable single‐tube system with a compact, easily accessible transmitter.
Dedicated to applications requiring optimal cleanability under hygienic conditions. -
Sensor features
Increased process safety – easily cleanable and fully self-drainable tube design. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs.
Large range of hygienic process connections. 3-A and EHEDG conform. Fast recovery from CIP/SIP. -
Transmitter features
Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology.
Compact hygienic dual-compartment housing with IP69 and up to 3 I/Os. Backlit display with touch control and WLAN access. Remote display available. -
Nominal diameter range
DN 8 to 50 (⅜ to 2″)
-
Wetted materials
Measuring tube: 1.4435 (316L)
Connection: 1.4435 (316L); 1.4404 (316/316L) -
Measured variables
Mass flow, density, temperature, volume flow, corrected volume flow, reference density, concentration
-
Max. measurement error
Mass flow (liquid): ±0.10 %
Volume flow (liquid): ±0.10 %
Density (liquid): ±0.0005 g/cm³ -
Measuring range
0 to 70 000 kg/h (0 to 2570 lb/min)
-
Max. process pressure
PN 40, Class 150, 20K
-
Medium temperature range
–50 to +150 °C (–58 to +302 °F)
-
Ambient temperature range
Standard: –40 to +60 °C (–40 to +140 °F)
Option: –50 to +60 °C (–58 to +140 °F) -
Sensor housing material
1.4301 (304), corrosion resistant
-
Transmitter housing material
AlSi10Mg, coated; stainless steel for hygenic transmitter design
-
Degree of protection
IP66/67, type 4X enclosure
IP69 -
Display/Operation
4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available -
Outputs
3 outputs:
4-20 mA HART (active/passive)
4-20 mA WirelessHART
4-20 mA (active/passive)
Pulse/frequency/switch output (active/passive)
Double pulse output (active/passive)
Relay output -
Inputs
Status input
4-20 mA input -
Digital communication
HART, PROFIBUS DP, PROFIBUS PA, FOUNDATION Fieldbus, Modbus RS485, PROFINET, PROFINET over Ethernet-APL, Ethernet/IP, OPC-UA
-
Power supply
DC 24 V
AC 100 to 230 V
AC 100 to 230 V / DC 24 V (non-hazardous area) -
Hazardous area approvals
ATEX, IECEx, cCSAus, NEPSI, INMETRO, EAC, UK Ex, KC
-
Product safety
CE, C-tick, EAC marking
-
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation) -
Pressure approvals and certificates
PED, CRN
-
Material certificates
3.1 material
-
Hygienic approvals and certificates
3-A, EHEDG, cGMP
-
Measuring principle
Coriolis
-
Sensor features
Increased process safety – easily cleanable and fully self-drainable tube design. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs.
Large range of hygienic process connections. 3-A and EHEDG conform. Fast recovery from CIP/SIP. -
Hygienic approvals and certificates
cGMP
-
Measuring principle
Coriolis
-
Product Headline
Easy‐to‐clean, self‐drainable single‐tube system
Proline Promass S 100 Coriolis flowmeter
Easy-to-clean flowmeter with self-drainable single-tube system and an ultra-compact transmitter
-
Measuring principle
Coriolis
-
Product headline
Easy-to-clean flowmeter with self-drainable single-tube system and an ultra-compact transmitter.
Dedicated to applications requiring optimal cleanability under hygienic conditions. -
Sensor features
Increased process safety – easily cleanable and fully self-drainable tube design. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs.
Large range of hygienic process connections. 3-A and EHEDG conform. Fast recovery from CIP/SIP. -
Transmitter features
Space‐saving transmitter – full functionality on the smallest footprint. Time‐saving local operation without additional software and hardware – integrated web server. Integrated verification – Heartbeat Technology.
Robust, ultra-compact transmitter housing. Highest degree of protection: IP69. Local display available. -
Nominal diameter range
DN 8 to 50 (⅜ to 2″)
-
Wetted materials
Measuring tube: 1.4435 (316L)
Connection: 1.4435 (316L); 1.4404 (316/316L) -
Measured variables
Mass flow, density, temperature, volume flow, corrected volume flow, reference density, concentration
-
Max. measurement error
Mass flow (liquid): ±0.10 %
Volume flow (liquid): ±0.10 %
Density (liquid): ±0.0005 g/cm³ -
Measuring range
0 to 70 000 kg/h (0 to 2570 lb/min)
-
Max. process pressure
PN 40, Class 150, 20K
-
Medium temperature range
–50 to +150 °C (–58 to +302 °F)
-
Ambient temperature range
Standard: –40 to +60 °C (–40 to +140 °F)
Option: –50 to +60 °C (–58 to +140 °F) -
Sensor housing material
1.4301 (304), corrosion resistant
-
Transmitter housing material
Compact: AlSi10Mg, coated
Compact/ultra‐compact: 1.4301 (304) -
Degree of protection
Standard: IP66/67, type 4X enclosure
Option: IP69 -
Display/Operation
4‐line backlit display available (no local operation)
Configuration via web browser and operating tools possible -
Outputs
4‐20 mA HART (active)
Pulse/frequency/switch output (passive) -
Inputs
None
-
Digital communication
HART, Modbus RS485, EtherNet/IP, PROFIBUS DP, PROFINET
-
Power supply
DC 20 to 30 V
-
Hazardous area approvals
ATEX, IECEx, cCSAus, INMETRO, NEPSI, EAC
-
Product safety
CE, C-Tick, EAC marking
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation) -
Pressure approvals and certificates
PED, CRN
-
Material certificates
3.1 material
-
Hygienic approvals and certificates
3‐A, EHEDG, cGMP
-
Measuring principle
Coriolis
-
Hygienic approvals and certificates
cGMP
-
Measuring principle
Coriolis
-
Product headline
Easy-to-clean flowmeter with self-drainable single-tube system and an ultra-compact transmitter.
Dedicated to applications requiring optimal cleanability under hygienic conditions. -
Sensor features
Increased process safety – easily cleanable and fully self-drainable tube design. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs.
Large range of hygienic process connections. 3-A and EHEDG conform. Fast recovery from CIP/SIP. -
Transmitter features
Space‐saving transmitter – full functionality on the smallest footprint. Time‐saving local operation without additional software and hardware – integrated web server. Integrated verification
Proline Promass Q 500 Coriolis flowmeter
Innovative specialist for challenging applications, as remote version with up to 4 I/Os
-
Measuring principle
Coriolis
-
Product headline
Innovative specialist for challenging applications, as remote version with up to 4 I/Os.
Secured measuring quality – unmatched accuracy of mass flow, volume flow and density.
Highest measurement performance for custody transfer, density and under tough process conditions. -
Sensor features
Optimized performance for liquids with entrained gas – MFT (Multi-Frequency Technology). Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs.
Mass flow: measurement error ±0.05 % (PremiumCal). Density: measurement error ±0.2 kg/m³. High turndown due to low pressure loss/zero point. -
Transmitter features
Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology.
Remote version with up to 4 I/Os. Backlit display with touch control and WLAN access. Standard cable between sensor and transmitter. -
Nominal diameter range
DN 25 to 250(1 to 10″)
-
Wetted materials
Measuring tube: 1.4404 (316/316L); stainless steel for cryogenic applications
Connection: 1.4404 (316/316L) -
Measured variables
Mass flow, density, temperature, volume flow, corrected volume flow, reference density, concentration
-
Max. measurement error
Mass flow (liquid): ±0.10 % (standard), 0.05 % (option)
Volume flow (liquid): ±0.10 %
Mass flow (gas): ±0.25 %
Density (liquid): +/-0.2 kg/m³ (standard), +/-0.1 kg/m³ (option) -
Measuring range
0 to 2 400 000 kg/h (0 to 88183 lb/min)
-
Max. process pressure
PN 100, Class 600, 63K
-
Medium temperature range
Standard: –50 to +205°C (–58 to +401 °F)
Option: –196 to +150 °C (–321 to +302 °F) -
Ambient temperature range
Standard: – 4 to +60 °C (–4 to +140 °F)
Option:–60 to +60 °C (–76 to +140 °F) -
Sensor housing material
1.4404 (316L), highest corrosion resistance
Sensor connection housing (standard): AlSi10Mg, coated
Sensor connection housing (option): 1.4301 (304); 1.4404 (316L); 1.4409 (CF3M) similar to 316L -
Transmitter housing material
AlSi10Mg, coated; 1.4409 (CF3M) similar to 316L; Polycarbonat
-
Degree of protection
Sensor remote version (standard): IP66/67, type 4X enclosure
Sensor remote version (option): IP69. Transmitter remote version: IP66/67, Type 4X enclosure -
Display/Operation
4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible -
Outputs
4 outputs:
4-20 mA HART (active/passive)
4-20 mA WirelessHART
4-20 mA (active/passive)
Pulse/frequency/switch output (active/passive)
Double pulse output (active/passive)
Relay output -
Inputs
Status input
4-20 mA input -
Digital communication
HART, PROFIBUS DP, PROFIBUS PA, FOUNDATION Fieldbus, Modbus RS485, PROFINET, PROFINET over Ethernet-APL, Ethernet/IP, OPC-UA
-
Power supply
DC 24 V
AC 100 to 230 V
AC 100 to 230 V / DC 24 V (non-hazardous area) -
Hazardous area approvals
ATEX, IECEx, cCSAus, NEPSI, INMETRO, EAC, KC
-
Product safety
CE, C-tick, EAC marking
-
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)
MI-005 Liquids other than water (Hydrocarbons, Liquified gases, Cryogenic liquids)
OIML R117 (Liquids other than water, Liquified gases, Cryogenic liquids)
NTEP (Liquids other than water, Cryogenic liquids)
MC (Liquids other than water, Cryogenic liquids) -
Marine approvals and certificates
LR approval, DNV approval, ABS approval, BV approval, CCS approval
Proline Promass Q 300 Coriolis flowmeter
Innovative specialist for challenging applications with a compact, easily accessible transmitter
-
Measuring principle
Coriolis
-
Product headline
Innovative specialist for challenging applications with a compact, easily accessible transmitter.
Highest measurement performance for custody transfer, density and under tough process conditions. -
Sensor features
Optimized performance for liquids with entrained gas – MFT (Multi-Frequency Technology). Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs.
Mass flow: measurement error ±0.05 % (PremiumCal). Density: measurement error ±0.2 kg/m³. High turndown due to low pressure loss/zero point. -
Transmitter features
Full access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology.
Compact dual-compartment housing with up to 3 I/Os. Backlit display with touch control and WLAN access. Remote display available. -
Nominal diameter range
DN 25 to 250(1 to 10″)
-
Wetted materials
Measuring tube: 1.4404 (316/316L); stainless steel for cryogenic applications
Connection: 1.4404 (316/316L) -
Measured variables
Mass flow, density, temperature, volume flow, corrected volume flow, reference density, concentration
-
Max. measurement error
Mass flow (liquid): ±0.10 % (standard), 0.05 % (option)
Volume flow (liquid): ±0.10 %
Mass flow (gas): ±0.25 %
Density (liquid): +/-0.2 kg/m³ (standard), +/-0.1 kg/m³ (option) -
Measuring range
0 to 2 400 000 kg/h (0 to 88183 lb/min)
-
Max. process pressure
PN 100, Class 600, 63K
-
Medium temperature range
Standard: –50 to +205°C (–58 to +401 °F)
Option: –196 to +150 °C (–321 to +302 °F) -
Ambient temperature range
Standard: – 40 to +60 °C (–40 to +140 °F)
Option:–50 to +60 °C (–58 to +140 °F) -
Sensor housing material
1.4404 (316L), highest corrosion resistance
-
Transmitter housing material
AlSi10Mg, coated; 1.4409 (CF3M) similar to 316L; stainless steel for hygenic transmitter design
-
Degree of protection
IP66/67, type 4X enclosure
IP69 -
Display/Operation
4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Remote display available -
Outputs
3 outputs:
4-20 mA HART (active/passive)
4-20 mA WirelessHART
4-20 mA (active/passive)
Pulse/frequency/switch output (active/passive)
Double pulse output (active/passive)
Relay output -
Inputs
Status input
4-20 mA input -
Digital communication
HART, PROFIBUS DP, PROFIBUS PA, FOUNDATION Fieldbus, Modbus RS485, PROFINET, PROFINET over Ethernet-APL, Ethernet/IP, OPC-UA
-
Power supply
DC 24 V
AC 100 to 230 V
AC 100 to 230 V / DC 24 V (non-hazardous area) -
Hazardous area approvals
ATEX, IECEx, cCSAus, NEPSI, INMETRO, EAC, UK Ex, KC
-
Product safety
CE, C-tick, EAC marking
-
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)
MI-005 Liquids other than water (Hydrocarbons, Liquified gases, Cryogenic liquids)
OIML R117 (Liquids other than water, Liquified gases, Cryogenic liquids)
NTEP (Liquids other than water, Cryogenic liquids)
MC (Liquids other than water, Cryogenic liquids) -
Marine approvals and certificates
LR approval, DNV approval, ABS approval, BV approval, CCS approval
-